Simplifying EPID dosimetry for IMRT treatment verification.

R Pecharromán-Gallego1, Anton Mans, Jan-Jakob Sonke, Joep C Stroom, Igor Olaciregui-Ruiz, Marcel van Herk, Ben J Mijnheer.   

Abstract

PURPOSE: Electronic portal imaging devices (EPIDs) are increasingly used for IMRT dose verification, both pretreatment and in vivo. In this study, an earlier developed backprojection model has been modified to avoid the need for patient-specific transmission measurements and, consequently, leads to a faster procedure.
METHODS: Currently, the transmission, an essential ingredient of the backprojection model, is estimated from the ratio of EPID measurements with and without a phantom/patient in the beam. Thus, an additional irradiation to obtain "open images" under the same conditions as the actual phantom/patient irradiation is required. However, by calculating the transmission of the phantom/patient in the direction of the beam instead of using open images, this extra measurement can be avoided. This was achieved by using a model that includes the effect of beam hardening and off-axis dependence of the EPID response on photon beam spectral changes. The parameters in the model were empirically obtained by performing EPID measurements using polystyrene slab phantoms of different thickness in 6, 10, and 18 MV photon beams. A theoretical analysis to verify the sensitivity of the model with patient thickness changes was performed. The new model was finally applied for the analysis of EPID dose verification measurements of step-and-shoot IMRT treatments of head and neck, lung, breast, cervix, prostate, and rectum patients. All measurements were carried out using Elekta SL20i linear accelerators equipped with a hydrogenated amorphous silicon EPID, and the IMRT plans were made using PINNACLE software (Philips Medical Systems).
RESULTS: The results showed generally good agreement with the dose determined using the old model applying the measured transmission. The average differences between EPID-based in vivo dose at the isocenter determined using either the new model for transmission and its measured value were 2.6 +/- 3.1%, 0.2 +/- 3.1%, and 2.2 +/- 3.9% for 47 patients treated with 6, 10, and 18 MV IMRT beams, respectively. For the same group of patients, the differences in mean gamma analysis (3% maximum dose, 3 mm) were 0.16 +/- 0.26%, 0.21 +/- 0.24%, and 0.02 +/- 0.12%, respectively. For a subgroup of 11 patients, pretreatment verification was also performed, showing similar dose differences at the isocenter: -1.9 +/- 0.9%, -1.4 +/- 1.2%, and -0.4 +/- 2.4%, with somewhat lower mean gamma difference values: 0.01 +/- 0.09%, 0.01 +/- 0.07%, and -0.09 +/- 0.10%, respectively. Clinical implementation of the new model would save 450 h/yr spent in measurement of open images.
CONCLUSIONS: It can be concluded that calculating instead of measuring the transmission leads to differences in the isocenter dose generally smaller than 2% (2.6% for 6 MV photon beams for in vivo dose) and yielded only slightly higher gamma-evaluation parameter values in planes through the isocenter. Hence, the new model is suitable for clinical implementation and measurement of open images can be omitted.

Entities:  

Mesh:

Year:  2011        PMID: 21452735     DOI: 10.1118/1.3547714

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  A dual two dimensional electronic portal imaging device transit dosimetry model based on an empirical quadratic formalism.

Authors:  Y I Tan; M Metwaly; M Glegg; S P Baggarley; A Elliott
Journal:  Br J Radiol       Date:  2015-05-13       Impact factor: 3.039

2.  Fast 3D dosimetric verifications based on an electronic portal imaging device using a GPU calculation engine.

Authors:  Jinhan Zhu; Lixin Chen; Along Chen; Guangwen Luo; Xiaowu Deng; Xiaowei Liu
Journal:  Radiat Oncol       Date:  2015-04-11       Impact factor: 3.481

3.  Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification.

Authors:  Ali Jomehzadeh; Parvaneh Shokrani; Mohammad Mohammadi; Alireza Amouheidari
Journal:  Adv Biomed Res       Date:  2016-11-28

4.  Feasibility of portal dosimetry for flattening filter-free radiotherapy.

Authors:  Robert W Chuter; Philip A Rixham; Steve J Weston; Vivian P Cosgrove
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

5.  A management method for the statistical results of patient-specific quality assurance for intensity-modulated radiation therapy.

Authors:  Satoshi Nakamura; Hiroyuki Okamoto; Akihisa Wakita; Rei Umezawa; Kana Takahashi; Koji Inaba; Naoya Murakami; Toru Kato; Hiroshi Igaki; Yoshinori Ito; Yoshihisa Abe; Jun Itami
Journal:  J Radiat Res       Date:  2017-07-01       Impact factor: 2.724

6.  Portal dosimetry in wedged beams.

Authors:  Hanno Spreeuw; Roel Rozendaal; Priscilla Camargo; Anton Mans; Markus Wendling; Igor Olaciregui-Ruiz; Jan-Jakob Sonke; Marcel Van Herk; Ben Mijnheer
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

7.  Electronic Portal Imaging Device-Based Three-Dimensional Volumetric Dosimetry for Intensity-modulated Radiotherapy Pretreatment Quality Assurance.

Authors:  Manikandan Arjunan; Sureka Chandra Sekaran; Biplab Sarkar; Saran Kumar Manavalan
Journal:  J Med Phys       Date:  2019 Jul-Sep

8.  Simple Electronic Portal Imager-Based Pretreatment Quality Assurance using Acuros XB: A Feasibility Study.

Authors:  Arjunan Manikandan; Sureka Chandra Sekaran; Biplab Sarkar; Sujatha Manikandan
Journal:  J Med Phys       Date:  2019-12-11

9.  Commissioning and implementation of an implantable dosimeter for radiation therapy.

Authors:  Ivan Buzurovic; Timothy N Showalter; Matthew T Studenski; Robert B Den; Adam P Dicker; Junsheng Cao; Ying Xiao; Yan Yu; Amy Harrison
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

10.  An Empirical Transmitted EPID Dosimetry Method using a Back-Projection Algorithm.

Authors:  Hashemi S M; Bahreyni M H; Mohammadi M; Nasseri S; Bayani S; Gholamhosseinian H; Salek R; Shahedi F; Momennezhad M
Journal:  J Biomed Phys Eng       Date:  2019-10-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.