Deformation vector fields (DVF)-driven image reconstruction for 4D-CBCT.

Jun Dang1, Ouyang Luo1, Xuejun Gu1, Jing Wang1.   

Abstract

BACKGROUND: High quality 4D-CBCT can be obtained by deforming a planning CT (pCT), where the deformation vector fields (DVF) are estimated by matching the forward projections of pCT and 4D-CBCT projections. The matching metric used in the previous study is the sum of squared intensity differences (SSID). The scatter signal level in CBCT projections is much higher than pCT, the SSID metric may not lead to optimal DVF.
OBJECTIVE: To improve the DVF estimation accuracy, we develop a new matching metric that is less sensitive to the intensity level difference caused by the scatter signal.
METHODS: The negative logarithm of correlation coefficient (NLCC) is used as the matching metric. A non-linear conjugate gradient optimization algorithm is used to estimate the DVF. A 4D NCAT phantom and an anthropomorphic thoracic phantom were used to evaluate the NLCC-based algorithm.
RESULTS: In the NCAT phantom study, the relative reconstruction error is reduced from 18.0% in SSID to 14.13% in NLCC. In the thoracic phantom study, the root mean square error of the tumor motion is reduced from 1.16 mm in SSID to 0.43 mm in NLCC.
CONCLUSION: NLCC metric can improve the image reconstruction and motion estimation accuracy of DVF-driven image reconstruction for 4D-CBCT.

Entities:  

Keywords:  4D-CBCT; DVF estimation from projection; DVF-driven image reconstruction

Mesh:

Year:  2015        PMID: 25567403     DOI: 10.3233/XST-140466

Source DB:  PubMed          Journal:  J Xray Sci Technol        ISSN: 0895-3996            Impact factor:   1.535


  2 in total

1.  Preliminary clinical evaluation of a 4D-CBCT estimation technique using prior information and limited-angle projections.

Authors:  You Zhang; Fang-Fang Yin; Tinsu Pan; Irina Vergalasova; Lei Ren
Journal:  Radiother Oncol       Date:  2015-03-26       Impact factor: 6.280

2.  Simultaneous 4D-CBCT reconstruction with sliding motion constraint.

Authors:  Jun Dang; Fang-Fang Yin; Tao You; Chunhua Dai; Deyu Chen; Jing Wang
Journal:  Med Phys       Date:  2016-10       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.