Literature DB >> 27462383

Evaluation of a Prototype Hybrid Vacuum Pump to Provide Vacuum-Assisted Suspension for Above-Knee Prostheses.

Matthew J Major1, Ryan Caldwell2, Stefania Fatone3.   

Abstract

Vacuum-assisted suspension (VAS) of prosthetic sockets utilizes a pump to evacuate air from between the prosthetic liner and socket, and are available as mechanical or electric systems. This technical note describes a hybrid pump that benefits from the advantages of mechanical and electric systems, and evaluates a prototype as proof-of-concept. Cyclical bench testing of the hybrid pump mechanical system was performed using a materials testing system to assess the relationship between compression cycles and vacuum pressure. Phase 1 in vivo testing of the hybrid pump was performed by an able-bodied individual using prosthesis simulator boots walking on a treadmill, and phase 2 involved an above-knee prosthesis user walking with the hybrid pump and a commercial electric pump for comparison. Bench testing of 300 compression cycles produced a maximum vacuum of 24 in-Hg. In vivo testing demonstrated that the hybrid pump continued to pull vacuum during walking, and as opposed to the commercial electric pump, did not require reactivation of the electric system during phase 2 testing. The novelty of the hybrid pump is that while the electric system provides rapid, initial vacuum suspension, the mechanical system provides continuous air evacuation while walking to maintain suspension without reactivation of the electric system, thereby allowing battery power to be reserved for monitoring vacuum levels.

Keywords:  prosthesis; socket; suspension; vacuum

Year:  2015        PMID: 27462383      PMCID: PMC4940611          DOI: 10.1115/1.4030507

Source DB:  PubMed          Journal:  J Med Device        ISSN: 1932-6181            Impact factor:   0.582


  13 in total

1.  A comparison of trans-tibial amputee suction and vacuum socket conditions.

Authors:  W J Board; G M Street; C Caspers
Journal:  Prosthet Orthot Int       Date:  2001-12       Impact factor: 1.895

2.  Walking in a vacuum-assisted socket shifts the stump fluid balance.

Authors:  J Goswami; R Lynn; G Street; M Harlander
Journal:  Prosthet Orthot Int       Date:  2003-08       Impact factor: 1.895

3.  Residual limb wounds or ulcers heal in transtibial amputees using an active suction socket system. A randomized controlled study.

Authors:  M Traballesi; A S Delussu; A Fusco; M Iosa; T Averna; R Pellegrini; S Brunelli
Journal:  Eur J Phys Rehabil Med       Date:  2012-05-28       Impact factor: 2.874

4.  Interface pressures during ambulation using suction and vacuum-assisted prosthetic sockets.

Authors:  Tracy L Beil; Glenn M Street; Steven J Covey
Journal:  J Rehabil Res Dev       Date:  2002 Nov-Dec

5.  Transfemoral interfaces with vacuum assisted suspension comparison of gait, balance, and subjective analysis: ischial containment versus brimless.

Authors:  Jason T Kahle; M Jason Highsmith
Journal:  Gait Posture       Date:  2014-05-09       Impact factor: 2.840

6.  Trans-tibial prosthesization in large area of residual limb wound: is it possible? A case report.

Authors:  Marco Traballesi; Tiziano Averna; Anna Sofia Delussu; Stefano Brunelli
Journal:  Disabil Rehabil Assist Technol       Date:  2009-09

7.  Transfemoral sockets with vacuum-assisted suspension comparison of hip kinematics, socket position, contact pressure, and preference: ischial containment versus brimless.

Authors:  Jason T Kahle; M Jason Highsmith
Journal:  J Rehabil Res Dev       Date:  2013

8.  Amputee Independent Prosthesis Properties--a new model for description and measurement.

Authors:  Matthew J Major; Martin Twiste; Laurence P J Kenney; David Howard
Journal:  J Biomech       Date:  2011-08-09       Impact factor: 2.712

9.  Methods for characterization of mechanical and electrical prosthetic vacuum pumps.

Authors:  Oluseeni Komolafe; Sean Wood; Ryan Caldwell; Andrew Hansen; Stefania Fatone
Journal:  J Rehabil Res Dev       Date:  2013

10.  A case series featuring extremely short below-knee stumps.

Authors:  Jose Andre Carvalho; Mauricio Dias Mongon; William Dias Belangero; Bruno Livani
Journal:  Prosthet Orthot Int       Date:  2011-12-14       Impact factor: 1.895

View more
  2 in total

1.  Effectiveness of elevated vacuum and suction prosthetic suspension systems in managing daily residual limb fluid volume change in people with transtibial amputation.

Authors:  Robert T Youngblood; Jacob T Brzostowski; Brian J Hafner; Joseph M Czerniecki; Katheryn J Allyn; Richard L Foster; Joan E Sanders
Journal:  Prosthet Orthot Int       Date:  2020-03-18       Impact factor: 1.895

2.  Real time monitoring of transtibial elevated vacuum prostheses: A case series on socket air pressure.

Authors:  Katherine R Schoepp; Jonathon S Schofield; David Home; Michael R Dawson; Edmond Lou; McNiel Keri; Paul D Marasco; Jacqueline S Hebert
Journal:  PLoS One       Date:  2018-10-22       Impact factor: 3.240

  2 in total