Literature DB >> 27453976

Isolation and Quantification of Polyamide Cyclic Oligomers in Kitchen Utensils and Their Migration into Various Food Simulants.

Yutaka Abe1, Motoh Mutsuga1, Hiroyuki Ohno2, Yoko Kawamura1, Hiroshi Akiyama1.   

Abstract

Small amounts of cyclic monomers and oligomers are present in polyamide (PA)-based kitchen utensils. In this study, we isolated eight PA-based cyclic monomers and oligomers from kitchen utensils made from PA6 (a polymer of ε-caprolactam) and PA66 (a polymer of 1,6-diaminohexane and adipic acid). Their structures were identified using high-resolution mass spectrometry and 1H- and 13C-nuclear magnetic resonance spectroscopy, and their residual levels in PA-based kitchen utensils and degree of migration into food simulants were quantified by high-performance liquid chromatography/mass spectrometry using purchased PA6 monomer and isolated PA66 monomers, and isolated PA6 and PA66 oligomers as calibration standards. Their total residual levels among 23 PA-based kitchen utensils made from PA6, PA66, and copolymers of PA6 and PA66 (PA6/66) ranged from 7.8 to 20 mg/g. Using water, 20% ethanol, and olive oil as food simulants, the total migration levels of the PA monomers and oligomers ranged from 0.66 to 100 μg/cm2 under most examined conditions. However, the total migration levels of the PA66 monomer and oligomers from PA66 and PA6/66 kitchen utensils into 20% ethanol at 95°C were very high (1,700 and 2,200 μg/cm2, respectively) due to swelling by high-temperature ethanol.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27453976      PMCID: PMC4959713          DOI: 10.1371/journal.pone.0159547

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Plastic materials are widely used in kitchen utensils such as dishes, cups, spoons, forks, measuring cups, and bottles. Among the various plastics used, polyamide (PA) materials are typically used for turners or ladles due to their high heat and oil resistance. There are many types of PA-based materials, such as PA6, PA66, and PA6/66 (copolymers of PA6 and PA66), all of which are commonly used in kitchen utensils. PA6 is produced by the ring-opening polymerization of ε-caprolactam (CPL), while PA66 is produced by the polycondensation of 1,6-diaminohexane (1,6-hexamethylenediamine, HMDA) and adipic acid (AA) (Fig 1). Small amounts of PA cyclic monomers and oligomers are produced as by-products during these processes and remain in the PA-based materials. Typical structures of PA cyclic monomers and oligomers are shown in Fig 2. For succinctness, we use the term ‘PA66 monomer’ for the cyclic condensation product of HMDA or AA.
Fig 1

Condensation reactions of PA6 and PA66.

Fig 2

Structures of PA monomers and oligomers.

The numbers for CPL, PA6 cyclic dimer, and PA66 cyclic monomer represent the positions shown in Table 1.

Structures of PA monomers and oligomers.

The numbers for CPL, PA6 cyclic dimer, and PA66 cyclic monomer represent the positions shown in Table 1.
Table 1

1H- and 13C-NMR data of PA cyclic monomers and oligomers.

Monomers and oligomersNucleusChemical shift data (ppm)
1234567
PA6 monomer(CPL)1H7.60 (s)3.23 (2H, m)1.64–1.70a (m)1.80 (2H, m)1.64–1.70a2.47-
(m)(2H, t)
13C-4430.831.824.537.6182
PA6 dimer1H7.59 (s)3.27 (4H, t)1.28 (4H, m)1.63 (4H, m)1.482.2-
(4H, m)(4H, t)
13C-4030.126.726.437.1176.5
PA6 trimer1H7.60 (s)3.21 (6H, t)1.34 (6H, m)1.64 (6H, m)1.532.22-
(6H, m)(6H, t)
13C-40.329.62726.337176.4
PA6 tetramer1H7.59 (s)3.20 (8H, t)1.33 (8H, m)1.641.522.21-
(8H, m)(8H, m)(8H, t)
13C-40.430.127.426.737.2176.3
PA6 pentamer1H7.60 (s)3.18 (10H, t)1.33 (10H, m)1.631.522.19-
(10H, m)(10H, m)(10H, t)
13C-40.530.327.626.837.4176.1
PA6 hexamer1H7.60 (s)3.20 (12H, t)1.34 (12H, m)1.641.532.21-
(12H, m)(12H, m)(12H, t)
13C-40.430.127.426.737.2176.3
PA66 monomer1H7.58a3.22 (4H, t)1.53 (4H, s)1.321.632.19-
(4H, s)(4H, s)(4H, s)
13C-39.929.127.427.337.5176.5
PA66 dimer1H7.59 (s)3.15 (8H, t)1.47 (8H, t)1.311.62.18-
(8H, s)(8H, s)(8H, t)
13C-40.430.427.526.737.2176
PA66 trimer1H7.58a3.17 (12H, t)1.50 (12H, s)1.341.622.21-
(12H, s)(12H, s)(12H, t)
13C-40.530.427.626.737.1175.9

aOverlapped with each other.

