Use of bioabsorbable staple reinforcement material in side-to-side anastomoses: Suture line reinforcement of the weak point of the anastomosis.

Masahiro Kimura1, Yukio Terashita1.   

Abstract

BACKGROUND: Few studies have been designed regarding optimal reinforcement of the crotch of a side-to-side anastomosis. The purpose of this study was to clarify the weak point of the side-to-side anastomosis and to evaluate the effect of bioabsorbable reinforcement material.
METHODS: Fresh pig small bowel was used for all experiments. A side-to-side anastomosis was performed using a linear stapler, and the burst pressure of the anastomosis was measured. Three separate experiments were done. In experiment 1, the weak point and the burst pressure of that point were defined. In experiment 2, the burst pressure of the side of the anastomosis was measured. In experiment 3, we evaluated the effect of Neoveil(®) to strengthen the weak point of the anastomosis.
RESULTS: The weak point of the side-to side anastomosis was the crotch and the burst pressure was 39.8 ± 5.7 mmHg. The burst pressure of the side of the anastomosis was 109.9 ± 7.9 mmHg. This was significantly higher than the burst pressure of the crotch (P = 0.008). The burst pressure of the crotch in the group with Neoveil(®) was 83.3 ± 14.9 mmHg. This pressure was significantly higher than the group with no Neoveil(®) reinforcement (P = 0.001).
CONCLUSION: These findings suggest that the use of Neoveil(®) as a buttressing material is associated with reinforced staple lines and increased crotch burst pressures compared to non-buttressed staple lines. Neoveil(®) was found to perform comparably to clinically available buttress materials in this ex vivo model. Reinforcement of the weak point of the side-to-side anastomosis with Neoveil(®)may lead to fewer anastomotic leaks.

Entities:  

Keywords:  Burst pressure; Crotch; Neoveil®; Reinforcement; Side-to-side anastomosis

Year:  2016        PMID: 27158487      PMCID: PMC4843102          DOI: 10.1016/j.amsu.2016.01.034

Source DB:  PubMed          Journal:  Ann Med Surg (Lond)        ISSN: 2049-0801


Introduction

There are many postoperative complications after gastrointestinal anastomosis, including leak, bleeding, and stenosis [1], [2]. Of these, anastomotic leak is the most dreaded complication and can be life threatening. Many techniques have been developed for use in the anastomosis of an intestinal tract aimed at minimizing postoperative complications [1]. With the development of stapling devices, the variation of anastomosis techniques between surgeons has decreased, as have postoperative complications. The growth of laparoscopic surgery has further accelerated the frequency of the use of automatic suture instruments. The side-to-side anastomosis is widely used given the ease by which this is fashioned [2], [3]. A side-to-side anastomosis is performed using linear staplers. The weak point of this anastomosis is the crotch area. The arrangement of the suture device leads to three rows of staples, making a robust anastomosis that is clinically sufficient to prevent leakage of intestinal contents, except for in the crotch area. Surgeons therefore typically reinforce the crotch with a hand thrown stitch [3]. However, laparoscopically, reinforcement is not easy. In thoracic surgery, there are reports that Neoveil® reduces air leak after wedge resections [4]. Therefore, we sought to examine whether Neoveil®reinforcement of the crotch of the side-to-side anastomosis strengthened this area and made it less prone to leak.

Material and methods

Fresh pig small bowel was used for all experiments. The specimens were obtained from an animal that had been sacrificed for use in approved non-gastrointestinal research studies. The specimens were used within 24 h after sacrifice. Each segment of the intestinal tract was 20 cm in length. A side-to-side anastomosis was performed between two intestine specimens using a linear stapler. The stapler was inserted from the edge of each segment of intestinal tract. A 16-Fr catheter was then placed into the lumen through one intestinal wall. Each side of the anastomosis was clamped by forceps. The sphygmomanometer and tubing for instillation of air was connected, and the anastomosis was submerged in water. Air was then blown into the intestinal lumen with a syringe (Fig. 1).
Fig. 1

System used to examine burst pressure. (a: Experiment 1, b: Experiment 2). Sphy: sphygmomanometer; Sy: syringe.

The burst pressure of the anastomosis was indicated by the presence of bubbles. All procedures were performed by the same surgeon. The stapling device used was the Endo GIA60AMT (Covidien, Tokyo, Japan). Suture length was 6 cm and the number of staplers fired was 90. These staplers consist of three rows of either 3/3.5/4 mm staples prior to firing. Post-fire heights are 1.25/1.5/1.75 mm.

