Falsely increased bispectral index values by convective air warming system during kidney transplantation.

Se Hun Kim1, Byeong-Cheol Lee2, Yong Han Kim3.   

Abstract

Bispectral index (BIS) is a reliable parameter for measuring depth of hypnotic level during anesthesia. Convective air warming system is an effective equipment to maintain normothermia during operation. We report falsely elevated BIS value due to convective air warming system while undergoing kidney transplantation.

Entities:  

Keywords:  Air warmer; Artifact; Bispectral index; Depth of anesthesia

Year:  2016        PMID: 27375736      PMCID: PMC4928445          DOI: 10.12669/pjms.323.9858

Source DB:  PubMed          Journal:  Pak J Med Sci        ISSN: 1681-715X            Impact factor:   1.088


INTRODUCTION

Bispectral index (BIS) has been used to measure sedative-hypnotic depth during anesthesia. However, BIS value may be influenced by several agents or movement to lead misinterpretation.1 Anesthesiologist makes an effort to maintain normal body temperature. Convective air warming system (WarmTouch™, Covidien, USA) is a useful device to prevent hypothermia. WarmTouch™ can produce high value of BIS inaccurately. We report a case of inexact reading with BIS influenced by WarmTouch™ during kidney transplantation.

CASE REPORT

A 38-year-old female patient (height: 153cm, weight: 55kg) was scheduled for kidney transplantation from cadaveric donor. She had been diagnosed with chronic kidney disease about two years ago and on hemodialysis for one year. Anesthesia was induced with propofol 100mg and atracurium 25mg with equipment of monitoring including BIS, invasive arterial pressure, pulse oximeter and electrocardiogram. After tracheal intubation, the anesthesia was maintained by 6% desflurane, 0.1μg/kg·min remifentanil and intermittent atracurium. Triple lumen venous catheter was placed in the right internal jugular vein with fluid infusion and central venous pressure monitor. Esophageal probe was inserted to record body temperature. Warming blanket with convective air warming system covered chest, both arms and head to keep body temperature normal (Fig.1). BIS value maintained from 35 to 50 and electromyography (EMG) was almost fixed at 27 during one and half hour of operation. BIS Signal Quality Index (BIS-SQI) and Suppression Ratio (SR) was 100 and 0, respectively. Blood pressure was 131 over 72 mmHg, and heart rate was 69 beats per minute. WarmTouch™ was operated for body heating because esophageal temperature dropped from 36.3 to 36.0ºC. BIS and EMG increased to 74 and 37, respectively. BIS-SQI and SR was 100 and 0, unchanged. Hemodynamics was steady state. The anesthesiologist added up desflurane to 8% concentration and inject atracurium 5mg. However, BIS was over 70 and EMG was 37. After turning off WarmTouch™, BIS and EMG returned to 45 and 27. BIS and EMG was raised again to same numerical value after switch on WarmTouch™. Heated and humidified breathing circuit (Heated Circuit Kit®, Ace Medical, South Korea) was applied for normothermia without activation of WarmTouch™. Esophageal temperature was maintained at 36.0 ºC to recovery of anesthesia. Total operation time was four hours and the patient did not recall any intraoperative event. The patient was discharged without any symptom or complication on postoperative 22 days.
Fig.1

BIS was surrounded by WarmTouch ™ and blanket.

BIS was surrounded by WarmTouch ™ and blanket.