The data were obtained in CD3OD:CDCl3 (1:1) solution.

aOverlapped with each other. The data were obtained in CD3OD:CDCl3 (1:1) solution. It has been reported that such monomers and oligomers migrate into food and food simulants from food packaging [1-4]. In these reports, PA cyclic monomers and oligomers were analyzed using high-performance liquid chromatography (HPLC) equipped with an ultraviolet (UV) detector at 210 nm [2-6]. Due to a lack of commercially available standards, the quantification of these cyclic oligomers was performed using relative response factors (RRFs) with reference to CPL. Unfortunately, the RRFs reported in previous publications are not consistent. Notably, Heimrich et al. [7] derived cyclic oligomer RRFs (UV 210 nm) relative to CPL using synthesized cyclic oligomers as standards, and their quantification was carried out using an HPLC equipped with a chemiluminescence nitrogen detector (CLND) and UV detector. However, CLND is not commonly available in many laboratories, and the selectivity and sensitivity of a UV detector may be insufficient for the determination of these oligomers. In this study, we isolated eight PA cyclic monomer and oligomers from PA6- and PA66-based kitchen utensils and identified their structures using high-resolution mass spectroscopy (HRMS) and 1H- and 13C-nuclear magnetic resonance (NMR) spectroscopy. Furthermore, their residual levels in PA6-, PA66-, and PA6/66-based kitchen utensils were quantified by liquid chromatography/mass spectrometry (LC/MS) using purchased CPL and isolated PA66 monomer and PA6 and PA66 oligomers as calibration standards. In addition, migration tests were performed using various food simulants, e.g., water was used for common foods, 20% ethanol for alcoholic beverages, and olive oil for fat and fatty foods.

Materials and Methods

Samples and reagents

Twenty-three kitchen utensils composed of PA (turners, ladles, a sesame grinder, a cake server, and a cake scraper) were purchased at retail stores, such as super markets, one-coin shops, variety store, in Setagaya (Tokyo) and Nagoya (Aichi), Japan. The sample information, such as color, labeling (material types, maximum usage temperature, and producing country), and their material types identified using LC/MS chromatograms of their extracts as mentioned in the ‘Results and Discussion’ section were shown in Table 2. The names of the manufacturing or sales companies of these samples were shown in S1 Table.
Table 2

PA kitchen utensil sample types.

SampleColorLabelMaterial*
MaterialMaximum usage temperature (°C)Producing country
Turner 1BlackNylon 66180Japan66
Turner 2GrayNylon210Japan66
Turner 3Dark grayNylon180China66
Turner 4BlackNylon 66210Japan66
Turner 5BlackNylon 66220Japan66
Turner 6BlackNylon 66, Glass fiber 15%220Portugal66
Turner 7BlackNylon180Japan66
Turner 8BlackNylon200Japan66
Turner 9OrangeNylon200China66
Turner 10Yellow-greenNylon180China6/66
Ladle 1GrayNylon180China6
Ladle 2Dark grayNylon180China66
Ladle 3BlackNylon 66210Japan66
Ladle 4GrayNylon 66210Japan66
Ladle 5BlackNylon210China6/66
Ladle 6BlackNylon, Glass fiber 15%240China66
Ladle 7BlackNylon 66200China66
Ladle 8BlackNylon200Japan66
Ladle 9OrangeNylon180Japan66
Ladle 10OrangeNylon 66200Japan66
Sesame grinderWhiteNylon90Vietnam6
Cake serverGrayNylon180China6/66
Cake scraperGrayNylon180China6/66

*Determined using LC/MS.

*Determined using LC/MS. ε-Caprolactam (99%) was purchased from Tokyo Chemical Industry Co., Ltd. (Tokyo, Japan). Methanol for HPLC analysis grade (> 99.8%) was obtained from Kanto Chemical Co., Inc. (Tokyo, Japan), and formic acid for HPLC analysis grade (ca. 99%) was obtained from Wako Pure Chemical Industries, Ltd. (Osaka, Japan). Ultrapure water was purified with the Milli-Q Gradient system equipped with a Millipak 0.22 μm filter (Millipore, Billerica, MA, USA).

LC/MS conditions

The LC/MS was an Acquity Series LC/MS system (Waters, Milford, MA, USA) equipped with an Acquity BEH C18 (100 mm × 2.1 mm, 1.7 μm, Waters). The LC/MS conditions were as follows: column temperature of 40°C; flow rate of 0.25 mL/min; mobile phase of methanol/0.1% formic acid (5:95) (1 min) running with a linear gradient to methanol/0.1% formic acid (95:5) (5 min) and methanol/0.1% formic acid (95:5) (10 min); MS electrospray capillary voltage of 3 kV; drying gas temperature of 400°C; flow rate of 600 L/hour (N2), cone gas flow rate of 50 L/hour (N2); electrospray ionization (positive) mode, SCAN (m/z 50–1000) or single ion recording (SIR) modes; and cone voltage of 30 V (SCAN mode). The SIR mode cone voltage and quantitative ions are shown in Table 3.
Table 3

Retention times and MS (SIR mode) conditions for quantification.

