H Kim1, C-T Chen1, N Eclov1, A Ronzhin2, P Murat2, E Ramberg2, S Los2, C-M Kao1.
Abstract
A strip-line and waveform sampling based readout is a signal multiplexing method that can efficiently reduce the readout channels while fully exploiting the fast time characteristics of photo-detectors such as the SiPM. We have applied this readout method for SiPM-based time-of-flight (TOF) positron emission tomography (PET) detectors. We have prototyped strip-line boards in which 8 SiPMs (pitch 5.2 mm) are connected by using a single strip-line, and the signals appearing at the ends of the strip-line are acquired by using the DRS4 waveform sampler at a nominal sampling frequency of 1-5 GS/s. Experimental tests using laser and LYSO scintillator are carried out to assess the performance of the strip-line board. Each SiPM position, which is inferred from the arrival time difference of the two signals at the ends of the strip-line, is well identified with 2.6 mm FWHM resolution when the SiPMs are coupled to LYSO crystals and irradiated by a 22Na source. The average energy and coincidence time resolution responding to 511 keV photons are measured to be ~32% and ~510 ps FWHM, respectively, at a 5.0 GS/s DRS4 sampling rate. The results show that the sampling rate can be lowered to 1.5 GS/s without performance degradation. These encouraging initial test results indicate that the strip-line and waveform sampling readout method is applicable for SiPM-based TOF PET development.Entities:
Keywords: Multiplexing; Positron Emission Tomography; Strip-Line Readout; Waveform Sampling
Year: 2016 PMID: 27746514 PMCID: PMC5058446 DOI: 10.1016/j.nima.2016.05.085
Source DB: PubMed Journal: Nucl Instrum Methods Phys Res A ISSN: 0168-9002 Impact factor: 1.455