Detection of spine structures with Bioimpedance Probe (BIP) Needle in clinical lumbar punctures.

Sanna Halonen1,2, Kari Annala3, Juho Kari4, Samuli Jokinen5, Aki Lumme6, Kai Kronström4, Arvi Yli-Hankala5,7.   

Abstract

Lumbar puncture is a relatively safe procedure, but some serious, even fatal, complications can occur. Needle guidance can increase puncture accuracy, decrease the number of attempts, and make the procedure easier. We tested the feasibility of a bioimpedance-based tissue-sensing technology for needle guidance in clinical use. The Bioimpedance Probe (BIP) Needle has a removable BIP stylet enabling measurement of bioimpedance spectra during the procedure. The BIP Needle is connected to a measurement device that uses tissue-classification software, and the device provides audiovisual feedback when it detects cerebrospinal fluid (CSF). We performed spinal anesthesia with the BIP Needle in 45 patients. The device performance and needle tip location were verified by an experienced anesthesiologist confirming CSF leakage. The device detected CSF in all cases (sensitivity of 100 %). Six cases with false detections lowered the specificity to 81 %, but in practice, most of these were easy to differentiate from true detections because their duration was short and they occurred during backward movement of the needle. The epidural spectrum differentiated as fatty tissue from surrounding tissues, but the ligamentum flavum was not clearly detectable in the data. The BIP Needle is a reliable tool for detecting CSF in lumbar puncture. It can make the puncture procedure smoother, as repeated CSF flow tests are avoided. The correct needle tip location is immediately detected, thus unnecessary needle movements close to spinal nerves are prevented. Physicians could benefit from the information provided by the BIP Needle, especially in patients with obesity or anatomic alterations.

Entities:  

Keywords:  Bioimpedance; Cerebrospinal fluid; Epidural; Monitoring; Needle guidance; Spinal anesthesia

Mesh:

Year:  2016        PMID: 27492427     DOI: 10.1007/s10877-016-9915-8

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  18 in total

1.  Measurement of bio-impedance with a smart needle to confirm percutaneous kidney access.

Authors:  D J Hernandez; V A Sinkov; W W Roberts; M E Allaf; A Patriciu; T W Jarrett; L R Kavoussi; D Stoianovici
Journal:  J Urol       Date:  2001-10       Impact factor: 7.450

2.  Linear and nonlinear electrode polarization and biological materials.

Authors:  H P Schwan
Journal:  Ann Biomed Eng       Date:  1992       Impact factor: 3.934

3.  The polarization impedance of common electrode metals operated at low current density.

Authors:  T Ragheb; L A Geddes
Journal:  Ann Biomed Eng       Date:  1991       Impact factor: 3.934

4.  Impedance-based tissue discrimination for needle guidance.

Authors:  Håvard Kalvøy; Lars Frich; Sverre Grimnes; Orjan G Martinsen; Per Kristian Hol; Audun Stubhaug
Journal:  Physiol Meas       Date:  2009-01-09       Impact factor: 2.833

5.  Detection of intraneural needle-placement with multiple frequency bioimpedance monitoring: a novel method.

Authors:  Håvard Kalvøy; Axel R Sauter
Journal:  J Clin Monit Comput       Date:  2015-04-23       Impact factor: 2.502

Review 6.  The dielectric properties of biological tissues: I. Literature survey.

Authors:  C Gabriel; S Gabriel; E Corthout
Journal:  Phys Med Biol       Date:  1996-11       Impact factor: 3.609

Review 7.  Failed epidural: causes and management.

Authors:  J Hermanides; M W Hollmann; M F Stevens; P Lirk
Journal:  Br J Anaesth       Date:  2012-06-26       Impact factor: 9.166

8.  Ultrasound imaging facilitates spinal anesthesia in adults with difficult surface anatomic landmarks.

Authors:  Ki Jinn Chin; Anahi Perlas; Vincent Chan; Danielle Brown-Shreves; Arkadiy Koshkin; Vandana Vaishnav
Journal:  Anesthesiology       Date:  2011-07       Impact factor: 7.892

9.  The electrical conductivity of human cerebrospinal fluid at body temperature.

Authors:  S B Baumann; D R Wozny; S K Kelly; F M Meno
Journal:  IEEE Trans Biomed Eng       Date:  1997-03       Impact factor: 4.538

10.  Serious complications associated with spinal and epidural anaesthesia in Finland from 2000 to 2009.

Authors:  M T Pitkänen; U Aromaa; D A Cozanitis; J G Förster
Journal:  Acta Anaesthesiol Scand       Date:  2013-01-11       Impact factor: 2.105

View more
  5 in total

Review 1.  Advances in tissue state recognition in spinal surgery: a review.

Authors:  Hao Qu; Yu Zhao
Journal:  Front Med       Date:  2021-05-15       Impact factor: 4.592

Review 2.  Journal of clinical monitoring and computing 2017 end of year summary: anesthesia.

Authors:  Jan F A Hendrickx; Andre M De Wolf; Stanley Skinner
Journal:  J Clin Monit Comput       Date:  2018-02-24       Impact factor: 2.502

3.  Bioimpedance spinal needle provides high success and low complication rate in lumbar punctures of pediatric patients with acute lymphoblastic leukemia.

Authors:  Satu Långström; Anu Huurre; Juho Kari; Olli Lohi; Harri Sievänen; Sauli Palmu
Journal:  Sci Rep       Date:  2022-04-26       Impact factor: 4.996

4.  Real-time detection of cerebrospinal fluid with bioimpedance needle in paediatric lumbar puncture.

Authors:  Harri Sievänen; Juho Kari; Sanna Halonen; Timo Elomaa; Outi Tammela; Hanna Soukka; Vesa Eskola
Journal:  Clin Physiol Funct Imaging       Date:  2021-03-20       Impact factor: 2.273

5.  Tissue Recognition Based on Electrical Impedance Classified by Support Vector Machine in Spinal Operation Area.

Authors:  Bingrong Chen; Yongwang Shi; Jiahao Li; Jiliang Zhai; Liang Liu; Wenyong Liu; Lei Hu; Yu Zhao
Journal:  Orthop Surg       Date:  2022-08-01       Impact factor: 2.279

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.