PeakCyclic monomers and oligomersRetention time (min)Quantitative ion (m/z)Cone voltage (V)
Peak APA6 dimer4.2227.330
Peak BPA6 monomer (CPL)4.4113.630
Peak CPA6 trimer4.9340.430
Peak DPA6 tetramer5.3453.330
Peak EPA6 pentamer5.5566.440
Peak FPA6 hexamer5.7701.840
Peak GPA66 monomer4.5227.330
Peak HPA66 dimer5.6453.330
Peak IPA66 trimer5.9701.840

Isolation and purification of PA cyclic monomers and oligomers

Three grams of the PA-based samples (PA6: ladle 1; PA66: turner 1) were powdered with liquid nitrogen using a frost shattering machine (JFC-300, Japan Analytical Industry Co., Ltd., Tokyo, Japan). Monomers and oligomers were extracted from the samples with methanol (30 mL) at 40°C overnight and the residue was subsequently removed by filtration with a cotton plug. The filtrates and washings were combined and evaporated using a rotary evaporator (EYERA, Tokyo Rikakikai Co., Ltd., Tokyo, Japan), and the resulting residue was dissolved in 10% methanol. The solution was subjected to chromatography using an OASIS HLB cartridge (6 g, Waters), which was pretreated with methanol and water. Subsequently, each 50 mL of 10, 20, 30, 40, and 50% methanol was passed through the cartridge and collected separately. The eluted and crudely purified monomers and oligomers in each fraction were determined using LC/MS (SCAN mode), and the solvent was subsequently evaporated. Next, using an OASIS HLB cartridge (200 mg, Waters), each 50 mL of 10, 20, 30, 40, and 50% methanol was passed through the cartridge and every 10 mL was collected, and isolated monomers and oligomers in each fraction were determined in the same way. These steps were repeated for each fraction until pure PA cyclic monomers and oligomers were obtained. CPL and PA6 could not be separated using this cartridge. To separate them, the hydrolysis of CPL and subsequent separation were performed. To hydrolyze the PA6 monomer to 6-aminohexanoic acid (AHA) and separate it from the PA6 dimer, 10 mL of 6 M HCl was added to the residue containing a mixture of PA6 monomer and PA6 dimer (50 mg). The mixture was subsequently heated at 110°C for 6 h and then neutralized with 10% NaOH. This solution was subjected to the OASIS HLB cartridge (200 mg) again, and the PA6 dimer was separated from AHA via elution with 10%–30% methanol.

Characterization of PA cyclic monomer and oligomers

NMR spectra were recorded on a JEOL Delta 800 spectrometer (800 MHz for 1H, 200 MHz for 13C) (JEOL Ltd., Tokyo, Japan) using a 1:1 (v/v) mixture of chloroform-d/methanol-d4 as the solvent. To elucidate the structures of the monomer and oligomers, 1D (1H and 13C) and 2D H-H Correlation Spectroscopy (COSY), 1H-Detected Multiple Quantum Coherence (HMQC), and 1H-detected Multi-Bond Heteronuclear Multiple Quantum Coherence (HMBC) NMR analyses were performed. Analysis by HRMS was performed by injecting the extracts from ladle 1 and turner 1 into a LC/time-of-flight (TOF)-MS (Aquity LC system, Micromass LCT Premier XE, Waters) system. The LC/TOF-MS conditions were the same as those used for LC/MS. The purity of each isolated monomer and oligomer was calculated as their peak purities on the total ion chromatograms (TIC) of LC/MS (SCAN mode ranging from m/z 50–1,000). The peak purities were calculated by dividing the peak area of each isolated monomer or oligomer by the total peak area on the TIC.

Quantification of residual levels

For the quantification of the residual levels of PA cyclic monomers and oligomers in kitchen utensils, and to efficiently extract the oligomers from the samples, suitable extraction solvents were selected among water, 50% methanol, methanol, 20% ethanol, ethanol, 2-propanol, and acetonitrile, which are all commonly used extraction solvents. Methanol was found to be the most efficient for the extraction of PA monomers and oligomers. A powdered sample of kitchen utensils (100 mg) using a frost shattering machine was suspended in methanol (5 mL) and heated overnight at 40°C. After filtration using filter paper (No. 5C), the residue was washed with methanol. The filtrates and washings were combined, and the final volume was adjusted to 10 mL with methanol. The test solution was prepared by diluting this solution 20 times with 50% methanol containing 0.1% formic acid. When the concentration of the monomers or oligomers in the test solution exceeded the calibration range, the test solution was diluted 400- or 2,000-fold with the same solution.

Quantification of migration levels

To quantify the migration levels of PA cyclic monomers and oligomers, migration tests were performed using various food simulants with reference to the Japanese Food Sanitation Law9). In the Law, water and 4% acetic acid were used as the food simulants for common foods based on the pH, 20% ethanol for alcoholic beverages and heptane for fatty foods. Among these simulants, we selected water for common foods, 20% ethanol for alcoholic beverages, and heptane for fatty foods. However, since heptane was not used in the EU as a food simulant, olive oil was used instead. The migration tests were performed using the total immersing method as follows. The samples were cut into 1 × 5 cm pieces and then placed into screw-capped glass tubes (100 mm i.d. × 300 mm) containing 20 mL of the simulant (water, 20% ethanol, or olive oil), which was preheated to 60°C. The samples were placed in a thermostatically controlled water bath (60°C and 95°C for all simulants) or an autoclave (Autoclave SS-245, Tomy Seiko Co., Ltd., Tokyo, Japan) (121°C for olive oil only) for 30 min. After the migration, the samples were removed from the simulants as soon as possible. The obtained water migration solutions were added to methanol and formic acid to reach a level of 50% methanol containing 0.1% formic acid, and subsequently subjected to LC/MS. The 20% ethanol migration solutions were subjected to LC/MS without any pretreatment. The olive oil migration solutions were not suitable for LC/MS analysis because of their insolubility in aqueous solutions. Therefore, to analyze these solutions using LC/MS, it was necessary to replace the olive oil with a suitable solvent. This was achieved using the methods reported by Begley et al. [2] and Soto et al. [3]. Briefly, the obtained olive oil solution (5 mL) was added to 50 mL of hexane. It was then transferred into a separatory funnel and 50 mL methanol (saturated with hexane) was added. After shaking the funnel vigorously, the methanol layer was collected and evaporated. The residue was dissolved in 5 mL of 50% methanol containing 0.1% formic acid and was then subjected to LC/MS. When the concentration of the monomers or oligomers in the test solution exceeded the calibration range, the test solution was diluted with the relevant solvent.

