| Literature DB >> 35166011 |
Alex T Price1, Nels C Knutson1, Taeho Kim1, Olga L Green1.
Abstract
Secondary external dose calculations for a 0.35 T magnetic resonance image-guided radiation therapy (MRgRT) are needed within the radiation oncology community to follow safety standards set forth within the field. We evaluate the commercially available software, RadCalc, in its ability to accurately perform monitor unit dose calculations within a magnetic field. We also evaluate the potential effects of a 0.35 T magnetic field upon point dose calculations. Monitor unit calculations were evaluated with (wMag) and without (noMag) a magnetic field considerations in RadCalc for the ViewRay MRIdian. The magnetic field is indirectly accounted for by using asymmetric profiles for calculation. The introduction of double-stacked multi-leaf collimator leaves was also included in the monitor unit calculations and a single transmission value was determined. A suite of simple and complex geometries with a variety field arrangements were calculated for each method to demonstrate the effect of the 0.35 T magnetic field on monitor unit calculations. Finally, 25 patient-specific treatment plans were calculated using each method for comparison. All simple geometries calculated in RadCalc were within 2% of treatment planning system (TPS) values for both methods, except for a single noMag off-axis comparison. All complex muilt-leaf collimator (MLC) pattern calculations were within 5%. All complex phantom geometry calculations were within 5% except for a single field within a lung phantom at a distal point. For the patient calculations, the noMag method average percentage difference was 0.09 ± 2.5% and the wMag average percentage difference was 0.08 ± 2.5%. All results were within 5% for the wMag method. We performed monitor unit calculations for a 0.35 T MRgRT system using a commercially available secondary monitor unit dose calculation software and demonstrated minimal impact of the 0.35 T magnetic field on monitor unit dose calculations. This is the first investigation demonstrating successful calculations of dose using RadCalc in the low-field 0.35 T ViewRay MRIdian system.Entities:
Keywords: MRI-guided radiation therapy; MRgRT; quality assurance; secondary dose calculations
Mesh:
Year: 2022 PMID: 35166011 PMCID: PMC8906210 DOI: 10.1002/acm2.13452
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
FIGURE 1(a) MRIdian MLCs that only have one single leaf in the middle of the field. This single leaf would have a transmission value of approximately 35.0%. A closed set of interdigitated single layered MLCs has a transmission value of approximately 7.8%. All of the other leaves have mostly double‐stacked MLC patterns which have a transmission of approximately <0.1%. (b) The MLCs as modeled in RadCalc version 7.1.4.0
FIGURE 2(a) The cross‐pattern used to test complex MLC patterns. The calculation points identified in (a) have corresponding calculation results in (b and c). The calculated doses and percentage differences are presented. (b) The noMag method and (c) the wMag method. In the name column, the first value for the description is the depth whereas the second value is the distance in cm off‐axis
FIGURE 3Complex heterogeneous phantoms used for calculation. (a) The lung phantom and (b) the bowel phantom
Disease site, modality, fractionation scheme, and comparison of the RadCalc percentage difference of the ViewRay plan dose and the version 7.1.4.0 calculated dose
| Methods of verifying calculated dose | |||
|---|---|---|---|
| Site | Modality | Fractionation scheme | RadCalc % difference |
| Abdomen | SBRT | 700 cGy × 5 | 2.4% |
|
| SBRT | 1000 cGy × 5 | 3.8% ( |
| Breast | SBRT | 850 cGy × 3 | 2.1% |
| Breast | SBRT | 850 cGy × 3 | –0.7% |
| Chestwall | IMRT | 200 cGy × 30 | –4.2% |
| Chestwall | SBRT | 700 cGy × 5 | 1.2% |
| Liver | SBRT | 1000 cGy × 5 | –0.6% |
| Liver | SBRT | 1000 cGy × 5 | –0.5% |
| Liver | SBRT | 1000 cGy × 5 | –1.1% |
| Liver | SBRT | 1000 cGy × 5 | 3.9% |
| Liver | SBRT | 1000 cGy × 5 | –2.6% |
| Liver | SBRT | 1000 cGy × 5 | 4.0% |
| Liver | SBRT | 1000 cGy × 5 | 0.4% |
| Lung | IMRT | 500 cGy × 12 | –1.0% |
|
| IMRT | 400 cGy × 15 | –0.3% ( |
|
| IMRT | 400 cGy × 15 | –2.5% ( |
| Pancreas | SBRT | 1000 cGy × 5 | 1.1% |
| Pancreas | SBRT | 1000 cGy × 5 | –3.1% |
| Pancreas | SBRT | 1000 cGy × 5 | 3.0% |
| Pancreas | SBRT | 1000 cGy × 5 | –1.2% |
| Pancreas | SBRT | 1000 cGy × 5 | 2.7% |
| Pancreas | SBRT | 1000 cGy × 5 | 2.9% |
|
| SBRT | 1000 cGy × 5 | –3.8% ( |
| Pancreas | SBRT | 1000 cGy × 5 | –4.3% |
| Periportal | SBRT | 700 cGy × 5 | 0.5% |
Bold treatment site represents significant heterogeneity within the beam path.
