Makiko Kondo1, Koji Sudo2, Takako Sano1, Takuya Kawahata3, Ichiro Itoda4, Shinya Iwamuro5, Yukihiro Yoshimura6, Natsuo Tachikawa6, Yoko Kojima3, Haruyo Mori3, Hiroshi Fujiwara7, Naoki Hasegawa7, Shingo Kato2.
Abstract
Accurate diagnosis of earlier HIV infection is essential for treatment and prevention. Currently, confirmation tests of HIV infection in Japan are performed using Western blot (WB), but WB has several limitations including low sensitivity and cross-reactivity between HIV-1 and HIV-2 antibodies. To address these problems, a new HIV testing algorithm and a more reliable confirmation and HIV-1/2 differentiation assay are required. The Bio-Rad Geenius HIV-1/2 Confirmatory Assay (Geenius) has recently been approved and recommended for use in the revised guidelines for diagnosis of HIV infection by the Center for Disease Control and Prevention (USA). We made comprehensive comparison of the performance of Geenius and the Bio-Rad NEW LAV BLOT 1 and 2 (NLB 1 and 2) which are WB kits for HIV-1 and HIV-2, respectively, to examine if Geenius is a suitable alternative to these WB assays which are now being used in HIV testing in Japan. A total of 166 HIV-1 positive samples (146 from patients with established HIV-1 infection and 20 from patients with acute infection), five HIV-1 seroconversion panels containing 21 samples and 30 HIV-2 positive samples were used. In addition, a total of 140 HIV negative samples containing 10 false-positives on screening tests were examined. The sensitivity of Geenius and NLB 1 for HIV-1 positive samples was 99.3% and 98.6%, respectively. Geenius provided more positive results in the samples from acute infections and detected positivity 0 to 32 days earlier in seroconversion panels than NLB 1. NLB 2 gave positive results in 12.3% of HIV-1 positive samples. The sensitivity of both Geenius and NLB 2 for HIV-2 positive samples was 100%. The specificity of Geenius, NLB 1 and NLB 2 was 98.5%, 81.5% and 90.0%, respectively. Geenius is an attractive alternative to WB for confirmation and differentiation of HIV-1 and HIV-2 infections. The adaptation of Geenius to the HIV testing algorithm may be advantageous for rapid diagnosis and the reduction of testing costs.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30379808 PMCID: PMC6209130 DOI: 10.1371/journal.pone.0198924
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Comparison of Geenius with NLB 1 and 2 results for established and acute HIV-1 infection samples.
| WB | Geenius | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| HIV-1 positive | HIV-1 indeterminate | HIV-2 positive | HIV-2 indeterminate | HIV positive untypable | HIV negative | Total | |||
| Established HIV-1 Infection | NLB 1 | Positive | 143 | 0 | 0 | 0 | 1 | 0 | 144 |
| Indeterminate | 1 | 1 | 0 | 0 | 0 | 0 | 2 | ||
| Negative | 0 | 0 | 0 | 0 | 0 | 0 | 0 | ||
| Total | 144 | 1 | 0 | 0 | 1 | 0 | 146 | ||
| NLB 2 | Positive | 18 | 0 | 0 | 0 | 0 | 0 | 18 | |
| Indeterminate | 122 | 1 | 0 | 0 | 1 | 0 | 124 | ||
| Negative | 4 | 0 | 0 | 0 | 0 | 0 | 4 | ||
| Total | 144 | 1 | 0 | 0 | 1 | 0 | 146 | ||
| Acute HIV-1 | NLB 1 | Positive | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Indeterminate | 7 | 6 | 0 | 0 | 0 | 3 | 16 | ||
| Negative | 0 | 0 | 0 | 0 | 0 | 4 | 4 | ||
| Total | 7 | 6 | 0 | 0 | 0 | 7 | 20 | ||
| NLB 2 | Positive | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Indeterminate | 5 | 3 | 0 | 0 | 0 | 0 | 8 | ||
| Negative | 2 | 3 | 0 | 0 | 0 | 7 | 12 | ||
| Total | 7 | 6 | 0 | 0 | 0 | 7 | 20 | ||
aFourth-generation enzyme immunoassay positive, NLB 1 positive and NAT positive at the time of initial diagnosis.
bFourth-generation enzyme immunoassay positive, NLB 1 negative and NAT positive at the time of sample collection
cOn ART at the time of sample collection.
