High Risk of Surgical Glove Perforation From Surgical Rotatory Instruments.

Ashton H Goldman1, Emanuel Haug1, John R Owen1, Jennifer S Wayne2, Gregory J Golladay1.   

Abstract

BACKGROUND: Surgical gloves can be damaged during the course of a procedure, which can place the surgeon and patient at risk. Glove perforation may not always be readily apparent, and determining the risk factors for glove perforation can aid the surgeon in deciding when a glove change is advisable. Time of wear and needle sticks have been well studied; however, other mechanisms including mechanical stress from surgical equipment have had limited evaluation to date. QUESTIONS/PURPOSES: We evaluated the risk of glove perforation in gloves that were caught in a surgical rotatory device (such as drills and reamers). The aims of our study were (1) to determine the percentage of undetected microperforations after entanglement on a rotatory tool during orthopaedic procedures, (2) to determine which kinds of rotatory devices most commonly cause such microperforations, and (3) to assess whether time of wear had an effect on the risk of perforation.
METHODS: From July 2014 to September 2015, 33 gloves were obtained from all orthopaedic subspecialties at our Level I trauma center if they were caught in a rotatory device greater than one revolution. Time of glove wear and location of the glove that was caught in a rotatory device were recorded. After an evaluation for macroperforations (≥ 5 mm), the gloves were evaluated for microperforations (< 5 mm) via the American Society for Testing and Materials (ASTM) one-liter load test. Time of wear was compared among gloves with macroperforations, microperforations, and no perforations.
RESULTS: The 33 gloves obtained came from 33 procedures. Seventeen of 33 (52 %) gloves had perforations. Seven of the 17 perforated gloves had macroperforations while 10 had microperforations. Eleven of 33 entanglements were caught by drills, nine by reamers, eight by K-wires, and the remaining five gloves were caught by various other instruments. Eight of 17 perforations were caused by drills, three by reamers, three by K-wires, and three by various other instruments. The average time of wear was 58 minutes which did not differ with pattern of glove damage.
CONCLUSION: Surgical gloves caught in a rotatory power instrument are likely to have been perforated, regardless of the amount of time that they had been worn. Visual inspection appears to be an inadequate test of glove integrity. When a glove becomes entangled in a rotary instrument such as a drill, pin, or reamer, a surgeon should change the gloves regardless of whether he or she believes a perforation is present. LEVEL OF EVIDENCE: Level IV, therapeutic study.

Entities:  

Mesh:

Year:  2016        PMID: 27339122      PMCID: PMC5052191          DOI: 10.1007/s11999-016-4948-3

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  18 in total

Review 1.  Risk of transmission of blood-borne viral infection in orthopaedic and trauma surgery.

Authors:  R Lemaire; J B Masson
Journal:  J Bone Joint Surg Br       Date:  2000-04

2.  Glove perforation in pediatric orthopedic practice.

Authors:  Ibrahim Al-Habdan; Mir Sadat-Ali
Journal:  J Pediatr Orthop       Date:  2003 Nov-Dec       Impact factor: 2.324

3.  A randomized prospective study of glove perforation in orthopaedic surgery: is a thick glove more effective?

Authors:  Chang Dong Han; Jieun Kim; Sung Hwan Moon; Byung Ho Lee; Hyuk Min Kwon; Kwan Kyu Park
Journal:  J Arthroplasty       Date:  2013-06-06       Impact factor: 4.757

4.  Risk of glove perforation with arthroscopic knot tying using different surgical gloves and high-tensile strength sutures.

Authors:  Adriana Martinez; Yung Han; Zeeshan M Sardar; Lorne Beckman; Thomas Steffen; Bruce S Miller; Paul A Martineau
Journal:  Arthroscopy       Date:  2013-07-19       Impact factor: 4.772

Review 5.  Gloves, extra gloves or special types of gloves for preventing percutaneous exposure injuries in healthcare personnel.

Authors:  Christina Mischke; Jos H Verbeek; Annika Saarto; Marie-Claude Lavoie; Manisha Pahwa; Sharea Ijaz
Journal:  Cochrane Database Syst Rev       Date:  2014-03-07

6.  Efficacy of double-gloving as a barrier to microbial contamination during total joint arthroplasty.

Authors:  S F McCue; E W Berg; E A Saunders
Journal:  J Bone Joint Surg Am       Date:  1981-06       Impact factor: 5.284

Review 7.  Double gloving to reduce surgical cross-infection.

Authors:  J Tanner; H Parkinson
Journal:  Cochrane Database Syst Rev       Date:  2006-07-19

8.  Surgeons' and residents' double-gloving practices at 2 teaching hospitals in Ontario.

Authors:  Ted Haines; Bernadette Stringer; Jeremy Herring; Achilleas Thoma; Kenneth A Harris
Journal:  Can J Surg       Date:  2011-04       Impact factor: 2.089

9.  Risk of sharp device-related blood and body fluid exposure in operating rooms.

Authors:  Douglas J Myers; Carol Epling; John Dement; Debra Hunt
Journal:  Infect Control Hosp Epidemiol       Date:  2008-12       Impact factor: 3.254

10.  Surgical glove bacterial contamination and perforation during total hip arthroplasty implantation: when gloves should be changed.

Authors:  J Beldame; B Lagrave; L Lievain; B Lefebvre; N Frebourg; F Dujardin
Journal:  Orthop Traumatol Surg Res       Date:  2012-05-10       Impact factor: 2.256

View more
  3 in total

1.  Damages with High Consequences: Analysis of Perforations in Surgical Latex Operation Gloves from Orthopedic Surgeries.

Authors:  Andreas Enz; Annett Klinder; Hannah Mittelmeier; Günther Kundt; Wolfram Mittelmeier; Sarah Zaatreh
Journal:  Eur J Microbiol Immunol (Bp)       Date:  2018-11-26

2.  Intraoperative damage to surgical gloves during various operations on the musculoskeletal system: a multicenter study.

Authors:  Andreas Enz; Tanja Kostuj; Philipp Warnke; Katrin Osmanski-Zenk; Wolfram Mittelmeier; Annett Klinder
Journal:  Arch Orthop Trauma Surg       Date:  2020-08-29       Impact factor: 3.067

3.  Wearable Natural Rubber Latex Gloves with Curcumin for Torn Glove Detection in Clinical Settings.

Authors:  Norfatirah Muhamad Sarih; Nuur Syuhada Dzulkafly; Simon Maher; Azura A Rashid
Journal:  Polymers (Basel)       Date:  2022-07-28       Impact factor: 4.967

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.