Calibration curve and limit of quantification

To construct calibration curves for the quantification of the residual and migration levels of the PA cyclic monomers and oligomers, standard solutions were prepared as follows. A mixture of the standard stock solution of PA monomers and oligomers was prepared using purchased CPL and isolated PA66 monomer and PA6 and PA66 oligomers with a concentration of 1,000 μg/mL in methanol. The mixtures of standard solutions were prepared with concentrations ranging from 0.001−0.100 μg/mL in 50% methanol containing 0.1% formic acid or 20% ethanol. The prepared mixtures of standard solutions were analyzed using LC/MS, and standard calibration curves were constructed using external standardization. For each PA monomer and oligomer, the limit of quantification (LOQ) was defined as the signal-to-noise ratio of peaks with intensities of 10/1.

Results and Discussion

Isolation and characterization of PA cyclic monomer and oligomers for discrimination of polyamide materials

To identify the polyamide materials, extracts from the samples were subjected to LC/MS (SCAN mode). As shown in Fig 3A–3C, three types of chromatograms were obtained. Among the twenty-three PA-based kitchen utensils, two, seventeen, and four samples exhibited the type of chromatograms shown in Fig 3A, 3B, and 3C, respectively. The compounds corresponding to peaks A, B, C, D, E, and F were detected in the chromatogram shown in Fig 3A, those corresponding to peaks G, H, and I were detected in the chromatogram shown in Fig 3B, and those corresponding to all peaks were detected in the chromatogram shown in Fig 3C.
Fig 3

LC/MS total ion chromatograms (SCAN mode) of the extracts.

(A) Ladle 1 (PA6), (B) turner 1 (PA66), and (C) cake scraper (PA6/66). LC/MS: liquid chromatography/mass spectrometry.

LC/MS total ion chromatograms (SCAN mode) of the extracts.

(A) Ladle 1 (PA6), (B) turner 1 (PA66), and (C) cake scraper (PA6/66). LC/MS: liquid chromatography/mass spectrometry. These compounds were predicted to be PA monomers and oligomers, and by comparing the retention time and mass spectrum of the purchased CPL standard, the compound corresponding to peak B was identified as CPL. To identify the compounds corresponding to peaks A, C, D, E, F, G, H, and I, the compounds were fractionated and purified using a hydrophilic–lipophilic–balanced (HLB) cartridge. The compounds corresponding to peaks C (25.3 mg), D (20.3 mg), E (18.7 mg), and F (5.5 mg) were purified from the ladle 1 extracts (Fig 3A), and those corresponding to peaks G (16.4 mg), H (20.2 mg), and I (4.2 mg) were purified from the turner 1 extracts (Fig 3B). As shown in Fig 4A, the compounds corresponding to peak A and CPL were not separable using this cartridge. However, as shown in Fig 4B, the majority of CPL was hydrolyzed to AHA under the described conditions (see the Material and Methods section), and following subsequent purification using the cartridge, the compound corresponding to peak A was successfully obtained (8.0 mg).
Fig 4

Hydrolysis of CPL to 6-aminohexanoic acid.

(A) Before and (B) after hydrolysis.

Hydrolysis of CPL to 6-aminohexanoic acid.

(A) Before and (B) after hydrolysis. To determine the structures of the purified compounds, HRMS and NMR analyses were performed. As shown in Table 4, HRMS revealed that the masses of the compounds corresponding to peaks A, C, D, E, and F were consistent with the calculated theoretical masses of the cyclic dimer, trimer, tetramer, pentamer, and hexamer of CPL, respectively. Furthermore, the masses of the compounds corresponding to peaks G, H, and I were, respectively, consistent with the calculated theoretical masses of the cyclic monomer, dimer, and trimer of HMDA and AA. The 1D and 2D NMR spectra strongly supported the estimated structures (Table 1). These HRMS results showed that the compounds corresponding to peaks A, C, D, E, and F were PA6 cyclic dimer, trimer, tetramer, hexamer, and pentamer, and those corresponding to peaks G, H, and I were PA66 cyclic monomer, dimer, and trimer, respectively (Fig 5). These results also showed that two samples containing only PA6 monomer and oligomers were identified as PA6, seventeen samples containing only PA66 monomer and oligomers were identified as PA66, and four samples containing both PA6 and PA66 monomers and oligomers were identified as PA6/66 on their chromatograms (Fig 3A–3C, Table 2).
Table 4

High-resolution mass data of PA6 cyclic oligomers and PA66 cyclic monomer and oligomers.

