Radio-opaque polyethylene for personalized craniomaxillofacial implants.

Marcin Kozakiewicz1, Leszek Olbrzymek2, Ludomir Stefanczyk3, Marek Olszycki3, Piotr Komorowski4,5, Bogdan Walkowiak4,5, Bartłomiej Konieczny6, Michał Krasowski6, Jerzy Sokołowski6.   

Abstract

OBJECTIVE: This study aimed to present a new possibility to create radio-opaque implant material for craniomaxillofacial reconstruction.
MATERIALS AND METHODS: The test disks made of the own compound of polyethylenes with addiction of 2, 4, and 6 % of weight TiO2 was investigated for cytotoxicity [each group 15 disks respectively]. Next, computed tomography of the disks was performed in environment of muscle and fat. Hardness, tensile modulus and strength, and compressive modulus and strength were tested too.
RESULTS: Deterioration of mechanical properties of the composites containing titanium dioxide was observed [hardness, tensile modulus and strength, compressive modulus and strength, respectively: 56.7 ± 1.6 shore D, 354 ± 52, 22.5 ± 1.3, 21.8 ± 1.1, and 2995 ± 327 MPa as addiction of 2 % TiO2; 52.0 ± 0.9 shore D, 347 ± 66, 18.0 ± 0.7, 14.2 ± 0.9, and 1396 ± 477 MPa as 4 % TiO2; 51.3 ± 1.3 shore D, 316 ± 9, 17.4 ± 0.2, 13.6 ± 0.6, and 1100 ± 144 MPa as 6 % TiO2 added]. The test disks revealed no cytotoxicity effect on human osteoblasts. The new material presents mild radio-opacity which was enough to observe the implant in relation to fat and muscle, but with no visible effect of beam hardening.
CONCLUSION: In view of the performed tests, the polyethylene enriched by titanium dioxide seems to be a proper material to consider manufacturing of craniomaxillofacial implants. CLINICAL RELEVANCE: Maxilloafacial surgery is still looking for new implantologic materials. The proposed one is a new way to manufacture an implant visible in computed tomography which does not interfere with its shape in radiological examination and makes it possible to observe the surrounding soft tissues.

Entities:  

Keywords:  Maxillofacial implants; personalized treatment; polyethylene; radiological opacity

Mesh:

Substances:

Year:  2016        PMID: 27766488     DOI: 10.1007/s00784-016-1978-x

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  19 in total

1.  Evaluation of tuned-aperture computed tomography depth discrimination for image series acquired variously with linear horizontal, linear vertical, and conical beam projection arrays.

Authors:  K Yamamoto; K Nishikawa; N Kobayashi; K Kuroyanagi; A G Farman
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2000-06

2.  Comparison of microfocus- and synchrotron X-ray tomography for the analysis of osteointegration around Ti6Al4V implants.

Authors:  R Bernhardt; D Scharnweber; B Müller; P Thurner; H Schliephake; P Wyss; F Beckmann; J Goebbels; H Worch
Journal:  Eur Cell Mater       Date:  2004-06-30       Impact factor: 3.942

3.  Single-and dual-energy CT with monochromatic synchrotron x-rays.

Authors:  F A Dilmanian; X Y Wu; E C Parsons; B Ren; J Kress; T M Button; L D Chapman; J A Coderre; F Giron; D Greenberg; D J Krus; Z Liang; S Marcovici; M J Petersen; C T Roque; M Shleifer; D N Slatkin; W C Thomlinson; K Yamamoto; Z Zhong
Journal:  Phys Med Biol       Date:  1997-02       Impact factor: 3.609

4.  In vitro and in vivo imaging of ultra-high-molecular-weight polyethylene orbital implants.

Authors:  Marek Olszycki; Marcin Kozakiewicz; Marcin Elgalal; Agata Majos; Ludomir Stefanczyk
Journal:  Int J Oral Maxillofac Implants       Date:  2015 Mar-Apr       Impact factor: 2.804

5.  Computer-aided orbital wall defects treatment by individual design ultrahigh molecular weight polyethylene implants.

Authors:  Marcin Kozakiewicz
Journal:  J Craniomaxillofac Surg       Date:  2013-07-05       Impact factor: 2.078

6.  Patient specific implants (PSI) in reconstruction of orbital floor and wall fractures.

Authors:  Thomas Gander; Harald Essig; Philipp Metzler; Daniel Lindhorst; Leander Dubois; Martin Rücker; Paul Schumann
Journal:  J Craniomaxillofac Surg       Date:  2014-10-31       Impact factor: 2.078

7.  Surgical management of upgaze diplopia in patients after posttraumatic orbital floor reconstruction.

Authors:  Piotr Loba; Marcin Kozakiewicz; Anna Broniarczyk-Loba
Journal:  J Craniomaxillofac Surg       Date:  2015-04-08       Impact factor: 2.078

8.  Complications following cranioplasty: incidence and predictors in 348 cases.

Authors:  Mario Zanaty; Nohra Chalouhi; Robert M Starke; Shannon W Clark; Cory D Bovenzi; Mark Saigh; Eric Schwartz; Emily S I Kunkel; Alexandra S Efthimiadis-Budike; Pascal Jabbour; Richard Dalyai; Robert H Rosenwasser; Stavropoula I Tjoumakaris
Journal:  J Neurosurg       Date:  2015-03-13       Impact factor: 5.115

9.  Posttraumatic enophthalmos: etiology, principles of reconstruction, and correction.

Authors:  Luigi Clauser; Manlio Galiè; Francesco Pagliaro; Riccardo Tieghi
Journal:  J Craniofac Surg       Date:  2008-03       Impact factor: 1.046

10.  Development and validation of a pedestrian lower limb non-linear 3-d finite element model.

Authors:  P J Schuster; C C Chou; P Prasad; G Jayaraman
Journal:  Stapp Car Crash J       Date:  2000-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.