| Literature DB >> 27512157 |
Iwona Matraszek-Zuchowska1, Barbara Wozniak1, Andrzej Posyniak1.
Abstract
In accordance with Commission Decision 2002/657/EC, confirmatory methods for the detection of prohibited substances should comply with specific requirements, including the criteria for confirmation. Two strategies: multiple reaction monitoring (MRM) and enhanced product ion (EPI) scanning functions were compared for confirming the anabolic compounds from synthetic stilbenes group in bovine urine samples. In the research, twenty samples fortified at the Recommended Concentration (RC) of 1 µg L-1 with diethylstilbestrol, dienestrol and hexestrol were analyzed by liquid chromatography-tandem mass spectrometry on a QTRAP 5500 instrument. The analytical procedure, validated in accordance with the Commission Decision 2002/657/EC, used in the official control of hormones in Poland was applied. The validation parameters were in agreement with 2002/657/EC performance criteria. The effectiveness of MRM and EPI scanning modes for confirmation purposes was evaluated based on the percentage of the results confirmed. In all urine samples recorded in the MRM mode, the confirmation criteria (retention time, relative intensities between transitions) have been fulfilled. The presence of stilbenes in all urine samples using EPI scan mode was confirmed too as evidenced by a good matching of stilbenes spectra in the samples to the reference spectra with critical match factor above 0.7. The results of the research show that EPI scanning function provides the same effectiveness for confirmation of banned compounds as the mostly used MRM scan mode and can be an additional tool to confirm the doubtful case results in the analysis of hormones residues, even at such low concentration levels.Entities:
Keywords: Confirmatory method; Enhanced product ion; LC–MS/MS; Multiple reaction monitoring; Stilbenes
Year: 2016 PMID: 27512157 PMCID: PMC4961723 DOI: 10.1007/s10337-016-3121-1
Source DB: PubMed Journal: Chromatographia ISSN: 0009-5893 Impact factor: 2.044
Equations calibration curves and results of the method validation for the DES, DIE, HEX in urine samples at a concentration 1 µg L−1
| Compound | DES | DIE | HEX |
|---|---|---|---|
| Calibration curve of standard | |||
| Linear range | 0.1–6 µg L−1 | ||
| Slope ± | 2.30 ± 0.25 | 1.30 ± 0.26 | 1.46 ± 0.22 |
| | −0.32 ± 0.19 | −0.07 ± 0.20 | −0.19 ± 0.17 |
| Linear correlation coefficient ( | 0.9990 | 0.9994 | 0.9985 |
| Coefficient of determination ( | 0.9980 | 0.9988 | 0.9970 |
| Standard error | 0.248 | 0.109 | 0.187 |
| Matrix-matched calibration curve | |||
| Linear range | 0.07–5 µg L−1 | ||
| Slope ± sb | 2.09 ± 0.32 | 1.3 ± 0.4 | 1.29 ± 0.17 |
| | −0.04 ± 0.27 | 0.11 ± 0.11 | 0.04 ± 0.16 |
| Linear correlation coefficient ( | 0.9999 | 0.9999 | 0.9998 |
| Coefficient of determination ( | 0.9998 | 0.9998 | 0.9996 |
| Standard error | 0.062 | 0.019 | 0.047 |