Experiment 1: weak point identification and burst pressure of a side-to-side anastomosis

After creation of the anastomosis, the stump of each end of the small intestine and the side of the anastomosis were clamped by forceps. The burst pressure of the crotch was recorded. Five anastomoses were completed.

Experiment 2: burst pressure of the side of the anastomosis

Experiment 2 was done to assess the strength of the side of the side-to-side anastomosis. The crotch was clamped and the side burst pressure recorded. Five anastomoses were completed.

Experiment 3: effect of buttress on the crotch with Neoveil®

Experiment 3 was designed to evaluate the effect of Neoveil®(Gunze) on the crotch. For that purpose, tube type Neoveil® was attached to the stapling device. One side of tube type Neoveil® is PGA (polyglycoic acid) sheet, and another side is expandable nylon mesh sheet. These are fixed with 3-0 vucryl. To reinforce only the crotch of the anastomosis, the Neoveil® was cut to a length of 1.5 cm (Fig. 3a and b). As we could not deny that the reinforcement effect of Neoveil® may just be that it makes the anastomosis thicker, we added an experiment to sandwich Neoveil® between intestinal tracts and then perform an anastomoses (Fig. 2, Fig. 3c,d). Five anastomoses in each group were completed.
Fig. 3

a: tube-type Neoveil® was attached to the stapling devices, b: side-to side anastomosis with stapler (with Neoveil®), c: intestinal lumen after the stapling with Neoveil®, d: sandwich Neoveil® between intestinal tracts.

Fig. 2

Schema of experiment 3. (A: without Neoveil®, B: with Neoveil® attached to the stapling devices, C: Neoveil® sandwiched between intestinal tracts).

Statistical analysis

Discrete variables were analyzed by the Mann–Whitney test and significance was indicated at p < 0.05.

Results

Experiment 1: weak point identification and burst pressure of a side-to side anastomosis

All bursting points were at the crotch. Burst pressures were 39.8 ± 5.7 mmHg (Table 1).
Table 1

Burst pressure of side-to-side anastomosis.

Pressure (mmHg)Leak point
39.8 ± 5.7Crotch
The burst pressures of the side were 109.9 ± 7.9 mmHg (Table 2). These were significantly higher than those of the crotch (P = 0.008).
Table 2

The burst pressure of side to side anastomosis (bursting pressure of the side).

Pressure (mmHg)
109.9 ± 7.9
All burst points were in the crotch. The burst pressures in the buttress group with Neoveil® were 83.3 ± 14.9 mmHg. The pressures were significantly higher than the group with no Neoveil (P = 0.001). On the other hand, in the group in which Neoveil® was sandwiched outside the intestinal tract, burst pressures were 39 ± 8.6 mmHg, similar to the group without Neoveil crotch reinforcement (Table 3).
Table 3

The bursting pressure of side to side anastomosis (with and without Neoveil®).

Pressure (mmHg)Leak pointP value
A: Neoveil (−)39.8 ± 5.7CrotchA vs B 0.0012A vs C 0.87
B: Neoveil® (+) inside83.3 ± 14.9CrotchB vs C 0.0017
C: Neoveil® (+) outside39 ± 8.6Crotch

Discussion

Creation of a gastrointestinal tract anastomosis is a fundamental and important surgical procedure. To reduce postoperative complications, such as bleeding and leak, surgical stapling instruments are being increasingly used for gastrointestinal anastomoses not only in laparoscopic surgery but also in open surgery. In the 1960s, Steichen and Ravitch first introduced stapling instruments [5]. Improvements in stapling instruments have been ongoing for half a century [6], [7], [8], [9]. For example, the suture device that included two suture lines on one side was improved to a suture device with three suture lines on one side, and suture strength has significantly increased. We performed a burst pressure experiment using the GIA™60-3.8 (2 line) and the Endo GIA™60-3.5 (three lines), and the burst pressure increased by approximately two-fold. With the development of a variety of instruments, various anastomosis methods have also been developed. Mechanical side-to-side anastomoses are the basis of most methods. Even in open surgery, mechanical side-to-side anastomoses are easy, quick, and cost-effective because these can be performed using a single linear stapling device, while an end-to-side reconstruction requires both linear and circular staplers, thus incurring additional cost. Despite these advances, stapled anastomoses have inherent weak points. The present study was designed to investigate the mechanical strength of the side-to-side anastomosis by measuring the leak pressure detected on an ex vivo porcine small bowel model. Experiment 1 identified the weak point of the side-to-side anastomosis to be the crotch, with a mean burst pressure of 93.8 mmHg. We previously performed an experiment to measure the strength of a side-to-side anastomosis using the esophagus and the small intestine of a pig, and identified the weak point to be the crotch when using a three-row stapler device. Another study before that showed that, with a two-row suture device, leaks occurred with equal frequency between the crotch and the side of the anastomosis. Thus, with the more widespread use of three-row stapling devices, reinforcement of the crotch has become important. There were no experiments, however, quantifying the burst strength difference between the side and the crotch. In experiment 2, we found that the mean burst pressure of the side was 109.9 mmHg, approximately 3 times that of the crotch. Clinically, the weakness of the crotch is widely recognized. As shown in Fig. 5, it is common to reinforce the crotch. Goto reported the effectiveness of buttressing of the crotch in a functional end-to-end anastomosis using pig intestine [3]. They reported that the burst pressure rose to 44 from 27.5 mmHg with a 3-0 silk suture to buttress. However, in laparoscopic surgery, crotch reinforcement with a suture is technically difficulty. For example, the crotch of the side-to-side anastomosis after total gastrectomy, and the crotch of the functional end-to-end anastomosis after large intestine excision, are difficult to reinforce. Therefore we looked for a method to reinforce the crotch in a manner that did not require a suture.
Fig. 5

