Jennifer R McKinney1, Marshall S Sussman2, Rahim Moineddin3, Afsaneh Amirabadi1, Tammy Rayner1, Andrea S Doria1,2.
Abstract
OBJECTIVES: To evaluate the accuracy of magnetic resonance imaging measurements of cartilage tissue-mimicking phantoms and to determine a combination of magnetic resonance imaging parameters to optimize accuracy while minimizing scan time.Entities:
Mesh:
Year: 2016 PMID: 27464298 PMCID: PMC4946528 DOI: 10.6061/clinics/2016(07)09
Source DB: PubMed Journal: Clinics (Sao Paulo) ISSN: 1807-5932 Impact factor: 2.365
Figure 1Layout of cartilage-mimicking phantoms (1-4) on a planar surface in ambient air (a) and embedded in a saline-Magnevist® bath (b).
Magnetic resonance imaging sequence parameters and statistically significant differences (p-values) between the mean direct measurements and the mean MRI measurements (shown in bold).
| Pulse sequence | T1 SE | GRE MPGR | FSPGR | ||||||
|---|---|---|---|---|---|---|---|---|---|
| MRI parameters | Matrix (pixels) | 320x320 | 320x320 | 512x512 | 320x320 | 512x512 | 512x512 | 512x512 | 1024x1024 |
| Slice thickness (mm) | 1.5 | 4 | 1.5 | 4 | 1.5 | 1.5 | 0.5 | 1.5 | |
| NEX | 2 | 2 | 2 | 2 | 2 | 1 | 1 | 1 | |
| Scan time (min) | 6:11 | 6:11 | 9:48 | 3:40 | 5:49 | 3:36 | 3:38 | 10:33 | |
| TE (ms) | 12 | 12 | 12 | 15 | 15 | Min full | Min full | Min full | |
| TR (ms) | 568 | 568 | 568 | 450 | 450 | N/A | N/A | N/A | |
| Bandwidth (kHz) | 31.2 | 31.2 | 31.2 | 15.6 | 15.6 | 31.2 | 31.2 | 31.3 | |
| FOV (mm) | 150 | 150 | 150 | 150 | 150 | 160 | 160 | 160 | |
| Dimension | 2D | 2D | 2D | 2D | 2D | 3D | 3D | 3D | |
| Differences ( | Phantom 1(Min width = 3.0 mm, m.d.e.m. = 9.94 mm) | 0.45 | 0.48 | 0.48 | 0.35 ( | 0.61 | 0.40 | 0.39 ( | 0.41 ( |
| Phantom 2(Min width = 2.5 mm, m.d.e.m. = 8.69 mm) | 0.25( | 0.25( | 0.31( | 0.05( | 0.42( | 0.19 ( | 0.32( | 0.23( | |
| Phantom 3(Min width = 1.5 mm, m.d.e.m. = 6.38 mm) | 0.042 ( | 0.33( | 0.06( | -0.017( | 0.008( | 0.09 ( | -0.050 ( | 0.07 ( | |
| Phantom 4(Min width = 0.5 mm, m.d.e.m. = 7.94 mm) | 0.20 ( | 3.12( | 3.17( | 0.11( | 3.19( | 0.16 ( | 0.41 ( | 0.26( | |
FOV field of view, FSPGR fast spoiled gradient-recalled echo, GRE MPGR multiplanar gradient-recalled echo, Min full minimum achievable TE with full echo acquisition, m.d.e.m. mean direct edge measurement, Min width minimum width, N/A not applicable, NEX number of excitations, T1 SE T1-weighted spin-echo, TE echo time, TR repetition time.
Figure 2Coronal MR images generated by various sequences and parameters.(a) T1-weighted spin-echo (SE), 320 x 320 pixels, 1.5-mm slice thickness; (b) T1 SE, 320 x 320 pixels, 4-mm slice thickness; (c) T1 SE, 512 x 512 pixels, 1.5-mm slice thickness; (d) Multiplanar gradient-recalled echo (GRE MPGR), 320 x 320 pixels, 1.5-mm slice thickness; (e) GRE MPGR, 512 x 512 pixels, 1.5-mm slice thickness; (f) Fast spoiled gradient-recalled echo (FSPGR), 512 x 512 pixels, 0.5-mm slice thickness; (g) FSPGR, 1024 x 1024 pixels, 1.5-mm slice thickness; (h) FSPGR, 512 x 512 pixels, 1.5-mm slice thickness.
Figure 3Signal-to-noise ratio (SNR) calculation using MR images of phantoms 1-4 embedded in a saline bath. SNR was calculated as SNR=0.65*(signal intensity/standard deviation), employing a correction factor of 0.65 for background noise. The grey ovals represent 1-mm2 regions of interest (ROIs).
Signal-to-noise ratios for the magnetic resonance imaging sequences employed in this study.
| T1 SE | GRE MPGR | FSPGR | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Matrix (pixels) | 320x320 | 320x320 | 512x512 | 320x320 | 512x512 | 512x512 | 512x512 | 1024x1024 | |
| Slice thickness (mm) | 1.5 | 4 | 1.5 | 4 | 1.5 | 1.5 | 0.5 | 1.5 | |
| SNR | Phantom 1 | 44.4 | 66.5 | 25.8 | 35.0 | 11.7 | 39.2 | 21.7 | 15.4 |
| Phantom 2 | 38.2 | 79.8 | 31.0 | 44.9 | 8.6 | 47.6 | 22.4 | 18.0 | |
| Phantom 3 | 27.0 | 90.7 | 34.1 | 19.2 | 10.0 | 31.9 | 21.1 | 12.3 | |
| Phantom 4 | 33.6 | 84.9 | 33.5 | 23.7 | 9.7 | 36.8 | 20.7 | 14.5 | |
FSPGR fast spoiled gradient-recalled echo, GRE MPGR multiplanar gradient-recalled echo, SNR signal-to-noise ratio, T1 SE T1-weighted spin-echo.
Summary of factors influencing the appropriateness of different MRI sequences for measuring maturing cartilage using phantom 1.
| Pulse sequence | T1 SE | GRE MPGR | FSPGR | |||||
|---|---|---|---|---|---|---|---|---|
| Matrix (pixels) | 320x320 | 320x320 | 512x512 | 320x320 | 512x512 | 512x512 | 512x512 | 1024x1024 |
| Slice thickness (mm) | 1.5 | 4 | 1.5 | 4 | 1.5 | 1.5 | 0.5 | 1.5 |
| FOV (mm) | 150 | 150 | 150 | 150 | 150 | 160 | 160 | 160 |
| Scan time (min) | 6:11 | 6:11 | 9:48 | 3:40 | 5:49 | 3:36 | 3:38 | 10:33 |
| Spatial resolution (mm) | 0.47 | 0.47 | 0.29 | 0.47 | 0.29 | 0.31 | 0.31 | 0.16 |
| 0.049 | 0.020 | 0.030 | 0.111 | 0.007 | 0.048 | 0.10 | 0.051 | |
| SNR | 44.4 | 66.5 | 25.8 | 35.0 | 11.7 | 39.2 | 21.7 | 15.4 |
FOV field of view, FSPGR fast spoiled gradient-recalled echo, GRE MPGR multiplanar gradient-recalled echo, SNR signal-to-noise ratio, T1 SE T1-weighted spin-echo.