A novel mouthpiece prevents bite injuries caused by intraoperative transcranial electric motor-evoked potential monitoring.

Kyoko Oshita1, Noboru Saeki2, Takayasu Kubo3, Hitoshi Abekura3, Nobuhiro Tanaka4, Masashi Kawamoto2.   

Abstract

PURPOSE: Intraoperative transcranial motor-evoked potential monitoring causes contraction of the masseter muscles, which may cause injuries to the oral cavity and damage to the orotracheal tube. We developed a mouthpiece made from vinyl-silicone impression material to prevent these injuries. The purpose of this study was to examine its efficacy and safety.
METHODS: Twenty-two patients undergoing spinal surgery under transcranial motor-evoked potential monitoring were fitted with bespoke vinyl-silicone mouthpieces by dentists before surgery. On induction of general anesthesia and orotracheal intubation, the mouthpiece was attached to the upper and lower dental arches. A lateral cervical X-ray was taken at the end of surgery to examine the condition of the orotracheal tube. The incidence of endotracheal tube deformation was compared with an historic control group of 20 patients in whom a conventional gauze bite block had been previously used before induction of the mouthpiece. The oral cavity was examined by a dentist the day before surgery and 3 days postoperatively, and intraoral injuries were recorded.
RESULTS: No endotracheal tube deformation was found in 22 patients fitted with the new mouthpiece. The incidence of tube deformation (none of 22 patients, 0 %) was significantly lower than in those who had been fitted with the gauze bite block (9 of 20 patients, 45.0 %; p < 0.001). Application of the mouthpiece resulted in no tongue or tooth injuries.
CONCLUSION: A novel mouthpiece reduced the incidence of damage to the endotracheal tube caused by intraoperative transcranial motor-evoked potential monitoring.

Entities:  

Keywords:  Bite injuries; Mouthpiece; Tc-MEP monitoring

Mesh:

Year:  2016        PMID: 27468733     DOI: 10.1007/s00540-016-2220-3

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  13 in total

1.  Prevention of airway injury during spine surgery: rethinking bite blocks.

Authors:  Stacie G Deiner; Irene P Osborn
Journal:  J Neurosurg Anesthesiol       Date:  2009-01       Impact factor: 3.956

2.  Protection of tongue from injuries during transcranial motor-evoked potential monitoring.

Authors:  Mohamed Mahmoud; James Spaeth; Senthilkumar Sadhasivam
Journal:  Paediatr Anaesth       Date:  2008-09       Impact factor: 2.556

3.  Tube-in-tube emergency airway management after a bitten endotracheal tube caused by repetitive transcranial electrical stimulation during spinal cord surgery.

Authors:  Andreas Duma; Klaus Novak; Wolfgang Schramm
Journal:  Anesthesiology       Date:  2009-11       Impact factor: 7.892

4.  [Adverse events during transcranial muscle evoked potential monitoring].

Authors:  Yoshie Kuwahara; Shima Taguchi; Masahiko Kuroda; Masashi Kawamoto
Journal:  Masui       Date:  2011-06

Review 5.  Safety of intraoperative transcranial electrical stimulation motor evoked potential monitoring.

Authors:  David B MacDonald
Journal:  J Clin Neurophysiol       Date:  2002-10       Impact factor: 2.177

6.  A report of two cases of lip and tongue bite injury associated with transcranial motor evoked potentials.

Authors:  Scott F Davis; Philip Kalarickal; Ted Strickland
Journal:  Am J Electroneurodiagnostic Technol       Date:  2010-12

7.  Transcranial electric motor evoked potential monitoring during spine surgery: is it safe?

Authors:  Daniel M Schwartz; Anthony K Sestokas; John P Dormans; Alexander R Vaccaro; Alan S Hilibrand; John M Flynn; P Mark Li; Suken A Shah; William Welch; Denis S Drummond; Todd J Albert
Journal:  Spine (Phila Pa 1976)       Date:  2011-06       Impact factor: 3.468

8.  The incidence of bite injuries associated with transcranial motor-evoked potential monitoring.

Authors:  Arvydas Tamkus; Kent Rice
Journal:  Anesth Analg       Date:  2012-04-20       Impact factor: 5.108

9.  The efficacy of motor evoked potentials in fixed sagittal imbalance deformity correction surgery.

Authors:  Jeremy A Lieberman; Russ Lyon; John Feiner; Serena S Hu; Sigurd H Berven
Journal:  Spine (Phila Pa 1976)       Date:  2008-06-01       Impact factor: 3.468

10.  An evidenced-based scoring index to determine the periodontal prognosis on molars.

Authors:  Preston D Miller; Mark L McEntire; Nicole M Marlow; Robert G Gellin
Journal:  J Periodontol       Date:  2013-05-31       Impact factor: 6.993

View more
  2 in total

Review 1.  Transcranial electric stimulation motor evoked potentials for cervical spine intraoperative monitoring complications: systematic review and illustrative case of cardiac arrest.

Authors:  Francisco Revilla-Pacheco; Shoko Watanabe; Joel Rodríguez-Reyes; Claudia Sánchez-Torres; Paul Shkurovich-Bialik; Tenoch Herrada-Pineda; Pamela Rodríguez-Salgado; Juvenal Franco-Granillo; Martín Calderón-Juárez
Journal:  Eur Spine J       Date:  2022-07-06       Impact factor: 2.721

2.  Bite injuries caused by transcranial electrical stimulation motor-evoked potentials' monitoring: incidence, associated factors, and clinical course.

Authors:  Sachiko Yata; Mitsuru Ida; Hiroko Shimotsuji; Yosuke Nakagawa; Nobuhiro Ueda; Tsunenori Takatani; Hideki Shigematsu; Yasushi Motoyama; Hiroyuki Nakase; Tadaaki Kirita; Masahiko Kawaguchi
Journal:  J Anesth       Date:  2018-10-05       Impact factor: 2.078

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.