PeakCyclic monomer and oligomersFormulaMass (m/z)
CalculatedObserved
Peak APA6 dimerC12H23N2O2227.1760227.1779
Peak CPA6 trimerC18H34N3O3340.2600340.2606
Peak DPA6 tetramerC24H45N4O4453.3441453.3427
Peak EPA6 pentamerC30H56N5O5566.4281566.4277
Peak FPA6 hexamerC36H67N6O6679.5122679.5108
Peak GPA66 monomerC12H23N2O2227.1760227.1767
Peak HPA66 dimerC24H45N4O4453.3441453.3438
Peak IPA66 trimerC36H67N6O6679.5122679.5123
Fig 5

LC/MS total ion chromatograms (SCAN mode) of isolated PA cyclic monomers and oligomers.

As shown in Fig 5, some small peaks corresponding to other oligomers were detected on the TIC of isolated monomers and oligomers, but those corresponding to impurities were not detected. From these results, it was assumed that isolated monomers or oligomers might be slightly mixed in with other oligomers, but not mixed with impurities. These assumptions were confirmed using NMR analysis in that only the signals corresponding to CPL, HMDA, or AA were detected in each NMR spectrum. Thus, the purities of each isolated monomer and oligomer were estimated using the TIC of LC/MS analysis as shown in the Materials and Methods section, and the purities of the PA6 cyclic dimer, trimer, tetramer, pentamer, and hexamer, and PA66 cyclic monomer, dimer, and trimer were estimated to be 84%, 95%, 80%, 93%, 94%, 91%, 84%, and 77%, respectively. Heimrich et al. [7] reported the identification of four PA linear oligomers and three PA cyclic oligomers (PA6 cyclic dimer, PA66 cyclic monomer, and PA66 cyclic dimer). However, as far as we know, our report is the first that presents the complete identification of the PA6 cyclic trimer, tetramer, and pentamer, and PA66 cyclic trimer. Although we monitored the linear oligomers using LC/MS in the SCAN mode, the linear oligomers could not be detected in this study. This was because the sum of the residual levels of linear oligomers would be less than 2% of those of cyclic oligomers [8].

Quantification of residual levels of PA cyclic monomers and oligomers in kitchen utensils

The method performance for the quantification of residual levels of PA monomers and oligomers in kitchen utensils was assessed. Good linearity was achieved over the concentration range of 0.001−0.050 μg/mL for the standard solutions of PA monomers and oligomers in 50% methanol containing 0.1% formic acid: correlation coefficients (R) were > 0.990. The LOQ for the residual levels of PA monomers and oligomers in PA kitchen utensils was estimated to be 0.002 mg/g. The repeatability and reproducibility of the determinations were assessed using the relative standard deviation (RSD) values of intraday (n = 5, RSDr) and interday precision (n = 2 × 5 days, RSDi) by determining the residual levels of PA monomers and oligomers in the ladle 1 and turner 1 samples. As shown in Table 5, their RSDr values ranged from 4.4% to 6.7%, and RSDi values ranged from 4.8% to 8.6%. These values showed that good linearity, repeatability, and reproducibility were obtained for the quantification of all PA monomers and oligomers.
Table 5

Repeatability and reproducibility of the determination method for PA cyclic monomers and oligomers.

Cyclic monomers and oligomersRSDrRSDi
PA6 monomer (CPL)6.15.2
PA6 dimer4.77.8
PA6 trimer4.46.0
PA6 tetramer6.75.5
PA6 pentamer6.24.8
PA6 hexamer6.37.4
PA66 monomer4.77.1
PA66 dimer6.28.2
PA66 trimer5.78.6

RSDr (%): Repeatability (n = 5).

RSDi (%): Reproducibility (n = 2 × 5 day).

RSDr (%): Repeatability (n = 5). RSDi (%): Reproducibility (n = 2 × 5 day). As shown in Table 6, the residual levels of PA monomers and oligomers in the 23 kitchen utensil samples were quantified.
Table 6

Residual levels of PA cyclic monomers and oligomers in PA kitchen utensils.

ArticlesMaterial typePA6 cyclic monomer and oligomersPA66 cyclic monomer and oligomersTotalPA6:PA66
MonomerDimerTrimerTetramerPentamerHexamerSubtotalMonomerDimerTrimerSubtotal
Ladle 164.82.82.84.62.81.4190.110.0880.0630.261999:1
Sesame grinder63.70.81.65.02.81.315NDNDNDND15100:0
Turner 1660.0260.0130.0190.0400.0110.0080.115.45.33.214141:99
Turner 2660.003NDNDNDNDND0.0034.74.53.913130:100*
Turner 366NDNDNDNDNDNDND4.44.43.713130:100*
Turner 4660.0560.002NDNDNDND0.0574.84.93.914140:100*
Turner 5660.0860.0060.0330.0600.0160.0210.223.03.02.98.99.12:98
Turner 6660.004NDNDNDNDND0.0042.83.23.29.19.10:100*
Turner 7660.0250.0050.0120.0290.0100.0090.0914.74.84.214141:99
Turner 8660.0620.0250.0410.0880.0260.0240.276.76.76.220201:99
Turner 966NDNDNDNDNDNDND5.25.33.414140:100*
Ladle 266NDNDNDNDNDNDND5.35.54.315150:100*
Ladle 3660.0290.0050.0150.0300.0110.0070.0962.72.62.47.77.81:99
Ladle 4660.011NDNDNDNDND0.0113.53.41.98.98.91:99
Ladle 6660.120.0100.0250.0500.0180.0160.243.64.44.212122:98
Ladle 7660.0230.0050.0110.0270.0090.0090.0835.25.34.615151:99
Ladle 8660.0330.0060.0130.0310.0100.0090.102.82.82.37.98.01:99
Ladle 9660.059NDNDNDNDND0.0596.05.94.616170:100*
Ladle 10660.064NDNDNDNDND0.0644.74.33.412121:99
Ladle 56/661.20.510.601.30.870.585.01.62.12.46.11145:55
Turner 106/660.630.430.270.800.480.232.83.53.73.2101322:78
Cake server6/661.60.320.521.40.820.455.13.73.93.8111731:69
Cake scraper6/660.740.110.280.580.230.102.14.24.13.7121415:85

