OBJECTIVE: To assess the results of anterograde transcatheter closure of a large sized patent ductus arteriosus using the Amplatzer Duct Occluder (ADO) device. STUDY
DESIGN: A case series. PLACE AND DURATION OF STUDY: Department of Cardiology, Armed Forces Institute of Cardiology and National Institute of Heart Diseases, Rawalpindi, from January 2005 to December 2007.
METHODOLOGY: All patients who underwent transcatheter closure of large sized PDA (> or = 4.0 mm), using the Amplatzer Duct Occluder (ADO) device, from January 2005 to December 2007, were included by convenient sampling technique. After the procedure, repeat aortogram was performed to confirm the appropriate position of the ADO and to evaluate for residual shunts.
RESULTS: A total of 100 patients had PDA diameter of > or = 4.00 mm. Mean age was 11.73 + 10.31 years and there were 37 males and 63 females. Mean PDA diameter was 6.011 + 2.078 mm. Forty-nine patients had type 'A', 19 had type 'B', 19 patients had type 'C', 5 had type 'D' and 8 patients had type 'E' PDA. Complete closure was achieved in 70 patients, 26 had trace shunt and 4 patients had small residual shunt immediately after the procedure. However, no patient had residual shunt 24 hours after the procedure.
CONCLUSION: ADO is a highly efficient prosthesis that can be safely applied in most patients with PDA particularly in patients with large sized PDA.

Entities:  

Mesh:

Year:  2009        PMID: 19409157     DOI: 05.2009/JCPSP.275278

Source DB:  PubMed          Journal:  J Coll Physicians Surg Pak        ISSN: 1022-386X            Impact factor:   0.711


  4 in total

1.  Transcatheter Closure of Patent Ductus Arteriosus in Children with the Occlutech Duct Occluder.

Authors:  Meki Bilici; Fikri Demir; Alper Akın; Mehmet Türe; Hasan Balık; Mahir Kuyumcu
Journal:  Pediatr Cardiol       Date:  2017-08-21       Impact factor: 1.655

2.  Long-term Results of Transcatheter Closure of Patent Ductus Arteriosus in Adolescents and Adults with Amplatzer Duct Occluder.

Authors:  Mostafa Behjati-Ardakani; Mansour Rafiei; Mohammad Amin Behjati-Ardakani; Mohammadreza Vafaeenasab; Mohammadtaghi Sarebanhassanabadi
Journal:  N Am J Med Sci       Date:  2015-05

3.  Long-term results of transcatheter closure of patent ductus arteriosus in infants using amplatzer duct occluder.

Authors:  Mostafa Behjati-Ardakani; Mohammad-Amin Behjati-Ardakani; Seyed-Habibolla Hosseini; Noormohammad Noori
Journal:  Iran J Pediatr       Date:  2013-08       Impact factor: 0.364

4.  Follow-Up Results of Device Occlusion of Patent Ductus Arteriosus.

Authors:  Hamid Amoozgar; Sara Salehi; Pouya Farhadi; Mohammad Reza Edraki; Mohammad Borzoee; Gholamhossein Ajami; Sirous Cheriki; Hamid Mohammadi
Journal:  Iran J Pediatr       Date:  2016-05-10       Impact factor: 0.364

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.