Satoshi Nakamura1, Hiroyuki Okamoto1, Akihisa Wakita1, Rei Umezawa1, Kana Takahashi1, Koji Inaba1, Naoya Murakami1, Toru Kato1, Hiroshi Igaki1, Yoshinori Ito1, Yoshihisa Abe1, Jun Itami1.
Abstract
There are many reports concerning patient-specific quality assurance (QA) for intensity-modulated radiation therapy (IMRT). However, reports about the statistical results of QA are lacking. Management methods for the results of the QA are needed, even though we have the ESTRO group recommendation that a tolerance limit of 1.96 standard deviation (SD) be established in each institution. The purpose of this study was to establish a management method for determining the tolerance limit and to report the statistical results of patient-specific QA. From April 2006 to March 2015, five linacs in the National Cancer Center, Tokyo, Japan, were used to treat 1185 patients with IMRT. Patient-specific QA was performed using an ion chamber, films, and some detectors. To establish a management method for the results, differences between the measured and calculated doses in the ion chamber were analyzed for each linac, each phantom, and each treatment site. The overall mean dose difference was 0.5 ± 1.3%, and the mean dose difference in each linac was 0.6 ± 1.2%, 0.9 ± 1.3%, -0.4 ± 1.4%, -0.1 ± 1.2% and -0.1 ± 0.9%. The difference between linacs and between treatment sites was significant (P < 0.001 and 0.01, respectively). The proportion of the dose difference within ±3% was 97.7%, and that was improved from 2006 to 2014. The results of the patient-specific QA should be managed for each linac and each treatment site in order to decide the suitable tolerance limit. Reports of statistical results will be helped if a new tolerance limit and action level will be considered.Entities:
Keywords: IMRT; Ppatient-specific QA; management method; statistical analysis; tolerance limit
Mesh:
Year: 2017 PMID: 27837121 PMCID: PMC5569959 DOI: 10.1093/jrr/rrw107
Source DB: PubMed Journal: J Radiat Res ISSN: 0449-3060 Impact factor: 2.724
Fig. 1.Distribution of the percentage dose difference between the dose measured with an ion chamber and the dose calculated with a treatment planning system.
Fig. 2.Pass rate for the patient-specific QA for all linacs. The criteria was the tolerance limit proposed by the ESTRO group, i.e. treatment plans can be passed if the measured and calculated ion chamber doses agree within ±3%.
Fig. 3.Box plot of the mean difference between the measured and calculated doses in each medical linac.
Fig. 4.(a) Box plot of the mean difference between the measured and calculated doses at each treatment site in linac E, (b) Box plot for the mean difference between the measured and calculated doses at each treatment site in linac B.
Fig. 5.Box plot of the plan's average speed of the MLC motion within jaw size at each treatment site.
Fig. 6.Box plot of the plan's average gap width of the MLC within jaw size at each treatment site.