Simultaneous 4D-CBCT reconstruction with sliding motion constraint.

Jun Dang1, Fang-Fang Yin2, Tao You1, Chunhua Dai1, Deyu Chen1, Jing Wang3.   

Abstract

PURPOSE: Current approaches using deformable vector field (DVF) for motion-compensated 4D-cone beam CT (CBCT) reconstruction typically utilize an isotropically smoothed DVF between different respiration phases. Such isotropically smoothed DVF does not work well if sliding motion exists between neighboring organs. This study investigated an anisotropic motion modeling scheme by extracting organ boundary local motions (e.g., sliding) and incorporated them into 4D-CBCT reconstruction to optimize the motion modeling and reconstruction methods.
METHODS: Initially, a modified simultaneous algebraic reconstruction technique (mSART) was applied to reconstruct high quality reference phase CBCT using all phase projections. The initial DVFs were precalculated and subsequently updated to achieve the optimized solution. During the DVF update, sliding motion estimation was performed by matching the measured projections to the forward projection of the deformed reference phase CBCT. In this process, each moving organ boundary was first segmented. The normal vectors of the boundary DVF were then extracted and incorporated for further DVF optimization. The regularization term in the objective function adaptively regularizes the DVF by (1) isotopically smoothing the DVF within each organ; (2) smoothing the DVF at boundary along the normal direction; and (3) leaving the tangent direction of boundary DVF unsmoothed (i.e., allowing for sliding motion). A nonlinear conjugate gradient optimizer was used. The algorithm was validated on a digital cubic tube phantom with sliding motion, nonuniform rotational B-spline based cardiac-torso (NCAT) phantom, and two anonymized patient data. The relative reconstruction error (RE), the motion trajectory's root mean square error (RMSE) together with its maximum error (MaxE), and the Dice coefficient of the lung boundary were calculated to evaluate the algorithm performance.
RESULTS: For the cubic tube and NCAT phantom tests, the REs are 10.2% and 7.4% with sliding motion compensation, compared to 13.4% and 8.9% without sliding modeling. The motion trajectory's RMSE and MaxE for NCAT phantom tests are 0.5 and 0.8 mm with sliding motion constraint compared to 3.5 and 7.3 mm without sliding motion modeling. The Dice coefficients for both NCAT phantom and the patients show a consistent trend that sliding motion constraint achieves better similarity for segmented lung boundary compared with the ground truth or patient reference.
CONCLUSIONS: A sliding motion-compensated 4D-CBCT reconstruction and the motion modeling scheme was developed. Both phantom and patient study demonstrated the improved accuracy and motion modeling accuracy in reconstructed 4D-CBCT.

Entities:  

Mesh:

Year:  2016        PMID: 27782722      PMCID: PMC5035309          DOI: 10.1118/1.4959998

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  27 in total

1.  Respiratory correlated cone beam CT.

Authors:  Jan-Jakob Sonke; Lambert Zijp; Peter Remeijer; Marcel van Herk
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

2.  Autoadaptive phase-correlated (AAPC) reconstruction for 4D CBCT.

Authors:  Frank Bergner; Timo Berkus; Markus Oelhafen; Patrik Kunz; Tinsu Pan; Marc Kachelriess
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

3.  Four-dimensional cone beam CT with adaptive gantry rotation and adaptive data sampling.

Authors:  Jun Lu; Thomas M Guerrero; Peter Munro; Andrew Jeung; Pai-Chun M Chi; Peter Balter; X Ronald Zhu; Radhe Mohan; Tinsu Pan
Journal:  Med Phys       Date:  2007-09       Impact factor: 4.071

4.  Image matching as a diffusion process: an analogy with Maxwell's demons.

Authors:  J P Thirion
Journal:  Med Image Anal       Date:  1998-09       Impact factor: 8.545

5.  Estimation of slipping organ motion by registration with direction-dependent regularization.

Authors:  Alexander Schmidt-Richberg; René Werner; Heinz Handels; Jan Ehrhardt
Journal:  Med Image Anal       Date:  2011-06-26       Impact factor: 8.545

6.  Deformation vector fields (DVF)-driven image reconstruction for 4D-CBCT.

Authors:  Jun Dang; Ouyang Luo; Xuejun Gu; Jing Wang
Journal:  J Xray Sci Technol       Date:  2015       Impact factor: 1.535

7.  A pilot evaluation of a 4-dimensional cone-beam computed tomographic scheme based on simultaneous motion estimation and image reconstruction.

Authors:  Jun Dang; Xuejun Gu; Tinsu Pan; Jing Wang
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-02-01       Impact factor: 7.038

8.  Piecewise-diffeomorphic image registration: application to the motion estimation between 3D CT lung images with sliding conditions.

Authors:  Laurent Risser; François-Xavier Vialard; Habib Y Baluwala; Julia A Schnabel
Journal:  Med Image Anal       Date:  2012-11-02       Impact factor: 8.545

9.  4D cone-beam CT reconstruction using multi-organ meshes for sliding motion modeling.

Authors:  Zichun Zhong; Xuejun Gu; Weihua Mao; Jing Wang
Journal:  Phys Med Biol       Date:  2016-01-13       Impact factor: 3.609

10.  Automated segmentation of a motion mask to preserve sliding motion in deformable registration of thoracic CT.

Authors:  Jef Vandemeulebroucke; Olivier Bernard; Simon Rit; Jan Kybic; Patrick Clarysse; David Sarrut
Journal:  Med Phys       Date:  2012-02       Impact factor: 4.071

View more
  3 in total

1.  Iterative reconstruction for dual energy CT with an average image-induced nonlocal means regularization.

Authors:  Houjin Zhang; Dong Zeng; Jiahui Lin; Hao Zhang; Zhaoying Bian; Jing Huang; Yuanyuan Gao; Shanli Zhang; Hua Zhang; Qianjin Feng; Zhengrong Liang; Wufan Chen; Jianhua Ma
Journal:  Phys Med Biol       Date:  2017-05-04       Impact factor: 3.609

2.  An unsupervised 2D-3D deformable registration network (2D3D-RegNet) for cone-beam CT estimation.

Authors:  You Zhang
Journal:  Phys Med Biol       Date:  2021-03-24       Impact factor: 4.174

3.  Fully Automatic Sliding Motion Compensated and Simultaneous 4D-CBCT via Bilateral Filtering.

Authors:  Jun Dang; Tao You; Wenzheng Sun; Hanguan Xiao; Longhao Li; Xiaopin Chen; Chunhua Dai; Ying Li; Yanbo Song; Tao Zhang; Deyu Chen
Journal:  Front Oncol       Date:  2021-01-18       Impact factor: 6.244

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.