OBJECTIVE: HIV seroconversion biomarkers are being used in cross-sectional studies for HIV incidence estimation. Bio-Rad Geenius HIV-1/2 Supplemental Assay is an immunochromatographic single-use assay that measures antibodies (Ab) against multiple HIV-1/2 antigens. The objective of this study was to determine whether the Geenius assay could additionally be used for recency estimation.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27509247 PMCID: PMC5110377 DOI: 10.1097/QAI.0000000000001146
Source DB: PubMed Journal: J Acquir Immune Defic Syndr ISSN: 1525-4135 Impact factor: 3.731
Estimated Test Properties (and 95% CIs) for Various Specimen Sets and Different Thresholds Cutoff
FIGURE 1.Antigen-specific band intensities. A, Box-and-whisker plots of band intensities, as a function of time since infection (in 6-month intervals), for p31, gp160, and gp41. B, Pairwise scatter plots of band intensities for p31, gp160, and gp41. Excludes treated subjects and SCOPE elite controllers.
FIGURE 2.Incidence assay results over time since infection. A, Box-and-whisker plots of the Geenius result, as a function of time since infection. Results are summarized for each 6-month interval after infection, and results for all specimens drawn more than 2 years after infection are captured in the rightmost box plot. B, Proportion of “recent” results (using a cutoff of 1.5) as a function of time since infection (with a 95% CI), stratified by HIV subtype (B, C, A1, and D). Excludes treated subjects and SCOPE elite controllers.
FIGURE 3.Reproducibility of assay results. Box-and-whisker plots of the 25 repeat measurements for the 3 blinded controls (A–C) and 6–10 repeat measurements for the 4 labeled controls (D–G), for each band used as well as the Geenius index result.
FIGURE 4.Sensitivity of assay results to variations in testing procedures. For each of the 4 specimens (distinguished by marker shape), the individual measurements are shown by markers, and mean measurements connected by dashed lines. The input plasma volume is varied (left plot) or number of drops of buffer is varied (right plot).