Literature DB >> 2759649

Definition and notation of hemoglobin oxygen saturation.

W G Zijlstra, B Oeseburg.   

Abstract

Oxygen saturation (SO2) should not be confused with oxyhemoglobin fraction (FHbO2). As automated multiple wavelength photometers may yield both SO2 and FHbO2, care should be taken that only SO2 is used in checking the results of pulse oximeters. Arterial SO2 is the quantity aimed at by pulse oximeters. Although the accuracy of the measurement may be less than that of the analysis of an arterial blood sample in vitro, the measured quantity remains SO2. Using a special symbol, like SpO2, obscures this fact and should be discouraged.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2759649

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

1.  Computationally two-dimensional finite-difference model for hollow-fibre blood-gas exchange devices.

Authors:  D A Baker; J E Holte; S V Patankar
Journal:  Med Biol Eng Comput       Date:  1991-09       Impact factor: 2.602

2.  The effect of dyshemoglobins on pulse oximetry: Part I, Theoretical approach and Part II, Experimental results using an in vitro test system.

Authors:  K J Reynolds; E Palayiwa; J T Moyle; M K Sykes; C E Hahn
Journal:  J Clin Monit       Date:  1993-04
  2 in total