Ultrasound transducer function: annual testing is not sufficient.

Mattias Mårtensson1, Mats Olsson, Lars-Åke Brodin.   

Abstract

AIMS: The objective was to follow-up the study 'High incidence of defective ultrasound transducers in use in routine clinical practice' and evaluate if annual testing is good enough to reduce the incidence of defective ultrasound transducers in routine clinical practice to an acceptable level. METHODS AND
RESULTS: A total of 299 transducers were tested in 13 clinics at five hospitals in the Stockholm area. Approximately 7000-15,000 ultrasound examinations are carried out at these clinics every year. The transducers tested in the study had been tested and classified as fully operational 1 year before and since then been in normal use in the routine clinical practice. The transducers were tested with the Sonora FirstCall Test System. There were 81 (27.1%) defective transducers found; giving a 95% confidence interval ranging from 22.1 to 32.1%. The most common transducer errors were 'delamination' of the ultrasound lens and 'break in the cable' which together constituted 82.7% of all transducer errors found. The highest error rate was found at the radiological clinics with a mean error rate of 36.0%. There was a significant difference in error rate between two observed ways the clinics handled the transducers. There was no significant difference in the error rates of the transducer brands or the transducers models.
CONCLUSION: Annual testing is not sufficient to reduce the incidence of defective ultrasound transducers in routine clinical practice to an acceptable level and it is strongly advisable to create a user routine that minimizes the handling of the transducers.

Mesh:

Year:  2010        PMID: 20530601     DOI: 10.1093/ejechocard/jeq072

Source DB:  PubMed          Journal:  Eur J Echocardiogr        ISSN: 1532-2114


  9 in total

1.  A multicentre survey of the condition of ultrasound probes.

Authors:  Nicholas J Dudley; Darren J Woolley
Journal:  Ultrasound       Date:  2016-08-01

Review 2.  Quality assurance of ultrasound systems: current status and review of literature.

Authors:  H Grazhdani; E David; O Ventura Spagnolo; F Buemi; A Perri; N Orsogna; S Gigli; R Chimenz
Journal:  J Ultrasound       Date:  2018-06-14

3.  [Diagnosis of and therapy for hepatocellular carcinoma].

Authors:  T F Greten; N P Malek; S Schmidt; J Arends; P Bartenstein; W Bechstein; T Bernatik; M Bitzer; A Chavan; M Dollinger; D Domagk; O Drognitz; M Düx; S Farkas; G Folprecht; P Galle; M Geißler; G Gerken; D Habermehl; T Helmberger; K Herfarth; R T Hoffmann; M Holtmann; P Huppert; T Jakobs; M Keller; J Klempnauer; F Kolligs; J Körber; H Lang; F Lehner; F Lordick; A Lubienski; M P Manns; A Mahnken; M Möhler; C Mönch; P Neuhaus; C Niederau; M Ocker; G Otto; P Pereira; G Pott; J Riemer; K Ringe; U Ritterbusch; E Rummeny; P Schirmacher; H J Schlitt; K Schlottmann; V Schmitz; A Schuler; H Schulze-Bergkamen; D von Schweinitz; D Seehofer; H Sitter; C P Straßburg; C Stroszczynski; D Strobel; A Tannapfel; J Trojan; I van Thiel; A Vogel; F Wacker; H Wedemeyer; H Wege; A Weinmann; C Wittekind; B Wörmann; C J Zech
Journal:  Z Gastroenterol       Date:  2013-11-15       Impact factor: 2.000

4.  Modeling and Experimental Characterization of Bonding Delaminations in Single-Element Ultrasonic Transducer.

Authors:  Wenxiang Ding; Maxime Bavencoffe; Marc Lethiecq
Journal:  Materials (Basel)       Date:  2021-04-27       Impact factor: 3.623

5.  Phantom-based quality assurance measurements in B-mode ultrasound.

Authors:  Vilma Mannila; Outi Sipilä
Journal:  Acta Radiol Short Rep       Date:  2013-12-11

6.  Method for automatic detection of defective ultrasound linear array transducers based on uniformity assessment of clinical images - A case study.

Authors:  Robert Lorentsson; Nasser Hosseini; Jan-Olof Johansson; Wiebke Rosenberg; Benny Stenborg; Lars Gunnar Månsson; Magnus Båth
Journal:  J Appl Clin Med Phys       Date:  2018-01-11       Impact factor: 2.102

7.  Evaluation of Imaging Parameters of Ultrasound Scanners: Baseline for Future Testing.

Authors:  Ewa Fabiszewska; Katarzyna Pasicz; Iwona Grabska; Witold Skrzyński; Wioletta Ślusarczyk-Kacprzyk; Wojciech Bulski
Journal:  Pol J Radiol       Date:  2017-12-15

8.  Reference values for fetal tissue velocity imaging and a new approach to evaluate fetal myocardial function.

Authors:  Nina Ne Elmstedt; Jonas Jj Johnson; Britta Bl Lind; Kjerstin Kfw Ferm-Widlund; Lotta Lh Herling; Magnus Mw Westgren; Lars-Åke Lab Brodin
Journal:  Cardiovasc Ultrasound       Date:  2013-08-16       Impact factor: 2.062

9.  Automated quality control of ultrasound based on in-air reverberation patterns.

Authors:  Pepijn van Horssen; Arnold Schilham; Dennis Dickerscheid; Niels van der Werf; Han Keijzers; Ronald van Almere; Joost Kuijer; Rob Peters; Mark Hofman
Journal:  Ultrasound       Date:  2017-09-29
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.