Comparison of Geenius with NLB 1 and 2 results for HIV-1.seroconversion panels.
| Patient | Sample | Days | Geenius | NLB 1 | NLB 2 | |
|---|---|---|---|---|---|---|
| HIV-1 | HIV-2 | |||||
| A | 1 | 0 | Neg | Neg | Neg | Neg |
| 2 | 9 | Pos (gp160, p24, gp41) | Neg | Ind (gp160, p68, p55, p24) | Ind (p26) | |
| 3 | 16 | Pos (gp160, p24, gp41) | Neg | Ind (gp160, p68, p55, p40, p31, p24, p18) | Ind (p26) | |
| 4 | 36 | Pos (gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| 5 | 42 | Pos (gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| 6 | 65 | Pos (p31, gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| 7 | 107 | Pos (p31, gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| B | 1 | 0 | Neg | Neg | Neg | Neg |
| 2 | 7 | Neg | Neg | Neg | Neg | |
| 3 | 40 | Pos (gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| 4 | 47 | Pos (p31, gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| 5 | 85 | Pos (p31, gP160, p24, gp41) | Neg | Pos | Ind (p26) | |
| C | 1 | 0 | Neg | Neg | Neg | Neg |
| 2 | 7 | Pos (gp160, gp41) | Neg | Ind (gp160, p24) | Neg | |
| 3 | 39 | Pos (gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| 4 | 126 | Pos (gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| D | 1 | 0 | Neg | Neg | Neg | Neg |
| 2 | 7 | Ind (gp41) | Neg | Ind (p52, p40, p24, p18) | Ind (p26) | |
| 3 | 33 | Pos (gp160, p24, gp41) | Neg | Pos | Ind (p26) | |
| E | 1 | 0 | Pos (gp160, p24, gp41) | Neg | Pos | Ind (p26, p16) |
| 2 | 8 | Pos (gp160, p24, gp41) | Neg | Pos | Ind (p26, p16) | |
aResults are shown by Pos (positive), Ind (indeterminate) or Neg (negative).
bTime from collecting the first sample.
cWhen a result was positive, reactive antigens are shown in parenthesis.
dWhen a result was indeterminate, reactive antigens are shown in parenthesis.
Comparison of Geenius with NLB 1 and 2 results for HIV-2 panel samples.
| Geenius | |||||||
|---|---|---|---|---|---|---|---|
| HIV-1 positive | HIV-2 positive | HIV-2 positive with HIV-1 cross-reactivity | HIV positive untypable | HIV negative | Total | ||
| NLB 1 | Positive | 0 | 0 | 2 | 0 | 0 | 2 |
| Indeterminate | 0 | 10 | 16 | 2 | 0 | 28 | |
| Negative | 0 | 0 | 0 | 0 | 0 | 0 | |
| Total | 0 | 10 | 18 | 2 | 0 | 30 | |
| NLB 2 | Positive | 0 | 10 | 18 | 2 | 0 | 30 |
| Indeterminate | 0 | 0 | 0 | 0 | 0 | 0 | |
| Negative | 0 | 0 | 0 | 0 | 0 | 0 | |
| Total | 0 | 10 | 18 | 2 | 0 | 30 | |
Comparison of Geenius with NLB 1 and 2 results for negative samples by fourth-generation immunoassay (n = 130).
| Geenius | ||||||||
|---|---|---|---|---|---|---|---|---|
| HIV-1 positive | HIV-1 indeterminate | HIV-2 positive | HIV-2 indeterminate | HIV positive untypable | HIV negative | Total | ||
| NLB 1 | Positive | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Indeterminate | 0 | 0 | 0 | 0 | 0 | 24 | 24 | |
| Negative | 0 | 1 | 0 | 1 | 0 | 104 | 106 | |
| Total | 0 | 1 | 0 | 1 | 0 | 128 | 130 | |
| NLB 2 | Positive | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Indeterminate | 0 | 0 | 0 | 1 | 0 | 12 | 13 | |
| Negative | 0 | 1 | 0 | 0 | 0 | 116 | 117 | |
| Total | 0 | 1 | 0 | 1 | 0 | 128 | 130 | |
Comparison of Geenius with NLB 1 and 2 results for HIV-1 Combo positive but NAT negative samples (n = 10)a.
| Sample. | Geeniusa | NLB 1 | NLB 2 | |
|---|---|---|---|---|
| HIV-1 | HIV-2 | |||
| 1 | Negative | Negative | Indeterminate (p18) | Negative |
| 2 | Negative | Negative | Indeterminate (p18) | Indeterminate (p26, p16) |
| 3 | Negative | Negative | Negative | Negative |
| 4 | Negative | Negative | Negative | Negative |
| 5 | Negative | Negative | Indeterminate (p24, p18) | Negative |
| 6 | Indeterminate (p31) | Negative | Indeterminate (p18) | Indeterminate (p26) |
| 7 | Negative | Negative | Negative | Negative |
| 8 | Negative | Negative | Negative | Indeterminate (p16) |
| 9 | Indeterminate (p31) | Negative | Negative | Indeterminate (p16) |
| 10 | Negative | Negative | Negative | Indeterminate (p16) |