Each value (mg/g) is the mean of two trials.

ND: < 0.002 mg/g

*0:100: PA6 ratios below 0.4.

Each value (mg/g) is the mean of two trials. ND: < 0.002 mg/g *0:100: PA6 ratios below 0.4. For the two PA6 samples, the PA6 monomer and oligomers were detected in both samples. Their residual levels ranged from 0.8 to 5.0 mg/g, and their subtotals were 19 and 15 mg/g. The PA66 monomer and oligomers were detected only in ladle 1 with the residual levels ranging from 0.063 to 0.11 mg/g, and their subtotal was 0.26 mg/g. For the seventeen PA66 samples, the PA66 monomer and oligomers were detected in all samples. Their residual levels ranged from 1.9 to 6.7 mg/g, and their subtotals ranged from 7.7 to 20 mg/g. The PA6 monomer and oligomers were detected in fourteen samples. Their residual levels ranged from 0.003 to 0.12 mg/g and their subtotals ranged from 0.003 to 0.27 mg/g. For the four PA6/66 samples, the PA6 and PA66 monomers and oligomers were detected in all samples. The residual levels of the PA6 monomer and oligomers ranged from 0.10 to 1.6 mg/g, and their subtotals ranged from 2.1 to 5.1 mg/g. The residual levels of the PA66 monomer and oligomers ranged from 1.6 to 4.2 mg/g, and their subtotals ranged from 6.1 to 12 mg/g. Comparing the subtotals of the residual levels of the PA6 monomer and oligomers with those of the PA66 monomer and oligomers, it is evident that the levels of the PA66 monomer and oligomers were higher. The total residual levels of PA monomers and oligomers in all samples ranged from 7.8 to 20 mg/g, and there appear to be no significant differences between the material types. Being common to all samples, the residual levels of the PA6 monomer and tetramer appeared to be greater than those of the other oligomers. Those of the PA66 monomer and dimer appeared to be similar while those of the trimer were slightly lower.

Quantification of migration levels of PA cyclic monomers and oligomers into simulants

Since the quantitative capability for the standard solutions of PA monomers and oligomers in 50% methanol containing 0.1% formic acid was already assessed in the previous section, those in 20% ethanol were assessed in this section. The calibration curves for all of them using 20% ethanol as the standard solution were constructed. All curves had good linearity (0.001−0.050 μg/mL), and the correlation coefficients (> 0.990) and the LOQ for the migration levels in all migration solutions were estimated to be 0.001 μg/mL (0.002 μg/ cm2). These results were similar to those of the curve using 50% methanol containing 0.1% formic acid. The obtained olive oil migration solutions required treatment prior to LC/MS analysis, as described in the Materials and Methods section. To evaluate the performance of the method for determining the migration levels of PA monomers and oligomers in olive oil, recovery tests (n = 3) using 5 mL of olive oil in which were spiked 50 ng of each monomer and oligomer were performed, and the recovery rates and RSD values were assessed. As shown in Table 7, the recovery rates of the PA6 monomer and oligomers ranged from 73% to 105% and those of the PA66 monomer and oligomers ranged from 81% to 95%. The RSD values ranged from 1.1 to 10.2%. These are considered acceptable recovery rates and RSD values for all PA monomers and oligomers, suggesting that the proposed method was reliable for their quantitative determination in olive oil.
Table 7

Recovery rates of PA cyclic monomers and oligomers from olive oil.

Cyclic monomers and oligomersRecovery rate (%)
PA6 monomer (CPL)98 ± 3 (3.3)
PA6 dimer101 ± 2 (2.1)
PA6 trimer105 ± 6 (5.5)
PA6 tetramer103 ± 3 (2.6)
PA6 pentamer87 ± 9 (9.8)
PA6 hexamer73 ± 8 (10.2)
PA66 monomer92 ± 1 (1.1)
PA66 dimer95 ± 7 (7.1)
PA66 trimer81 ± 6 (7.9)

Each value is given as the mean ± SD (RSD%) of three trials.

Each oligomer (50 ng) was spiked in 5 mL of olive oil.

