Reliability of a method for analyzing three-dimensional knee kinematics during gait.

David R Labbe1, Nicola Hagemeister, Mathieu Tremblay, Jacques de Guise.   

Abstract

BACKGROUND: Different attachment systems have been proposed in an effort to reduce skin movement artifacts when recording knee bone movement during gait. One such system, called exoskeleton, has shown promising accuracy but little is known concerning its reliability. The objective of this study was to determine the intra- and inter-observer reliability of this attachment system for recording 3D knee kinematics during gait.
METHODS: Two separate studies were conducted. The intra-observer study involved one observer who reinstalled the exoskeleton on 15 healthy subjects and recorded gait kinematics four times for each subject. The inter-observer study also involved 15 healthy subjects and for each of these subjects, three observers reinstalled and recorded gait kinetics three times in randomized order.
FINDINGS: In the intra-observer setting, ICC values were 0.92, 0.94 and 0.88 for knee flexion/extension, abduction/adduction and internal/external tibial rotation, respectively. In the inter-observer setting, the corresponding values were 0.94, 0.92 and 0.89.
INTERPRETATION: The high ICC values found indicate very high reliability of the exoskeleton for recording 3D knee kinematics despite reinstallation. Moreover, the values between both settings are very similar which indicates that reliability is independent of the observer who performs the installation. Therefore, evaluations may be carried out by several different clinicians without impacting reliability.

Mesh:

Year:  2007        PMID: 18155553     DOI: 10.1016/j.gaitpost.2007.11.002

Source DB:  PubMed          Journal:  Gait Posture        ISSN: 0966-6362            Impact factor:   2.840


  9 in total

Review 1.  The KneeKG system: a review of the literature.

Authors:  Sébastien Lustig; Robert A Magnussen; Laurence Cheze; Philippe Neyret
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-01-04       Impact factor: 4.342

2.  Gait changes of the ACL-deficient knee 3D kinematic assessment.

Authors:  B Shabani; D Bytyqi; S Lustig; L Cheze; C Bytyqi; P Neyret
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-07-16       Impact factor: 4.342

3.  Gait knee kinematic alterations in medial osteoarthritis: three dimensional assessment.

Authors:  Dafina Bytyqi; Bujar Shabani; Sebastien Lustig; Laurence Cheze; Natyra Karahoda Gjurgjeala; Philippe Neyret
Journal:  Int Orthop       Date:  2014-03-12       Impact factor: 3.075

Review 4.  Gait analysis methodology for the measurement of biomechanical parameters in total knee arthroplasties. A literature review.

Authors:  Georgios I Papagiannis; Athanasios I Triantafyllou; Ilias M Roumpelakis; Panayiotis J Papagelopoulos; George C Babis
Journal:  J Orthop       Date:  2018-02-02

5.  Effects of Postoperative Rehabilitation on Gait Parameters and Electromyography Variables in Acute and Chronic Anterior Cruciate Ligament Reconstruction Surgery in Football Players.

Authors:  Gopal Nambi; Walid Kamal Abdelbasset; Anju Verma; Shereen H Elsayed; Osama R Aldhafian; Naif Bin Nwihadh; Mohamed A Omar; Tohamy G T Hassan; Ayman K Saleh
Journal:  Evid Based Complement Alternat Med       Date:  2021-08-13       Impact factor: 2.629

6.  Kinematic Analysis of Gait Following Intra-articular Corticosteroid Injection into the Knee Joint with an Acute Exacerbation of Arthritis.

Authors:  Saurabh Mehta; Barbara L Shay; Tony Szturm; Hani S El-Gabalawy
Journal:  Physiother Can       Date:  2011-10-20       Impact factor: 1.037

7.  Femorotibial alignment measured during robotic assisted knee surgery is reliable: radiologic and gait analysis.

Authors:  Etienne Deroche; Alexandre Naaim; Timothy Lording; Raphael Dumas; Elvire Servien; Laurence Cheze; Sébastien Lustig; Cécile Batailler
Journal:  Arch Orthop Trauma Surg       Date:  2021-07-03       Impact factor: 3.067

8.  Clinical and Biomechanical Evaluations of Staged Bilateral Total Knee Arthroplasty Patients with Two Different Implant Designs.

Authors:  Alexandre Renaud; Alexandre Fuentes; Nicola Hagemeister; Martin Lavigne; Pascal-André Vendittoli
Journal:  Open Orthop J       Date:  2016-05-31

9.  Knee kinematics in anatomic anterior cruciate ligament reconstruction with four- and five-strand hamstring tendon autografts.

Authors:  Anders Sideris; Ali Hamze; Nicky Bertollo; David Broe; William Walsh
Journal:  Orthop Rev (Pavia)       Date:  2018-09-05
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.