Petra M Härtl1, Marius Treutwein2, Matthias G Hautmann1, Manuel März1, Fabian Pohl1, Oliver Kölbl1, Barbara Dobler1.
Abstract
INTRODUCTION: A sweeping beam technique for total body irradiation in standard treatment rooms and for standard linear accelerators (linacs) is introduced, which does not require any accessory attached to the linac. Lung shielding is facilitated to reduce the risk of pulmonary toxicity. Additionally, the applicability of a commercial radiotherapy planning system (RTPS) is examined.Entities:
Keywords: 3D treatment planning; Dosimetry; Lung shielding; Sweeping beam technique; Total body irradiation
Mesh:
Year: 2016 PMID: 27287010 PMCID: PMC4902948 DOI: 10.1186/s13014-016-0658-y
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1Linac in starting position and patient setup (Alderson phantom) on the couch
Parameters of the fixed beam fields (according IEC 61217 [36])
| Field | Gantry Angle | Collimator Angle | Wedge | X1 |
| Y1 | Y2 | Monitor Units |
|---|---|---|---|---|---|---|---|---|
| Cranial | 330° | 270° | 60° | −16.0 | 16.0 | −14.5 | 14.5 | 316 |
| Caudal 1 | 36° | 90° | No | −16.0 | 16.0 | −8.0 | 20.0 | 60 |
| Caudal 2 | 36° | 90° | 60° | −16.0 | 16.0 | −15.0 | 15.0 | 268 |
| Caudal 3 | 36° | 90° | No | −16.0 | 16.0 | −15.0 | 20.0 | 50 |
Fig. 2Profiles along the main axis measured with ionization chamber in a cuboid phantom for three different heights. Starting from position 135 cm to the right the height is 11 cm in all three cases. The dose is normalized at position 60 cm. For the standard height of 21 cm a measurement with Gafchromic™ film and the RTPS calculation are demonstrated
Dependency of the number of MU for a given dose value on the phantom height
| Stack height in cm | 11 | 13 | 15 | 17 | 19 | 21 | 23 | 25 | 27 | 29 |
| Normalized number of MU | 95 % | 95 % | 97 % | 98 % | 99 % | 100 % | 101 % | 103 % | 104 % | 106 % |
Fig. 3Transmittance of the lung shields as a function of the thickness of the MCP96 material. The dots represent the measured values, the line an exponential fit
Fig. 4Gamma evaluations of three Gafchromic films in the Alderson phantom (a head region, b thorax, c abdomen) compared to the RTPS calculations. Read pixels do not fulfill the gamma criterion (3 mm, 3 %). Below the corresponding CT slices
Fig. 5DVH of the Alderson phantom and a representative patient, containing the lungs as organs at risk and the combined head and trunk region
Average and standard deviation of the in-vivo dosimetry for 10 patients
| Forehead | Neck ventral | Chest midline | Reference ventral | Reference dorsal | Abdomen | Thigh ventral | Ankle | |
|---|---|---|---|---|---|---|---|---|
| Dose in Gy | 2.01 ± 0.06 | 2.00 ± 0.05 | 2.04 ± 0.09 | 2.11 ± 0.08 | 2.04 ± 0.06 | 2.10 ± 0.10 | 2.12 ± 0.17 | 2.04 ± 0.10 |