Abstract

Infusion catheters, when used in combination with balloons, are subjected to pressure created by inflation of the balloon. The compression can reduce the catheter flow area and cause elevated shear stresses in the fluid. A model and experiments were developed with a range of applied balloon pressures to investigate whether such situations may cause cell lysis during stem-cell infusion through single-lumen catheters. It was found that for balloon inflation pressures in excess of ~7 atm, it is possible for cell injury to occur, although the critical pressure depends on the fluid viscosity. The study was then applied to a multi-lumen catheter which was designed to resist compression. That device was able to handle very elevated balloon pressures and flow rates before cell lysis became a concern.

Entities:  

Mesh:

Year:  2014        PMID: 25413890     DOI: 10.1007/s12265-014-9595-z

Source DB:  PubMed          Journal:  J Cardiovasc Transl Res        ISSN: 1937-5387            Impact factor:   4.132


  20 in total

1.  A discussion on the threshold limit for hemolysis related to Reynolds shear stress.

Authors:  M Grigioni; C Daniele; G D'Avenio; V Barbaro
Journal:  J Biomech       Date:  1999-10       Impact factor: 2.712

2.  Deformation and fragmentation of human red blood cells in turbulent shear flow.

Authors:  S P Sutera; M H Mehrjardi
Journal:  Biophys J       Date:  1975-01       Impact factor: 4.033

3.  Radiolabeled cell distribution after intramyocardial, intracoronary, and interstitial retrograde coronary venous delivery: implications for current clinical trials.

Authors:  Dongming Hou; Eyas Al-Shaykh Youssef; Todd J Brinton; Ping Zhang; Pamela Rogers; Erik T Price; Alan C Yeung; Brian H Johnstone; Paul G Yock; Keith L March
Journal:  Circulation       Date:  2005-08-30       Impact factor: 29.690

Review 4.  Methods of stem cell delivery in cardiac diseases.

Authors:  Emerson C Perin; Javier López
Journal:  Nat Clin Pract Cardiovasc Med       Date:  2006-03

Review 5.  Catheter-based delivery of cells to the heart.

Authors:  Warren Sherman; Timothy P Martens; Juan F Viles-Gonzalez; Tomasz Siminiak
Journal:  Nat Clin Pract Cardiovasc Med       Date:  2006-03

6.  An in vitro study of the hemodynamic effects of catheter injections.

Authors:  T W Morris; C S Walike
Journal:  Invest Radiol       Date:  1989-05       Impact factor: 6.016

7.  Red blood cell damage by shear stress.

Authors:  L B Leverett; J D Hellums; C P Alfrey; E C Lynch
Journal:  Biophys J       Date:  1972-03       Impact factor: 4.033

8.  Hemolysis near an ultrasonically pulsating gas bubble.

Authors:  J A Rooney
Journal:  Science       Date:  1970-08-28       Impact factor: 47.728

9.  Hemolysis near a transversely oscillating wire.

Authors:  A R Williams; D E Hughes; W L Nyborg
Journal:  Science       Date:  1970-08-28       Impact factor: 47.728

Review 10.  Stem cells for cardiac regeneration by cell therapy and myocardial tissue engineering.

Authors:  Jun Wu; Faquan Zeng; Richard D Weisel; Ren-Ke Li
Journal:  Adv Biochem Eng Biotechnol       Date:  2009       Impact factor: 2.635

View more
  2 in total

1.  Minimizing thermal damage to vascular nerves while drilling of calcified plaque.

Authors:  Seifollah Gholampour; Keyvan Hajirayat
Journal:  BMC Res Notes       Date:  2019-06-14

Review 2.  Stem Cell Therapy for the Heart: Blind Alley or Magic Bullet?

Authors:  Arne A N Bruyneel; Apurv Sehgal; Sophia Malandraki-Miller; Carolyn Carr
Journal:  J Cardiovasc Transl Res       Date:  2016-08-19       Impact factor: 4.132

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.