Hamid Amoozgar1, Sara Salehi1, Pouya Farhadi2, Mohammad Reza Edraki1, Mohammad Borzoee1, Gholamhossein Ajami1, Sirous Cheriki1, Hamid Mohammadi1.
Abstract
BACKGROUND: Transcatheter patent ductus arteriosus (PDA) closure is an established procedure.Entities:
Keywords: Amplatzer Ductal Occlude; Coil Occlusion; Patent Ductus Arteriosus; Transcatheter Occlusion; Treatment Outcome
Year: 2016 PMID: 27617065 PMCID: PMC4988101 DOI: 10.5812/ijp.3621
Source DB: PubMed Journal: Iran J Pediatr ISSN: 2008-2142 Impact factor: 0.364
Baseline Characteristics of the Patients[a]
| Characteristics | Data |
|---|---|
|
| 43.2 ± 49.2 (1 - 312) |
| 0 - 12 | 14.3 (35.3) |
| > 12 | 261 (74.7) |
|
| 143/261 (0.5/1) |
|
| 17.8 ± 14.1 |
|
| 3.02 ± 2.8 |
|
| 7.62 ± 3.45 |
|
| 9.5 ± 2.53 |
|
| 4.37 ± 4.1 |
|
| 295/6/8/3/9/83 |
|
| |
|
| 220 (54.5) |
|
| 184 (45.5) |
Abbreviations: PDA, patent ducts arteriosus; MPAP, mean pulmonary artery pressure.
aValues are expressed as mean ± SD.
Closure Rate in Successful Deployment Group[a]
| Value | Immediately After Procedure | 2 Weeks | 2 Months | 6 Months | At Time of The Study |
|---|---|---|---|---|---|
|
| 290 (73.7) | 325 (84) | 368 (93.6) | 388 (98.7) | 390 (99.5) |
|
| 161 (75.6) | 179 (85.4) | 201 (94.3) | 209 (98.1) | 211 (99.2) |
|
| 141 (78.3) | 146 (81.1) | 167 (92.7) | 179 (99.4) | 179 (99.7) |
Abbreviation: ADO, amplatzer duct occluder.
aValues are expressed as No. (%).
Complications Among the Patients With Unsuccessful Catheterization
| Value | Age, Month | Weight | Type of Device | Time of Embolization | Retrieval |
|---|---|---|---|---|---|
| 1 | 6 | 5 | Nit-Occlud coil 6 × 5 | Just after deployment | Left in place |
| 2 | 7 | 5.5 | Nit-Occlud coil 7 × 6 | 2 weeks after | Retrieved by snaring |
| 3 | 11 | 7 | Nit-Occlud coil 7 × 6 | Just after deployment | Retrieved by snaring |
| 4 | 36 | 11 | Nit-Occlud coil 7 × 6 | Just after deployment | Retrieved by snaring |
| 5 | 8 | 6 | Nit-Occlud coil 9 × 6 | 1 day after | Surgical removal |
| 6 | 6 | 4.5 | Nit-Occlud coil 9 × 6 | Just after deployment | Surgical removal |
| 7 | 5 | 3.5 | ADO 4 × 6 | Just after deployment | Surgical removal |
| 8 | 8 | 5.8 | ADO 6 × 8 | Just after deployment | Surgical removal |
| 9 | 6 | 4.8 | ADO 6 × 8 | Just after deployment | Retrieved by snaring |
| 10 | 7 | 5 | ADO 8 × 10 | Just after deployment | Retrieved by snaring |
| 11 | 38 | 13 | ADO 12 × 14 | 2 weeks after | Retrieved by snaring |
Abbreviation: ADO, amplatzer duct occluder.
Frequency of Complicationsa
| Value | Coil, (213) | ADO, (180) |
|---|---|---|
| Embolization | 6 (2.8) | 5 (2.7) |
| Accidental separation of the devices | 1 (0.5) | 0 |
| Bradycardia or hypotension | 0 | 4 (2.2) |
| Narrowing aortic isthmus | 0 | 3 (1.6) |
| Mild LPA stenosis | 4 (1.8) | 0 |
Abbreviation: LPA, left pulmonary artery.