Shear modulus reconstruction by ultrasonically measured strain ratio.

Chikayoshi Sumi1, Hidenori Matsuzawa2.   

Abstract

PURPOSE: In addition to a description of our three previously developed one-dimensional (1D) methods from the viewpoint of shear modulus reconstruction using the strain ratio, two new methods for stabilizing the 1D methods are described, together with their limitations. As confirmed using human in vivo breast tissues, method 1 for evaluating the strain ratio itself is useful when the measurement accuracy of the strain distribution is high. However, because tissues having high shear moduli, such as scirrhous carcinoma, often form singular points/regions, both methods 2 and 3 using the strain ratio (initial estimate) and a regularization method are effective for realizing a unique, stable, useful shear modulus reconstruction. Because method 3 carries out implicit integration only at singular points/regions, whereas method 2 carries out implicit integration throughout the region of interest (ROI), the smaller number of singular points enables more rapid shear modulus reconstruction by method 3 than by method 2. Like method 1, method 3 is also useful when the measurement accuracy of the strain distribution is high. However, when evaluating strain distribution in an ROI with a high spatial resolution to obtain a shear modulus reconstruction having a high spatial resolution, shear modulus reconstructions obtained by methods 1, 2, and 3 often become laterally unstable due to the instability and low accuracy of the strains in the reference regions (reference strains), i.e., regularization in methods 2 and 3 cannot reduce the instability in the initial estimate.
METHODS: To cope with this instability, (i) the reconstruction obtained by calculating the strain ratio should be low-pass filtered; for breast tissues, in particular, the reconstruction of the inverse shear modulus should be low-pass filtered, not the reconstruction of the shear modulus. (ii) Otherwise, when using homogeneous regions as a reference, such as a block of reference material, fatty tissue, or parenchyma, evaluation of the reference strains with a low spatial resolution is effective.
RESULTS: Although such evaluation yields a stable reconstruction with a high spatial resolution compared with that obtained by the low-pass filtering of the strain ratio, we confirmed through simulations that, when reducing artifacts due to a 1D reconstruction of the shear modulus, the evaluation yields a low-accuracy reconstruction value of inhomogeneity. In contrast, in such a case the low-pass filtering of the strain ratio yields a more accurate reconstruction value.
CONCLUSION: All the above-mentioned methods using the strain ratio realize real-time shear modulus reconstruction and should be selected appropriately in conventional ultrasonic imaging equipment by considering the application of the reconstruction (i.e., in accordance with the measurement accuracy of the strains and the occurrence of artifacts).

Entities:  

Keywords:  human breast scirrhous carcinoma; one-dimensional shear modulus reconstruction; stabilization; strain ratio; ultrasound

Year:  2007        PMID: 27278479     DOI: 10.1007/s10396-007-0151-1

Source DB:  PubMed          Journal:  J Med Ultrason (2001)        ISSN: 1346-4523            Impact factor:   1.314


  30 in total

1.  Assessment of thermal tissue ablation with MR elastography.

Authors:  T Wu; J P Felmlee; J F Greenleaf; S J Riederer; R L Ehman
Journal:  Magn Reson Med       Date:  2001-01       Impact factor: 4.668

2.  Visualization and quantification of breast cancer biomechanical properties with magnetic resonance elastography.

Authors:  D B Plewes; J Bishop; A Samani; J Sciarretta
Journal:  Phys Med Biol       Date:  2000-06       Impact factor: 3.609

3.  Reconstructions of shear modulus, Poisson's ratio, and density using approximate mean normal stress lambda epsilon alpha alpha as unknown.

Authors:  Chikayoshi Sumi
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2006-12       Impact factor: 2.725

4.  Ultrasonic axial strain measurement for lateral tissue deformation.

Authors:  Chikayoshi Sumi
Journal:  Ultrasound Med Biol       Date:  2007-07-30       Impact factor: 2.998

5.  Effective shear modulus reconstruction obtained with approximate mean normal stress remaining unknown.

Authors:  Chikayoshi Sumi
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2007-11       Impact factor: 2.725

6.  Identifying the mechanical properties of tissue by ultrasound strain imaging.

Authors:  Emre Turgay; Septimiu Salcudean; Robert Rohling
Journal:  Ultrasound Med Biol       Date:  2006-02       Impact factor: 2.998

7.  A theoretical framework for performance characterization of elastography: the strain filter.

Authors:  T Varghese; J Ophir
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  1997       Impact factor: 2.725

8.  Elastography: a quantitative method for imaging the elasticity of biological tissues.

Authors:  J Ophir; I Céspedes; H Ponnekanti; Y Yazdi; X Li
Journal:  Ultrason Imaging       Date:  1991-04       Impact factor: 1.578

9.  In vivo elastographic investigation of ethanol-induced hepatic lesions.

Authors:  Kenneth Hoyt; Flemming Forsberg; Christopher R B Merritt; Ji-Bin Liu; Jonathan Ophir
Journal:  Ultrasound Med Biol       Date:  2005-05       Impact factor: 2.998

10.  Fundamental mechanical limitations on the visualization of elasticity contrast in elastography.

Authors:  H Ponnekanti; J Ophir; Y Huang; I Céspedes
Journal:  Ultrasound Med Biol       Date:  1995       Impact factor: 2.998

View more
  2 in total

1.  Determination of lateral modulation apodization functions using a regularized, weighted least squares estimation.

Authors:  Chikayoshi Sumi
Journal:  Int J Biomed Imaging       Date:  2010-05-09

2.  A Novel Nonlinear Parameter Estimation Method of Soft Tissues.

Authors:  Qianqian Tong; Zhiyong Yuan; Mianlun Zheng; Xiangyun Liao; Weixu Zhu; Guian Zhang
Journal:  Genomics Proteomics Bioinformatics       Date:  2017-12-13       Impact factor: 7.691

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.