Central Corneal Thickness Reproducibility among Ten Different Instruments.

Luisa Pierro1, Lorenzo Iuliano, Marco Gagliardi, Alessandro Ambrosi, Paolo Rama, Francesco Bandello.   

Abstract

PURPOSE: To assess agreement between one ultrasonic (US) and nine optical instruments for the measurement of central corneal thickness (CCT), and to evaluate intra- and inter-operator reproducibility.
METHODS: In this observational cross-sectional study, two masked operators measured CCT thickness twice in 28 healthy eyes. We used seven spectral-domain optical coherence tomography (SD-OCT) devices, one time-domain OCT, one Scheimpflug camera, and one US-based instrument. Inter- and intra-operator reproducibility was evaluated by intraclass correlation coefficient (ICC), coefficient of variation (CV), and Bland-Altman test analysis. Instrument-to-instrument reproducibility was determined by ANOVA for repeated measurements. We also tested how the devices disagreed regarding systemic bias and random error using a structural equation model.
RESULTS: Mean CCT of all instruments ranged from 536 ± 42 μm to 577 ± 40 μm. An instrument-to-instrument correlation test showed high values among the 10 investigated devices (correlation coefficient range 0.852-0.995; p values <0.0001 in all cases). The highest correlation coefficient values were registered between 3D OCT-2000 Topcon-Spectral OCT/SLO Opko (0.995) and Cirrus HD-OCT Zeiss-RS-3000 Nidek (0.995), whereas the lowest were seen between SS-1000 CASIA and Spectral OCT/SLO Opko (0.852). ICC and CV showed excellent inter- and intra-operator reproducibility for all optic-based devices, except for the US-based device. Bland-Altman analysis demonstrated low mean biases between operators.
CONCLUSIONS: Despite highlighting good intra- and inter-operator reproducibility, we found that a scale bias between instruments might interfere with thorough CCT monitoring. We suggest that optimal monitoring is achieved with the same operator and the same device.

Entities:  

Mesh:

Year:  2016        PMID: 27571223     DOI: 10.1097/OPX.0000000000000974

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  4 in total

1.  Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation.

Authors:  Xi-Fang Zhang; Meng Li; Yan Shi; Xiu-Hua Wan; Huai-Zhou Wang
Journal:  Int J Ophthalmol       Date:  2020-04-18       Impact factor: 1.779

2.  Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology.

Authors:  Erhan Ozyol; Pelin Özyol
Journal:  Indian J Ophthalmol       Date:  2017-06       Impact factor: 1.848

3.  Sirius Scheimpflug-Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis.

Authors:  Yili Jin; Colm McAlinden; Yong Sun; Daizong Wen; Yiran Wang; Jinjin Yu; Ke Feng; Benhao Song; Qinmei Wang; Shihao Chen; Jinhai Huang
Journal:  Eye Vis (Lond)       Date:  2021-02-18

4.  Corneal and Epithelial Thickness Mapping: Comparison of Swept-Source- and Spectral-Domain-Optical Coherence Tomography.

Authors:  Cristina Georgeon; Ilanite Marciano; Roxane Cuyaubère; Otman Sandali; Nacim Bouheraoua; Vincent Borderie
Journal:  J Ophthalmol       Date:  2021-10-05       Impact factor: 1.909

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.