Sebastian Bidhult1,2, George Kantasis1,3, Anthony H Aletras1,3, Håkan Arheden1, Einar Heiberg1,2, Erik Hedström4,5.
Abstract
BACKGROUND: Determination of the relaxation time constants T1 and T2 with quantitative magnetic resonance imaging is increasingly used for both research and clinical practice. Recently, groups have been formed within the Society of Cardiovascular Magnetic Resonance to address issues with relaxometry. However, so far they have avoided specific recommendations on methodology due to lack of consensus and current evolving research. Standardised widely available software may simplify this process. The purpose of the current study was to develop and validate vendor-independent T1 and T2 mapping modules and implement those in the versatile and widespread software Segment, freely available for research and FDA approved for clinical applications.Entities:
Keywords: Mapping; Quantitative magnetic resonance imaging; T1; T2
Mesh:
Year: 2016 PMID: 27501697 PMCID: PMC4977731 DOI: 10.1186/s12880-016-0148-6
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Supported sequences and signal fitting models
| Supported signal models | ||
|---|---|---|
| Supported sequences | 3-parameter fit model | 2-parameter fit model |
| T1 spin-echo IR (magnitude images) | S(t) = |A (1 − B exp(−t/T1))| | S(t) = |A (1 − 2 exp(−t/T1))| |
| T1 PSIR | S(t) = A (1 − B exp(−t/T1)) | S(t) = A (1 − 2 exp(−t/T1)) |
| T1 saturation recovery balanced SSFP | S(t) = A (1 − B exp(−t/T1)) | S(t) = A (1 − exp(−t/T1)) |
| T1 MOLLI/T1 Look-Locker correction (magnitude images) | S(t) = |A (1 − B exp(−t/T1*))|; T1 = T1* (B − 1) | n/a |
| T1 MOLLI/T1 Look-Locker correction (PSIR images) | S(t) = A (1 − B exp(−t/T1*)); T1 = T1* (B − 1) | n/a |
| T2 spin echo (multi-echo and single-echo) | S(t) = A exp(−t/T2) + B; B > 0 | S(t) = A exp(−t/T2) |
| T2-prepared balanced SSFP | S(t) = A exp(−t/T2) + B; B > 0 | S(t) = A exp(−t/T2) |
IR inversion recovery, PSIR phase sensitive inversion recovery, MOLLI modified Look-Locker inversion recovery, SSFP steady-state free precession
Fig. 1The image shows the graphical user interface of the T1 mapping module (the T2 mapping module shares the design). An example of a cardiac pre-contrast MOLLI T1 map at 1.5 T in a healthy volunteer is shown. Both pixel-wise mapping, for the whole image and ROI-based, and ROI-based global mean mapping are available. The residuals of the curve fit for both T1 and T2 mapping can be visualised, indicating areas within the image that have a higher deviation from the curve and thus less accurate T1 and T2 values
Typical MRI sequence parameters
| TE (ms) | TR (ms)/delay between contrast preparations pulses | FA (°) | FOV (mm) | Matrix | Preparation pulse delays (ms) | iPAT/SENSE factor | Receiver BW (kHz) | ACQ time (hh:mm:ss) | |
|---|---|---|---|---|---|---|---|---|---|
| T1 spin-echo IR (magnitude images) | 5.8 | 10,000/10,000 | 90 | 241 × 241 | 128 × 128 | [21, 60, 100, 200, 300, 500, 660, 900, 1050, 1300, 1600, 2000, 2250, 2500, 3000, 3500, 4300] | off | 64 | 06:45:20 |
| T1 PSIR | 1.11 | 2.4/40,000 | 35 | 360 × 270 | 192 × 144 | [150, 300, 400, 500, 700, 900, 1100, 1300, 1800, 2000, 2500, 3000, 3700, 4300, 5000, 6300] + 1 reference image without IR-preparation | 2 | 208.32 | 00:11:20 |
| T1 saturation recovery balanced SSFP | 1.11 | 2.4/40,000 | 35 | 360 × 270 | 192 × 144 | [150, 300, 400, 500, 700, 900, 1100, 1300, 1800, 2000, 2500, 3000, 3700, 4300, 5000, 5600, 6300, 8000] + 1 reference image without SR preparation | 2 | 208.32 | 00:12:40 |
| T1 MOLLI/T1 Look-Locker correction (magnitude and PSIR images) | 1.1 | 2.4/7920 | 35 | 360 × 270 | 192 × 144 | [130, 210, 1130, 1210, 2130, 2210, 3130, 4130] | 2 | 208.32 | 00:00:11 |
