Literature DB >> 27183061

Cone Beam Computed Tomography vs Multislice Computed Tomography in Computer-Aided Design/Computer-Assisted Manufacture Guided Implant Surgery Based on Three-Dimensional Optical Scanning and Stereolithographic Guides: Does Image Modality Matter?

Gerlig Widmann, Bastian Fischer, Johannes Peter Michael Berggren, Albina Dennhardt, Peter Schullian, Bale Reto, Wolfgang Puelacher.   

Abstract

PURPOSE: Cone beam computed tomography (CBCT) is increasingly overtaking multislice computed tomography (MSCT) for implant planning and computer-aided design/computer-assisted manufacturing (CAD/CAM) surgical guide fabrication. The purpose of this study was to evaluate whether the image modality has a significant influence on the accuracy of image-fusion stereolithographic guides.
MATERIALS AND METHODS: A total of 240 implants were placed in 30 polymer models using surgical guides fabricated using CBCT or MSCT and optical scanning of dental casts and diagnostic wax-up. Postsurgical image data were fused with the planning data for evaluation of the following errors: total error (Euclidean error), mesiodistal and buccolingual error, depth error, and angular error. The CBCT and MSCT results were statistically compared using a t test (P = .05).
RESULTS: CBCT showed a statistically significantly higher total error with mean (± SD) of 0.36 ± 0.13 mm vs 0.27 ± 0.13 mm (P = .000), mesiodistal error with 0.20 ± 0.14 mm vs 0.16 ± 0.11 mm (P = .018), and depth error with 0.17 ± 0.12 mm vs 0.07 ± 0.07 mm (P = .000). Buccolingual and angular errors did not statistically significantly differ with 0.16 ± 0.11 mm vs 0.15 ± 0.12 mm (P = .613), and 0.96 ± 0.47 degrees vs 0.85 ± 0.59 degrees (P = .111).
CONCLUSION: Using CBCT for image-fusion stereolithographic guides may provide lower accuracy than MSCT. The results should be confirmed on alternative CBCT scanners. Due to the total volume of tissues, image artifacts, and patient movements, the advantages seen using MSCT on models could be nonexistent or exacerbated on patients. Patient studies are required to demonstrate clinical relevance.

Entities:  

Mesh:

Year:  2016        PMID: 27183061     DOI: 10.11607/jomi.4222

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  3 in total

Review 1.  Advantages and limitations of implant surgery with CAD/CAM surgical guides: A literature review.

Authors:  Gokce-Soganci Unsal; Ilser Turkyilmaz; Samantha Lakhia
Journal:  J Clin Exp Dent       Date:  2020-04-01

2.  Comparison of Dental Panoramic Radiography and CBCT for Measuring Vertical Bone Height in Different Horizontal Locations of Posterior Mandibular Alveolar Process.

Authors:  Shoaleh Shahidi; Barbad Zamiri; Masoud Abolvardi; Marzieh Akhlaghian; Maryam Paknahad
Journal:  J Dent (Shiraz)       Date:  2018-06

3.  Primary oral manifestation of Langerhans cell histiocytosis refractory to conventional therapy but susceptible to BRAF-specific treatment: a case report and review of the literature.

Authors:  Norbert Neckel; Andrej Lissat; Arendt von Stackelberg; Nadine Thieme; Mohemed-Salim Doueiri; Birgit Spors; Benedicta Beck-Broichsitter; Max Heiland; Jan-Dirk Raguse
Journal:  Ther Adv Med Oncol       Date:  2019-10-19       Impact factor: 8.168

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.