Measurement of bio-impedance with a smart needle to confirm percutaneous kidney access.

D J Hernandez1, V A Sinkov, W W Roberts, M E Allaf, A Patriciu, T W Jarrett, L R Kavoussi, D Stoianovici.   

Abstract

PURPOSE: The traditional method of percutaneous renal access requires freehand needle placement guided by C-arm fluoroscopy, ultrasonography, or computerized tomography. This approach provides limited objective means for verifying successful access. We developed an impedance based percutaneous Smart Needle system and successfully used it to confirm collecting system access in ex vivo porcine kidneys.
MATERIALS AND METHODS: The Smart Needle consists of a modified 18 gauge percutaneous access needle with the inner stylet electrically insulated from the outer sheath. Impedance is measured between the exposed stylet tip and sheath using Model 4275 LCR meter (Hewlett-Packard, Sunnyvale, California). An ex vivo porcine kidney was distended by continuous gravity infusion of 100 cm. water saline from a catheter passed through the parenchyma into the collecting system. The Smart Needle was gradually inserted into the kidney to measure depth precisely using a robotic needle placement system, while impedance was measured continuously.
RESULTS: The Smart Needle was inserted 4 times in each of 4 kidneys. When the needle penetrated the distended collecting system in 11 of 16 attempts, a characteristic sharp drop in resistivity was noted from 1.9 to 1.1 ohm m. Entry into the collecting system was confirmed by removing the stylet and observing fluid flow from the sheath. This characteristic impedance change was observed only at successful entry into the collecting system.
CONCLUSIONS: A characteristic sharp drop in impedance signifies successful entry into the collecting system. The Smart Needle system may prove useful for percutaneous kidney access.

Mesh:

Year:  2001        PMID: 11547124

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450


  3 in total

1.  Design of Bioimpedance Spectroscopy Instrument With Compensation Techniques for Soft Tissue Characterization.

Authors:  Robert E Dodde; Grant H Kruger; Albert J Shih
Journal:  J Med Device       Date:  2015-06       Impact factor: 0.582

2.  Detection of spine structures with Bioimpedance Probe (BIP) Needle in clinical lumbar punctures.

Authors:  Sanna Halonen; Kari Annala; Juho Kari; Samuli Jokinen; Aki Lumme; Kai Kronström; Arvi Yli-Hankala
Journal:  J Clin Monit Comput       Date:  2016-08-04       Impact factor: 2.502

3.  Biopsy Needle Integrated with Electrical Impedance Sensing Microelectrode Array towards Real-time Needle Guidance and Tissue Discrimination.

Authors:  Jaeho Park; Won-Mook Choi; Kyuyoung Kim; Won-Il Jeong; Joon-Beom Seo; Inkyu Park
Journal:  Sci Rep       Date:  2018-01-10       Impact factor: 4.379