[Certification of an in-house manufactured translational couch unit for total body irradiation in accordance with the medical devices act].

Christian Scherf1, Petra Härtl, Marius Treutwein, Ludwig Bogner.   

Abstract

UNLABELLED: A new translational couch unit with extended potentials of dose optimization by variable velocity and a comfortable user interface with integrated patient administration was developed at the university clinic of Regensburg.
MATERIALS AND METHODS: The concept and construction were elaborated in legal accordance with the in-house manufacture conditions mentioned in the German Medical Devices Act. In particular we have implemented a concept of functional safety based on a controller unit, an independent monitoring unit and self-testing procedures. Redundant safety relevant hard- and software components are continuously checked against each other. In case of any malfunction the translation movement and the linear accelerator are stopped. Gap-free continuation of the treatment will be possible after elimination of the cause of the interrupt. RESULTS AND
CONCLUSION: After the testing of the implemented functional safety features including the risk assessment and management, electrical safety, electromagnetic compatibility by accredited technical experts the translational couch system complies with the requirements of the Medical Devices Act and can be operated in terms of in-house application. The presented certification procedure can motivate other radiation departments to develop own systems for in-house usage.

Entities:  

Mesh:

Year:  2007        PMID: 17665737     DOI: 10.1016/j.zemedi.2006.11.003

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  2 in total

1.  Three-dimensional printers applied for the production of beam blocks in total body irradiation treatment.

Authors:  Manuel Maerz; Marius Treutwein; Jan Nabo; Barbara Dobler
Journal:  J Appl Clin Med Phys       Date:  2022-03-15       Impact factor: 2.243

2.  Total body irradiation-an attachment free sweeping beam technique.

Authors:  Petra M Härtl; Marius Treutwein; Matthias G Hautmann; Manuel März; Fabian Pohl; Oliver Kölbl; Barbara Dobler
Journal:  Radiat Oncol       Date:  2016-06-10       Impact factor: 3.481