| T2 spin echo (single-echo) | [6, 12, 20, 30, 40, 50, 70, 90, 120, 140, 180, 300, 400, 600, 1000] | 10,000/10,000 | 90 | 241 × 241 | 128 × 128 | n/a | off | 64 | 05:20:00 |
| T2 spin echo (multi-echo) | [9, 18, 27, 36, 45, 54, 63, 72, 81, 90, 99, 108, 117, 126, 135, 144] | 1500/n/a | 90 | 160 × 160 | 256 × 256 | n/a | off | 62.5 | 00:06:24 |
| T2-prepared balanced SSFP | 1.11 | 2.4/20,000 | 35 | 360 × 270 | 192 × 144 | T2p range = 25 − 200; ΔT2p = 5; | 2 | 208.32 | 00:12:20 |
IR inversion recovery, PSIR phase sensitive inversion recovery, MOLLI modified Look-Locker inversion recovery, SSFP steady-state free precession
Computational times for pixel-wise mapping in the complete image and in a selected ROI, respectively
| Pixel-wise (complete image) | Pixel-wise (ROI only) | |
|---|---|---|
| T1 spin-echo IR (magnitude images) | 2.5 s (3 parameters, 17 images, 128 × 128 images, 13 435 refined pixels) | 0.03 s (3 parameters, 17 images, 200 pixels, 152 refined pixels) |
| T1 PSIR | 11.7 s (3 parameters, 17 images, 192 × 144 matrix, 26 597 refined pixels) | 0.15 s (3 parameters, 17 images, 200 pixels, 200 refined pixels) |
| T1 saturation recovery balanced SSFP | 11.2 s (3 parameters, 15 images, 192 × 144 matrix, 26 594 refined pixels) | 0.09 s (3 parameters, 15 images, 202 pixels, 202 refined pixels) |
| T1 MOLLI/T1 Look-Locker correction (magnitude and PSIR images) | 6.4 s (3 parameters, 9 images, 192 × 144 matrix, 25 128 refined pixels) | 0.05 s (3 parameters, 9 images, 202 pixels, 198 refined pixels) |
| T2 spin echo (single-echo and multi-echo) | 1.12 s (2 parameters, 15 images, 128 × 128 matrix, 9 175 refined pixels) | 0.04 s (2 parameters, 15 images, 207 pixels, 160 refined pixels) |
| T2-prepared balanced SSFP | 11.7 s (3 parameters, 37 images, 192 × 144 matrix, 16 322 refined pixels) | 0.14 s (3 parameters, 37 images, 207 pixels, 184 refined pixels) |
The ROI-based global mean fitting takes less than 2 s for all sequences and is not listed. Performance was tested at a 2.4 GHz, 8 GB RAM, SSD HDD standard laptop running the MS Windows 7 64-bit operating system
IR inversion recovery, PSIR phase sensitive inversion recovery, MOLLI modified Look-Locker inversion recovery, SSFP steady-state free precession
Fig. 2Modified Bland-Altman analyses of the phantom validation data at 1.5 T (left column) and 3 T (right column). All curve fits were performed using the 3-parameter fit. The dotted and dashed lines represent bias and 95 % limits of agreement. A generally high agreement was found. The T1 outliers dependent on low T2 (<60 ms) found using the MOLLI sequence at 1.5 T are encircled in red. bSSFP = balanced steady-state free precession; MOLLI = modified Look-Locker inversion recovery; SR = saturation recovery; IR = inversion recovery
Fig. 3Example curve fits for T1 and T2 at 1.5 T in two phantoms. The solid lines represent estimated relaxation curves. bSSFP = balanced steady-state free precession; MOLLI = modified Look-Locker inversion recovery; SR = saturation recovery
Fig. 4Application of the proposed T1 and T2 mapping modules on images acquired in humans and ex vivo. Data from 3 major vendors were used for reconstruction of T1 and T2 maps. Arrows point to findings or structures. a Siemens Aera 1.5 T myocarditis (T1, MOLLI); b Siemens Aera 1.5 T healthy foetal brain (T1, inversion-recovery bSSFP); c Siemens Aera 1.5 T healthy foetal blood oxygenation in descending aorta (T2, T2-prepared bSSFP); d Siemens Prism 3 T ex vivo healthy placenta (T1, inversion-recovery bSSFP); e Philips 1.5 T healthy kidney (T1, MOLLI); f GE 750w 3 T healthy knee cartilage (T2, multi-echo spin-echo)