Abbreviation: SBRT, Stereotactic body radiation thearpy.
FIGURE 4Box and whisker plot of the 25 patient results with varying levels of transmission values. Note that 1.5% transmission is the data on the left, 1.625% MLC transmission is in the middle, and 1.75% MLC transmission is on the right
Comparison of the noMag and wMag methods for large fields and changing source to surface distance (SSD) scenarios
| Large field comparison | ||||||
|---|---|---|---|---|---|---|
| noMag method | wMag method | |||||
| Location | 70SSD | 80SSD | 90SSD | 70SSD | 80SSD | 90SSD |
|
| 0.5% | 0.4% | 0.5% | 0.5% | 0.4% | –0.5% |
|
| –0.2% | 0.7% | 1.3% | 0.0% | 0.7% | 0.9% |
|
| –1.3% | –0.6% | 0.3% | –1.7% | –0.8% | 0.3% |
|
| 0.0% | –0.1% | 0.9% | –1.0% | –0.9% | 0.4% |
|
| 0.9% | –0.6% | 0.1% | –1.9% | –1.4% | –0.5% |
Note: All points are within 2% which is within MPPG5a tolerances.
Abbreviations: MPPG5a, Medical Physics Practice Guideline 5.a; OAX, off‐axis.
Comparison of the noMag and wMag methods for small fields
| Small field comparison | ||
|---|---|---|
| Location | noMag method | wMag method |
|
| 0.9% | 0.5% |
|
| 2.2% | 1.6% |
|
| 0.3% | 0.8% |
|
| 1.4% | 1.1% |
|
| 0.7% | 0.4% |
Note: All points are within 2% which is within MPPG5a tolerances except for single OAX noMag comparison.
Abbreviations: MPPG5a, Medical Physics Practice Guideline 5.a; OAX, off‐axis.
Comparison of the noMag and wMag methods for variety of field arrangements for both the lung and bowel phantom
| Heterogeneity comparison | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| noMag method | wMag method | |||||||||||
| Location | AP‐L | 2Fd‐L | 4Fd‐L | AP‐B | 2Fd‐B | 4Fd‐B | AP‐L | 2Fd‐L | 4Fd‐L | AP‐B | 2Fd‐B | 4Fd‐B |
| O | 3.96% | 3.98% | 4.76% | 1.67% | 1.48% | 1.03% | 4.18% | 3.96% | 4.76% | 1.65% | 1.48% | 0.87% |
| A | N/A | –3.47% | –2.86% | N/A | 0.94% | 0.60% | N/A | –3.40% | –2.36% | N/A | 1.04% | 1.05% |
| B | N/A | N/A | N/A | N/A | 1.80% | –2.86% | N/A | N/A | N/A | N/A | 1.87% | –0.17% |
| C | N/A | N/A | N/A | N/A | 1.12% | –2.82% | N/A | N/A | N/A | N/A | 1.16% | –0.09% |
| D | N/A | –3.04% | –2.42% | N/A | 0.60% | 0.33% | N/A | –3.04% | –1.97% | N/A | 0.64% | 0.76% |
| E | 0.03% | N/A | –2.82% | 0.58% | N/A | 3.24% | 0.15% | N/A | –2.16% | 0.58% | N/A | 3.63% |
| F | N/A | N/A | N/A | 1.27% | N/A | –2.04% | N/A | N/A | N/A | 1.26% | N/A | 0.36% |
| G | N/A | N/A | N/A | 3.23% | N/A | –0.60% | N/A | N/A | N/A | 3.25% | N/A | 1.87% |
| H | –8.47% | N/A | –4.87% | 1.69% | N/A | 1.25% | –8.46% | N/A | –4.27% | 1.67% | N/A | 1.64% |
Note: “‐L” denotes the lung phantom, whereas the “‐B” denotes the bowel phantom.
Comparison of the noMag and wMag methods for calculation points at 1 cm from the gas heterogeneity in the bowel phantom
| Near heterogeneity comparison | ||
|---|---|---|
| Location | noMag method | wMag method |
| A | 2.65% | 2.79% |
| B | 3.33% | 3.43% |
| C | 4.03% | 4.15% |
| D | 1.70% | 1.82% |
FIGURE 5Box and whisker plot of the 25 patient results with using different calculation methods