Noise Power Characteristics of a Micro-Computed Tomography System.

Muhammad U Ghani1, Liqiang Ren, Molly Wong, Yuhua Li, Bin Zheng, Xiujiang John Rong, Kai Yang, Hong Liu.   

Abstract

OBJECTIVE: The aim of this study was to investigate the noise power properties of a micro-computed tomography (micro-CT) system under different operating conditions.
METHODS: A commercial micro-CT was used in the study that used a flat panel detector with a 127-μm-pixel pitch and a micro-focus x-ray tube. Conical tubes of various diameters were used under different acquisition conditions. Multidimensional noise power spectrums were used as a metric to investigate the noise properties of the system. Noise power spectrum was calculated from the difference data generated by subtraction of 2 identical scans. The noise properties with respect to various parameters that include the impact of number of projections, x-ray spectra, milliampere-second, slice location, object diameter, voxel size, geometric magnification (M), back-projection filters, and reconstruction magnification (Mrecon) were studied.
RESULTS: At a same isocentric exposure rate of 270 mR/s, the noise power was much lower for the image reconstructed with 3672 views (122 seconds) as compared with the 511 views (17 seconds), whereas at a fixed isocentric exposure of 4600 mR, the noise power levels were almost similar. Image noise with a 50-kV beam was higher as compared with the 90-kV beam at a same isocentric exposure. Image noise from a 16-mm-diameter conical tube was much lower as compared with the 28- and 56-mm tubes under identical isocentric exposures. The choice of back-projection filter influences noise power spectrum curves in terms of width and amplitudes. Reconstruction magnification applied during the reconstruction process increased the noise power at lower spatial frequencies but reduced the noise power at higher spatial frequencies. It can be established that, for small details corresponding to high spatial frequencies, reconstruction magnification can provide an improved signal-to-noise ratio. At all spatial frequencies, the in-plane images had lower noise power levels as compared with the z-plane images.
CONCLUSIONS: The noise power properties investigated in this study provide important image quality references for refined cone beam system development, optimization, and operations.

Entities:  

Mesh:

Year:  2017        PMID: 27680419      PMCID: PMC5233553          DOI: 10.1097/RCT.0000000000000483

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  25 in total

1.  Determination of the presampled MTF in computed tomography.

Authors:  J M Boone
Journal:  Med Phys       Date:  2001-03       Impact factor: 4.071

2.  Estimating local noise power spectrum from a few FBP-reconstructed CT scans.

Authors:  Rongping Zeng; Marios A Gavrielides; Nicholas Petrick; Berkman Sahiner; Qin Li; Kyle J Myers
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071

3.  Noise power properties of a cone-beam CT system for breast cancer detection.

Authors:  Kai Yang; Alexander L C Kwan; Shih-Ying Huang; Nathan J Packard; John M Boone
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

4.  Local and global 3D noise power spectrum in cone-beam CT system with FDK reconstruction.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

5.  Characterization of imaging performance in differential phase contrast CT compared with the conventional CT--noise power spectrum NPS(k).

Authors:  Xiangyang Tang; Yi Yang; Shaojie Tang
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

6.  Multimodal imaging analysis of tumor progression and bone resorption in a murine cancer model.

Authors:  Maria L Mouchess; Yasuyoshi Sohara; Marvin D Nelson; Yves A DeCLerck; Rex A Moats
Journal:  J Comput Assist Tomogr       Date:  2006 May-Jun       Impact factor: 1.826

7.  Detectability in computed tomographic images.

Authors:  K M Hanson
Journal:  Med Phys       Date:  1979 Sep-Oct       Impact factor: 4.071

8.  Cone beam CT for determining breast cancer margin: an initial experience and its comparison with mammography and specimen radiograph.

Authors:  Juan Yao; Chris Shaw; C J Lai; John Rong; Jian Wang; Wenya Liu
Journal:  Int J Clin Exp Med       Date:  2015-09-15

9.  Cone beam breast CT with multiplanar and three dimensional visualization in differentiating breast masses compared with mammography.

Authors:  Binghui Zhao; Xiaohua Zhang; Weixing Cai; David Conover; Ruola Ning
Journal:  Eur J Radiol       Date:  2014-07-15       Impact factor: 3.528

10.  Role of computed tomography voxel size in detection and discrimination of calcium and iron deposits in atherosclerotic human coronary artery specimens.

Authors:  Alexander C Langheinrich; Marian Kampschulte; Christine Crössmann; Regina Moritz; Wigbert S Rau; Rainer M Bohle; Erik L Ritman
Journal:  J Comput Assist Tomogr       Date:  2009 Jul-Aug       Impact factor: 1.826

View more
  1 in total

Review 1.  Tutorial on X-ray photon counting detector characterization.

Authors:  Liqiang Ren; Bin Zheng; Hong Liu
Journal:  J Xray Sci Technol       Date:  2018       Impact factor: 1.535

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.