Gangarapu Sri Krishna1, Vuppu Srinivas2, K M Ayyangar3, Palreddy Yadagiri Reddy4.
Abstract
Recently, Eclipse treatment planning system (TPS) version 8.8 was upgraded to the latest version 13.6. It is customary that the vendor gives training on how to upgrade the existing software to the new version. However, the customer is provided less inner details about changes in the new software version. According to manufacturer, accuracy of point dose calculations and irregular treatment planning is better in the new version (13.6) compared to the old version (8.8). Furthermore, the new version uses voxel-based calculations while the earlier version used point dose calculations. Major difference in intensity-modulated radiation therapy (IMRT) plans was observed between the two versions after re-optimization and re-calculations. However, minor difference was observed for IMRT cases after performing only re-calculations. It is recommended TPS quality assurance to be performed after any major upgrade of software. This can be done by performing dose calculation comparisons in TPS. To assess the difference between the versions, 25 clinical cases from the old version were compared keeping all the patient data intact including the monitor units and comparing the differences in dose calculations using dose volume histogram (DVH) analysis. Along with DVH analysis, uniformity index, conformity index, homogeneity index, and dose spillage index were also compared for both versions. The results of comparative study are presented in this paper.Entities:
Keywords: Conformity index; dose spillage index; dose volume histogram; homogeneity index; treatment planning version; uniformity index
Year: 2016 PMID: 27651566 PMCID: PMC5019038 DOI: 10.4103/0971-6203.189489
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Variations in doses for a case of cervix for both versions (8.8 and 13.6) - for IMRT technique (with re-optimization and re-calculations). ▴ Triangles are for version 13.6 and ▪ rectangles for version 8.8
Summary of average values of DVH indices of 25 cases for 5 sites between the two Eclipse versions (8.8 and 13.6)
Summary of doses to normal structures for 25 clinical cases
Figure 2Isodose curves on an axial slice at isocentre plane for a case of cervix between versions 8.8 and 13.6. Lower part of the figure shows the isodose distribution on coronal and sagittal plane of the corresponding slice
Figure 3Comparison of dose volume histogram curves of intensity-modulated radiation therapy plan of representative patient for planning target volume, rectum, bladder, left femoral head, and right femoral head for 8.8 and 13.6 version treatment planning system. ▴ Triangles are for version 13.6 and ▪ rectangles for version 8.8