| Literature DB >> 33474604 |
David Stillström1,2, Raluca-Maria Sandu3, Jacob Freedman4,5.
Abstract
PURPOSE: Evaluate the accuracy of multiple electrode placements in IRE treatment of liver tumours using a stereotactic CT-based navigation system.Entities:
Keywords: Ablation; Hepatocellular carcinoma; IRE; Liver; Navigation; Stereotactic
Mesh:
Year: 2021 PMID: 33474604 PMCID: PMC8172499 DOI: 10.1007/s00270-020-02762-5
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740
Fig. 1View of CAS-ONE work screen (top) and adjustable arm (bottom)
Fig. 2a Angular and lateral error between planned and validated electrode placement b Angular error between two validated electrodes in a pair c Relationship between the centre reference electrode and four validated electrodes
Patient and tumour characteristics
| Male | 58 (81%) |
| Female | 14 (19%) |
| Age (y), mean (±SD) | 65.0 ± 11.0 |
| Colorectal liver metastases | 30 (35.7%) |
| Hepatocellular carcinoma | 36 (42.9%) |
| Cholangiocarcinoma | 4 (4.8%) |
| Livermetastes from CCC | 2 (2.4%) |
| Leiomyosarcoma | 1 (1.2%) |
| Sarcoma | 1 (1.2%) |
| Adrenocortical carcinoma | 7 (8.3%) |
| Pankreas NET | 2 (2.4%) |
| Tumour diameter (mm), median (min-max) | 19 (2–60) |
| Number of electrodes/tumour, median (min-max) | 4 (2–6) |
| Number of electrodes, total | 300 |
| DPL, Patient radiation dose (mGy x cm), mean (±SD) | 1654.9 ± 686.0 |
Fig. 3Distribution of tumours in the liver
Lateral and angular errors
| Median | 4.05 |
| Range | 0.2–13.64 |
| Median | 3.62 |
| Range | 0.18–18.91 |
| Median | 2.25 |
| Range | 0–16.11 |
| Median | 2.48 |
| Range | 0.01–8.68 |
Fig. 4a Lateral error in mm between planned and validated electrode placement b Angular error in degrees between planned and validated electrode placement c Angular error in degrees between planned and validated electrode pairs d Angular error in degrees, comparing planned angel between electrode and reference electrode with validated angel between electrode and reference electrode
Angular errors between validated electrode pairs
| Median | 3.80 |
| Range | 0.28–17.18 |
Fig. 5Angular error in degrees between two validated electrodes in a pair