Performance of basic ventricular tachycardia detection algorithms in implantable cardioverter defibrillators: implications for device programming.

M H Anderson1, F D Murgatroyd, K Hnatkova, B Xie, S Jones, E Rowland, D E Ward, A J Camm, M Malik.   

Abstract

Around 20% of patients with third generation implantable cardioverter defibrillators receive inappropriate therapy, usually triggered by atrial fibrillation. This is because the criteria used for ventricular tachycardia detection by current implantable cardioverter defibrillators are based on the analysis of a sequence of RR intervals and may be inappropriately satisfied by supraventricular tachyarrhythmias. Algorithms for ventricular tachycardia detection were challenged against the full RR interval sequences from 482 spontaneous episodes of atrial fibrillation and 260 spontaneous episodes of ventricular tachycardia to determine their ability to discriminate between the arrhythmias. The sensitivities and specificities of the algorithms were calculated over a wide range of programmable parameters. For a given window length and detection interval, the most stringent algorithms, that required all beats to be classified as "fast", were more specific than those allowing a proportion of "normal" intervals, even after adjustment for differing sensitivity. These differences were less marked for faster tachycardias. Specificity increased with the detection window length to a limit of approximately 18 beats. We conclude that ventricular tachycardia is detected with the highest specificity if all beats in an analyzed sequence are required to be "fast," even after lengthening of the tachycardia detection interval to maintain sensitivity. Further improvement in algorithm performance may require the incorporation of criteria such as tachycardia onset and stability.

Entities:  

Mesh:

Year:  1997        PMID: 9455760     DOI: 10.1111/j.1540-8159.1997.tb05469.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

Review 1.  Sensing and detection in Medtronic implantable cardioverter defibrillators.

Authors:  Mark L Brown; Charles D Swerdlow
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.