Martina Scharitzer1, Johannes Lenglinger2,3, Wolfgang Schima4, Michael Weber5, Claudia Ringhofer3, Peter Pokieser6.
Abstract
OBJECTIVES: To compare videofluoroscopy that included a tablet test with impedance planimetry (EndoFLIP®) for the evaluation of oesophageal stenosis in patients with dysphagia.Entities:
Keywords: Cineradiography; Deglutition disorders; Impedance planimetry; Oesophageal stenosis; Tablets
Mesh:
Substances:
Year: 2016 PMID: 27553930 PMCID: PMC5334389 DOI: 10.1007/s00330-016-4516-y
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Flowchart of patient enrolment
Fig. 2A 19-year-old patient with dysphagia and achalasia. a Impaction of the tablet at videofluoroscopy for >3 min. b EndoFLIP lumen at 50 ml shows a minimal diameter of 12.4 mm (blue)
Fig. 3Peptic stenosis in a 37-year-old patient with gastroesophageal reflux and dysphagia. a The tablet was impacted at the level of the narrowing (arrows). b EndoFLIP lumen at the oesophagogastric junction is 11.7 mm (blue) and 12.8 mm (dark green) 2 cm more proximal, corresponding to the videofluoroscopically detected stenosis
Placebo tablet transit correlated with oesophageal diameter as measured by impedance planimetry
| Oesophageal tablet transit | Total |
| |||
|---|---|---|---|---|---|
| Impaction or localized delay > 15 s | Short delay ≤ 15 s or normal passage | ||||
| Impedance planimetry diameter | ≥15.1 mm | 6 (46.2 %) | 7 (53.8 %) | 13 (100 %) | 0.035 |
| <15.1 mm | 34 (79.1 %) | 9 (20.9 %) | 43 (100 %) | ||
| Total | 40 (71.4 %) | 16 (28.6 %) | 56 (100 %) | ||
*p values evaluated by Fisher–Halton–Freeman test
Findings of single- and double-contrast radiography correlated to impedance planimetry diameter
| Impedance planimetry diameter | Reader 1 | Reader 2 | ||||||||||||
| Single-contrast | Double-contrast |
| Single-contrast | Double-contrast |
| |||||||||
| Stenosis | No stenosis | Uncertain | Stenosis | No stenosis | Uncertain | 0.577 | Stenosis | No Stenosis | Uncertain | Stenosis | No stenosis | Uncertain | 0.470 | |
| <15.1 mm | 30 | 9 | 4 | 31 | 10 | 2 | 32 | 9 | 2 | 32 | 10 | 1 | ||
| ≥15.1 mm | 9 | 4 | 0 | 8 | 5 | 0 | 9 | 4 | 0 | 8 | 5 | 0 | ||
| Total | 39 | 13 | 4 | 39 | 15 | 2 | 41 | 13 | 2 | 40 | 15 | 1 | ||
*p values evaluated by Fisher-Halton Freeman test
Fig. 4Average oesophageal diameter (mm) for different filling volume of impedance planimetry bag (ml), error bars represent 95 % confidence intervals