An in vitro study of the hemodynamic effects of catheter injections.

T W Morris1, C S Walike.   

Abstract

The injection jet from a catheter has been shown to elevate distal pressure and flow. These disturbances in flow present a problem for the application of digital arteriography to measure physiologic flow. The authors used an in vitro model to determine the primary factors responsible for the increases in distal flow with the goal that these disturbances might be estimated and minimized. Experiments were performed to evaluate the importance of the cross-sectional area of the catheter and vessel, the importance of absolute vessel flow, and the importance of distal resistance. A lumped resistance model predicted the flow changes to within 3% of the preinjection vessel flow. In the model system the elevation of distal flow in percent was equal to (the injection rate divided by the tube flow rate) multiplied by (the proximal resistance divided by the total resistance) multiplied by 100. Because the resistance ratio generally is small, injections at physiologic flow rates should produce acceptably small increases in distal flow. The use of a large diameter catheter or the presence of a proximal stenotic lesion would increase the ratio.

Mesh:

Substances:

Year:  1989        PMID: 2745016     DOI: 10.1097/00004424-198905000-00004

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  4 in total

1.  The mixability of angiographic contrast with arterial blood.

Authors:  Baruch B Lieber; Chander Sadasivan; Qing Hao; Jaehoon Seong; Liliana Cesar
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

2.  Impact of balloon inflation pressure on cell viability with single and multi lumen catheters.

Authors:  N Dib; D B Schwalbach; B D Plourde; R E Kohler; D Dana; J P Abraham
Journal:  J Cardiovasc Transl Res       Date:  2014-11-21       Impact factor: 4.132

3.  An in vitro study of pressure increases during contrast injections in diagnostic cerebral angiography.

Authors:  Samantha Marfoglio; Brandon Kovarovic; Wei Hou; David J Fiorella; Chander Sadasivan
Journal:  Interv Neuroradiol       Date:  2021-02-25       Impact factor: 1.764

4.  Effect of saline injection mixing on accuracy of conductance lumen sizing of peripheral vessels.

Authors:  Hyo Won Choi; Benjamin Jansen; David Birrer; Ghassan S Kassab
Journal:  PLoS One       Date:  2013-09-13       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.