Each value is given as the mean ± SD (RSD%) of three trials. Each oligomer (50 ng) was spiked in 5 mL of olive oil. For kitchen utensils that are generally used at room temperature, such as sesame grinders and cake servers, migration tests were performed at 60°C for 30 min (reference to the official Japanese method conditions [9]) using water, 20% ethanol, and olive oil. As shown in Table 8, the results suggested that the PA monomers and oligomers more readily migrated into 20% ethanol than into water but migrated little into olive oil. The migration levels of PA monomers and oligomers decreased with increasing molecular weight. Specifically, there was a strongly negative linear correlation between the migration level and the molecular weight of the PA66 monomer and oligomers into water and 20% ethanol (R = 0.987 and 0.990, respectively).
Table 8

Migration levels of PA cyclic monomers and oligomers from the sesame grinder (PA6) and cake server (PA6/66) into three simulants at 60°C.

PA cyclic monomers and oligomersSesame grinder (PA6)Cake server (PA6/66)
Water20% EthanolOlive oilResidual levelWater20% EthanolOlive oilResidual level
PA6 monomer20210.603.72.53.30.0281.6
PA6 dimer2.57.70.0670.80.490.900.0030.32
PA6 trimer3.69.80.0341.60.541.30.0030.52
PA6 tetramer6.2200.0085.00.812.1ND1.4
PA6 pentamer1.56.1ND2.80.0690.69ND0.82
PA6 hexamer0.151.5ND1.30.0250.11ND0.45
Subtotal34660.71154.58.40.0335.1
PA66 monomerNDNDNDND9.7240.0833.7
PA66 dimerNDNDNDND4.2110.0043.9
PA66 trimerNDNDNDND0.332.3ND3.8
SubtotalNDNDNDND14370.08712
Total34660.711519450.1217

Each value (migration level: μg/cm2; residual level: mg/g) is mean of two trials.

ND (migration level): < 0.002 μg/cm2; ND (residual level): < 0.002 mg/g.

Each value (migration level: μg/cm2; residual level: mg/g) is mean of two trials. ND (migration level): < 0.002 μg/cm2; ND (residual level): < 0.002 mg/g. The migration levels of the PA6 tetramer from the sesame grinder and of the PA66 monomer from the cake server into 20% ethanol were similar to that of CPL from the sesame grinder. The migration limit of CPL to 20% ethanol is regulated in Japan [i.e., migration limit of 15 μg/mL (30 μg/cm2)] [9] and in the EU [by regulation 10/2011/EU; specific migration limit of 15 mg/kg)] [10]. However, there are no regulations regarding the migration levels of the PA6 tetramer and PA66 monomer. In Japan, such migration levels are only regulated by a limitation of the total residue upon evaporation [30 μg/mL (60 μg/cm2)], and in the EU by the overall migration limit (10 mg/dm2) of all migrated substances. The total migration level from the sesame grinder into 20% ethanol is 66 μg/cm2, which appears to exceed the Japanese regulation limit. However, the test for the level of residue upon evaporation measures the amount of the residual substance that is not volatilized from the test solution after the solution is evaporated to dryness on a water bath. Under these conditions, some PA monomers and oligomers would be volatilized, so the total level of residue upon evaporation would not exceed the regulatory limit. For the ladles, which are generally used at temperatures ranging from room temperature to 100°C, the migration tests were performed using a water bath heated at 95°C for 30 min using water, 20% ethanol, and olive oil. Typical results of the migration tests for the three selected samples made from PA6, PA66, and PA6/66 are shown in Table 9. Similar to the migration test performed at 60°C, the PA6 monomers and oligomers more readily migrated into 20% ethanol than into water but migrated little into olive oil.
Table 9

Migration levels of PA cyclic monomers and oligomers from ladles 1 (PA6), 8 (PA66), and 5 (PA6/66) into three simulants at 95°C.

PA cyclic monomers and oligomersLadle 1 (PA6)Ladle 8 (PA66)Ladle 5 (PA6/66)
Water20% EthanolOlive oilResidual levelWater20% EthanolOlive oilResidual levelWater20% EthanolOlive oilResidual level
PA6 monomer18230.674.80.440.470.0710.03318200.441.2
PA6 dimer0.852.30.0422.80.0260.0670.0230.0061.53.20.0440.51
PA6 trimer1.23.00.0252.80.0450.120.0020.0132.35.20.0460.60
PA6 tetramer3.4100.0174.60.0920.30ND0.0313.89.70.0281.3
PA6 pentamer3.011ND2.80.0070.019ND0.0100.163.9ND0.87
PA6 hexamer1.37.4ND1.40.0100.036ND0.0090.202.1ND0.58
Subtotal28560.75190.621.00.0960.1026440.565.0
PA66 monomer0.034ND0.0030.11398600.542.84610001.61.6
PA66 dimer0.031NDND0.088205600.0182.8267000.152.1
PA66 trimer0.008NDND0.0635.1260ND2.37.2480ND2.4
Subtotal0.073ND0.0030.266417000.567.97822001.76.1
Total28560.7519.36517000.668.010422002.311.1

Each value (migration level: μg/cm2; residual level: mg/g) is mean of two trials.

ND (migration level): < 0.002 μg/cm2; ND (residual level): < 0.002 mg/g.