DISCUSSION

BIS monitor is a helpful device for adequate depth of anesthesia during operation. The value of BIS is superior to clinical signs for hypnotic effect of anesthesia.2 However, several anesthetics, devices and clinical conditions could affect BIS value.1 BIS was increased after ketamine bolus (0.4mg/kg) followed by continuous infusion (1mg/kg·h).3 Halothane anesthesia reflects higher BIS than sevoflurane in breast surgery, therefore halothane inhalation based on only BIS has a possibility of overdose.4 BIS of halothane at 1 MAC is 56.1 whereas isoflurane is 33.2 in children.5 Some of the electronic devices could interfere BIS parameter. Forced air warmer (Bair Hugger®, Augustine Medical, USA) elevated BIS without hemodynamic change during hepatectomy with isoflurane, sufentanyl, and atracurium.6 Falsely increased BIS was also noted during cardiac surgery heated by Bair Hugger® with sevoflurane, fentanyl and rocuronium anesthesia without memory recall.7 Endoscopic shaver attributed falsely elevated BIS with sevoflurane, remifentanil and rocuronium during endoscopic shoulder surgery.8 Pacemaker-induced artifact appeared after deep hypothermic circulatory arrest with propofol, remifentanil and rocuronium in cardiac surgery.9 BIS increased falsely in patients with electromagnetic operating device (VTI surgery). Perioperative normothermia is very important to prevent following complications: delayed recovery, myocardial ischemia, shivering, hypertension, tachycardia, and prolonged coagulation time. Various methods could be adapted for maintenance of temperature including forced air blanket, water mattress, warm intravenous fluid, and heated airway circuit.10 WarmTouch™ is an effective method for preventing complication of hypothermia. The vibration of air warmer could interfere BIS value through high EMG signal. BIS represented higher value than real depth of hypnotic level. The ventilation outlet of WarmTouch™ was set close to BIS as it seems in Fig.1. In conclusion, BIS can be interpreted incorrectly by convective air warming system, therefore, deserves a special attention.
  9 in total

1.  Bispectral index increases and decreases are not always signs of inadequate anesthesia.

Authors:  B Guignard; M Chauvin
Journal:  Anesthesiology       Date:  2000-03       Impact factor: 7.892

2.  Falsely increased bispectral index during endoscopic shoulder surgery attributed to interferences with the endoscopic shaver device.

Authors:  Thomas M Hemmerling; Brigitte Migneault
Journal:  Anesth Analg       Date:  2002-12       Impact factor: 5.108

3.  Bispectral index scale is higher for halothane than sevoflurane during intraoperative anesthesia.

Authors:  Johnathan J Edwards; Roy G Soto; David M Thrush; Robert F Bedford
Journal:  Anesthesiology       Date:  2003-12       Impact factor: 7.892

4.  Falsely increased bispectral index values in a series of patients undergoing cardiac surgery using forced-air-warming therapy of the head.

Authors:  Thomas M Hemmerling; Joanne D Fortier
Journal:  Anesth Analg       Date:  2002-08       Impact factor: 5.108

5.  Assessment of depth of anesthesia: PRST score versus bispectral index.

Authors:  Jasmina Smajic; Mirsada Praso; Mirsad Hodzic; Samir Hodzic; Amna Srabovic-Okanovic; Nedim Smajic; Zinka Djonlagic
Journal:  Med Arh       Date:  2011

Review 6.  Different conditions that could result in the bispectral index indicating an incorrect hypnotic state.

Authors:  Ashraf A Dahaba
Journal:  Anesth Analg       Date:  2005-09       Impact factor: 5.108

7.  The Bispectral Index in children: comparing isoflurane and halothane.

Authors:  A J Davidson; C Czarnecki
Journal:  Br J Anaesth       Date:  2004-01       Impact factor: 9.166

8.  A comparison of bispectral index and ARX-derived auditory evoked potential index in measuring the clinical interaction between ketamine and propofol anaesthesia.

Authors:  H E M Vereecke; M M R F Struys; E P Mortier
Journal:  Anaesthesia       Date:  2003-10       Impact factor: 6.955

9.  Falsely increased bispectral index score during deep hypothermic circulatory arrest in cardiac surgery.

Authors:  Jae Ouk Bang; Hyo Jung Son; Eun Ho Lee; Kyung-Don Hahm; In-Cheol Choi
Journal:  Korean J Anesthesiol       Date:  2012-10-12
  9 in total
  2 in total

1.  Influence of electrocautery-induced electromagnetic interference on quantitative electroencephalographic monitoring of hypnosis during general anesthesia: comparison between the ADMS® and the BIS VISTATM.

Authors:  Dae-Hee Kim; Ji-Young Yoo; Jong-Yeop Kim; Soo-Hwan Ahn; Seongsu Kim; Sang-Kee Min
Journal:  Korean J Anesthesiol       Date:  2018-04-24

2.  Data Driven Investigation of Bispectral Index Algorithm.

Authors:  Hyung-Chul Lee; Ho-Geol Ryu; Yoonsang Park; Soo Bin Yoon; Seong Mi Yang; Hye-Won Oh; Chul-Woo Jung
Journal:  Sci Rep       Date:  2019-09-24       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.