Alexandre Renaud1, Alexandre Fuentes2, Nicola Hagemeister3, Martin Lavigne1, Pascal-André Vendittoli1.
Abstract
BACKGROUND: Various implants of total knee arthroplasty (TKA) are used in clinical practice and each presents specific design characteristics. No implant managed this day to reproduce perfectly the biomechanics of the natural knee during gait.Entities:
Keywords: arthroplasty; design; function; gait; kinematic; knee replacement; total knee
Year: 2016 PMID: 27347244 PMCID: PMC4897418 DOI: 10.2174/1874325001610010155
Source DB: PubMed Journal: Open Orthop J ISSN: 1874-3250
Descriptive characteristics of the groups.
| Descriptive characteristics | Bilateral TKA | Asymptomatic | |
|---|---|---|---|
| Age (years) | 68.2 | 37.4 | < 0.001 |
| Weight (kg) | 90.9 | 73.0 | 0.001 |
| Height (cm) | 166.0 | 172.3 | 0.054 |
| Body mass index (kg/m2) | 33.1 | 24.4 | < 0.001 |
| Gender | 8 females | 13 females | 0.575 |
Kinematic parameters results for NexgenTM TKAs, TriathlonTM TKAs and asymptomatic knees.
| TriathlonTM TKA | NexgenTM TKA | Asymptomatic knees | |||
|---|---|---|---|---|---|
| Total ROM (°) | 47.1 | 48.2 | 0.4 | 55.4 | 0.001 |
| Stance phase ROM (°) | 15.3 | 13.6 | 0.4 | 17.4 | 0.03 |
| Swing phase ROM (°) | 41.9 | 43.9 | 0.1 | 51.1 | 0.001 |
| Peak knee flexion (°) | 50.4 | 52.8 | 0.3 | 57.6 | 0.003 |
| Total ROM (°) | 10.1 | 9.6 | 0.6 | 9.9 | 0.9 |
| Stance phase ROM (°) | 4.6 | 4.7 | 0.7 | 5.4 | 0.4 |
| Swing phase ROM (°) | 9.0 | 8.6 | 0.7 | 9.0 | 0.9 |
| Peak stance angle (°) | 4.9 | 6.6 | 0.3 | 3.4 | 0.05 |
| Peak swing angle (°) | 9.0 | 9.2 | 0.9 | 3.6 | 0.001 |
| ROM IC – max mid stance (°) | 0.4 | 1.6 | 0.03 | 1.1 | 0.1 |
| Δ mean mid stance – IC (°) | -0.4 | 0.8 | 0.02 | 0.3 | 0.1 |
| Total ROM (°) | 12.7 | 10.9 | 0.2 | 12.6 | 0.5 |
| Stance phase ROM (°) | 8.5 | 6.7 | 0.2 | 8.9 | 0.2 |
| Swing phase ROM (°) | 11.8 | 10.1 | 0.2 | 10.9 | 0.6 |
| Peak stance internal rotation (°) | 2.9 | 1.9 | 0.6 | 4.5 | 0.1 |
* paired t test TriathlonTM vs NexgenTM
* one-way ANOVA TriathlonTM vs NexgenTM vs asymptomatic knees
KOOS.
| Sub total | TriathlonTM | NexgenTM | |
|---|---|---|---|
| Symptoms and stiffness | 80.0 | 74.5 | 0.21 |
| Pain | 79.8 | 79.8 | 1.0 |
| Function, daily living | 79.7 | 80.4 | 0.86 |
| Function, sports and recreational activities | 29.3 | 22.3 | 0.28 |
| Quality of life | 67.1 | 69.6 | 0.64 |
| 72.7 | 72.0 | 0.86 |
Joint perception.
| Joint perception | TriathlonTM | NexgenTM |
|---|---|---|
| Native or natural joint | 2 (13.3 %) | 1 (6.7 %) |
| Artificial joint with no restriction | 4 (26.7 %) | 3 (20 %) |
| Artificial joint with minimal restriction | 7 (46.7 %) | 9 (60 %) |
| Artificial joint with major restriction | 2 (13.3 %) | 2 (13.3 %) |
| Non functional joint | 0 (0%) | 0 (%) |