Each value (migration level: μg/cm2; residual level: mg/g) is mean of two trials. ND (migration level): < 0.002 μg/cm2; ND (residual level): < 0.002 mg/g. On the other hand, the PA66 monomer and oligomers, surprisingly, migrated from ladle 8 (PA66) and ladle 5 (PA6/66) into 20% ethanol at levels ranging from 260 to 1,000 μg/cm2. These high levels of migration were probably due to the hydrolysis of the amide linkage of PA66 caused by the swelling ethanol and water. For turners, which are generally used at temperatures over 100°C, the migration tests were performed at 121°C for 30 min using olive oil. Typical results for the migration tests of three selected samples made from PA66 and PA6/66 are shown in Table 10. For the PA6 monomer and oligomers, the migration levels were roughly correlated with the residual levels. For the PA66 monomer and oligomers, although the migration levels were also correlated with the residual levels, the migration levels drastically decreased with increasing degree of polymerization.
Table 10

Migration levels of PA cyclic monomers and oligomers from turners 1 and 2 (PA66), and 10 (PA6/66) into olive oil at 121°C.

PA cyclic monomers and oligomersTurner 1 (PA66)Turner 2 (PA66)Turner 10 (PA6/66)
Migration levelResidual levelMigration levelResidual levelMigration levelResidual level
PA6 monomer0.0800.0260.0700.0030.730.63
PA6 dimer0.0120.013NDND0.510.43
PA6 trimer0.0130.019NDND0.170.27
PA6 tetramer0.0100.040NDND0.160.80
PA6 pentamerND0.011NDND0.0060.48
PA6 hexamerND0.008NDNDND0.23
Subtotal0.120.120.0700.0031.62.8
PA66 monomer6.25.46.24.77.33.5
PA66 dimer0.975.30.874.50.983.7
PA66 trimer0.0073.20.0073.90.0063.2
Subtotal7.2147.1138.310
Total7.3147.2139.913

Each value (migration level: μg/cm2; residual level: mg/g) is mean of two trials.

ND (migration level): < 0.002 μg/cm2; ND (residual level): < 0.002 mg/g.

Each value (migration level: μg/cm2; residual level: mg/g) is mean of two trials. ND (migration level): < 0.002 μg/cm2; ND (residual level): < 0.002 mg/g.

Conclusions

Five PA6 cyclic oligomers and three PA66 cyclic monomer and oligomers were successfully identified in kitchen utensils composed of PA6 and PA66, and their structures were characterized using HRMS and NMR analyses. Furthermore, the residual and migration levels of PA monomers and oligomers in kitchen utensils were quantified using a purchased PA6 monomer and isolated PA monomer and oligomers. The total residual levels of PA monomers and oligomers in PA materials ranged from 7.8 to 20 μg/cm2, and no significant differences were observed among the total residual levels in PA6, PA66, and PA6/66 materials. The migration tests suggested that these monomers and oligomers readily migrated into alcoholic beverages possibly due to their features of easy swelling, but hardly migrated into fatty foods possibly due to their low solubility and swelling in the oil. Furthermore, the migrant substances were primarily cyclic monomers and low molecular weight cyclic oligomers. In addition, the PA66-based materials might be easily swelled and degraded, potentially leading to a large amount of PA66 cyclic monomer and oligomers migrating on contact with alcohols at high temperature.

Supporting information for kitchen utensil samples.

(DOCX) Click here for additional data file.
  5 in total

1.  Determination of potential migrants present in Nylon 'microwave and roasting bags' and migration into olive oil.

Authors:  H Soto-Valdez; J W Gramshaw; H J Vandenburg
Journal:  Food Addit Contam       Date:  1997-04

2.  Analysis of migrants from nylon 6 packaging films into boiling water.

Authors:  C T Barkby; G Lawson
Journal:  Food Addit Contam       Date:  1993 Sep-Oct

3.  Determination of migrants in and migration from nylon food packaging.

Authors:  T H Begley; M L Gay; H C Hollifield
Journal:  Food Addit Contam       Date:  1995 Sep-Oct

4.  Cyclic oligomers in polyamide for food contact material: quantification by HPLC-CLND and single-substance calibration.

Authors:  M Heimrich; M Bönsch; H Nickl; T J Simat
Journal:  Food Addit Contam Part A Chem Anal Control Expo Risk Assess       Date:  2012-02-13

5.  Migration from polyamide 'microwave and roasting bags' into roast chicken.

Authors:  J W Gramshaw; H Soto-Valdez
Journal:  Food Addit Contam       Date:  1998-04
  5 in total
  3 in total

1.  Migration study of caprolactam from polyamide 6 sheets into food simulants.

Authors:  Hyun Ju Song; Yoonjee Chang; Ji Sou Lyu; Mi Yong Yon; Haeng-Shin Lee; Se-Jong Park; Jae Chun Choi; MeeKyung Kim; Jaejoon Han
Journal:  Food Sci Biotechnol       Date:  2018-05-29       Impact factor: 2.391

2.  Prediction of Collision Cross Section Values: Application to Non-Intentionally Added Substance Identification in Food Contact Materials.

Authors:  Xue-Chao Song; Nicola Dreolin; Tito Damiani; Elena Canellas; Cristina Nerin
Journal:  J Agric Food Chem       Date:  2022-01-18       Impact factor: 5.279

3.  Development of an LC-MS method for the semiquantitative determination of polyamide 6 contaminations in polyolefin recyclates.

Authors:  Andrea Schweighuber; Markus Gall; Jörg Fischer; Yi Liu; Hermann Braun; Wolfgang Buchberger
Journal:  Anal Bioanal Chem       Date:  2020-11-26       Impact factor: 4.478

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.