| Apparent recovery (%) ( | 103.5 | 110.5 | 107.0 |
| Repeatability (R.S.D., %) ( | 6.7 | 4.0 | 2.6 |
| Within-lab reproducibility (R.S.D., %) ( | 14.1 | 12.6 | 3.4 |
| Decision limit (CCα, μg L−1) | 0.07 | 0.07 | 0.07 |
| Detection capability (CCβ, μg L−1) | 0.12 | 0.12 | 0.11 |
| Measurement uncertainty at 1μg L−1 ( | 0.30 | 0.32 | 0.22 |
| Matrix effect (%) | −9.1 | 0.0 | −11.6 |
| Samples fulfilling confirmation criteria at 1 µg L−1(%) ( | 100.0 | 100.0 | 95.8 |
n number of samples tested
The confirmation results of DES, DIE and HEX in twenty bovine urine samples spiked at 1 µg L−1, registered in the MRM mode
| Compound | MRM transition ( | Average relative ion intensity ± SD for standards | Average relative ion intensity ± SD for urine spiked | Samples fulfilling confirmation criteria (%) |
|---|---|---|---|---|
| DES |
| 0.866 ± 0.007 | 0.866 ± 0.018 | 100.0 |
| DES-d6 |
| – | – | – |
| DIE | 265.1 > 236.2 | 0.17 ± 0.05 | 0.161 ± 0.002 | 100.0 |
| DIE-d2 |
| – | – | – |
| HEX | 269.0 > 133.0 | 0.487 ± 0.008 | 0.491 ± 0.008 | 100.0 |
| HEX-d4 |
| – | – | – |
aThe most intense transitions in bold used for quantification
Fig. 1The XIC of MRM chromatograms and EPI spectra of DES (CE −43; 10,000 cps), DIE (CE −40; 30,000 cps) and HEX (CE −32; 5000 cps) in standard solution corresponding to 1 μg L−1 in urine sample
Statistics of library search results of Fit, RevFit and Purity for stilbenes obtained in twenty standard solutions with a concentration of DES, DIE and HEX corresponding to 0.1–1 μg L−1 in the urine sample and in urine samples spiked at 1 μg L−1
| Analyte | Fit index | Standard | Matrix | Meanstandard − Meanmatrix (%) | CVstandard − CVmatrix | ||
|---|---|---|---|---|---|---|---|
| Mean | CV (%) | Mean | CV (%) | ||||
| DES | Fit | 0.89 | 8.0 | 0.94 | 4.0 | −5.68 | 3.9 |
| RevFit | 0.94 | 5.0 | 0.87 | 7.7 | 6.63 | −2.7 | |
| Purity | 0.85 | 7.4 | 0.84 | 8.5 | 0.93 | −1.0 | |
| DIE | Fit | 0.97 | 3.5 | 0.98 | 1.7 | −0.77 | 1.8 |
| RevFit | 0.98 | 2.3 | 0.96 | 4.0 | 2.30 | −1.7 | |
| Purity | 0.96 | 4.4 | 0.95 | 4.3 | 0.78 | 0.0 | |
| HEX | Fit | 0.86 | 7.3 | 0.91 | 3.5 | −6.33 | 3.8 |
| RevFit | 0.91 | 5.3 | 0.86 | 9.2 | 4.88 | −3.9 | |
| Purity | 0.83 | 7.4 | 0.82 | 7.5 | 0.70 | −0.1 | |
The confirmation results of DES, DIE and HEX in twenty bovine urine samples spiked at 1 µg L−1, registered in the EPI mode
| Compound | Samples fulfilling confirmation criteria | |||
|---|---|---|---|---|
| Fit ≥ 0.9 | 0.9 > Fit ≥ 0.8 | 0.8 > Fit ≥ 0.7 | (%) | |
| DES | 45.0 | 35.0 | 20.0 | 100.0 |
| 25.0 | 55.0 | 20.0 | ||
| DIE | 95.0 | 5.0 | – | 100.0 |
| 90.0 | 10.0 | – | ||
| HEX | 35.0 | 50.0 | 15.0 | 100.0 |
| – | 80.0 | 20.0 | ||
Fit; RevFit ≥ 0.9, Purity ≥ 0.9—very good matching of spectra to reference. 0.9 > Fit; RevFit ≥ 0.8, 0.9 > Purity ≥ 0.8—good matching of spectra to reference. 0.8 > Fit; RevFit ≥ 0.7, 0.8 > Purity ≥ 0.7—satisfactory matching of spectra to reference