Operative findings. (a: crotch, b: reinforcement suture).

One such possibility is the use of buttress reinforcement materials. Buttress reinforcement materials are comprised of various synthetic polymers and biologically derived materials [10], [11], [12], [13], [14], [15], [16], [17]. Examples of clinically available buttress reinforcement materials are bovine pericardium (Peri-strips), expanded-poly-tetrafluoroethylene carbonate (TMC), copolymer (Gore Seamguard), urinary bladder matrix (MatriStem), and small intestinal submucosa (Surgisis). These products have demonstrated some success in reducing leakage and bleeding complications associated with staple lines. Application of reinforcement material in the staple line is thought to moderate tension of the staple line because it acts as a neutralization plate. Further, the buttressing materials seal off the staple holes and narrow the spaces in between each staple. Thus, leakage, bleeding, and tearing at the staple line can be reduced, especially in diseased and fragile tissue. However, choice of material should be considered carefully. Although all types of staple line reinforcement seem equally adequate in reducing complications at the staple line, the material itself can cause problems. For example, bovine pericardium material is prone to erosion and migration due to inflammatory response and carries the risk of animal source contamination. Neoveil® (polyglycolic acid felt: PGAF) has been described and tested in many clinical situations. Neoveil® is an absorbable reinforcement material, and its advantages have been widely recognized. PGAF is a soft and thin (0.15 mm depth) new absorbable material. In thoracic surgery, Neoveil® reduces and prevents air leakage from the lung. In addition, Neoveil® has been used for reinforcement of a suture line on the trachea. In digestive surgery, Neoveil® has been used for fragile organs such as liver and the gastrointestinal tract. Other than Neoveil®, there has been a report using BioGlue. Nandakumar reported the effectiveness of BioGlue® for the reinforcement of the gastrojejunostomy in obesity surgery [18]. They reported that the burst pressure significantly increased in the reinforced gastrojejunostomies from 27.4 to 59.1 mmHg. However, a more consistent method is needed as it is hard to apply BioGlue uniformly on an anastomosis. There are some reports describing Neoveil® for gastrointestinal enterotomy closure of the small intestine and the stomach, but there are no reports that used Neoveil® for intestinal side-to-side anastomosis. Neoveil® is thin and soft, but has memory when applied to a stapling device. However, when we use Neoveil® for the full length of the anastomosis, closure of the enterotomy after staple firing becomes difficult (Fig. 4). As highlighted in our experiment, though, only the crotch needs be reinforced as opposed to the entire staple line.
Fig. 4

Neoveil® for the full length of the anastomosis (arrow: protruding Neoveil®).

While not part of its intended use, we inserted Neoveil® between intestinal tracts to be anastomosed and performed a burst pressure examination to show that the increase in tissue thickness did not affect pressure as an independent variable. There were no significant differences between the use or non-use of Neoveil®, and thus proved that the increase in burst strength of the crotch was not due to an increase in the tissue thickness. Anastomotic failure depends on various parameters including tissue thickness, collagen content, blood flow, type of staple cartridge, ischemia, and tension [19]. It is not clear how burst pressure affects leak rate. However, clinically, many surgeons have experienced that the crotch in the anastomosis using a linear stapler is weak. We have shown that the burst pressure of the crotch can be nearly doubled with Neoveil® reinforcement. Laparoscopic operations and anastomoses with stapler devices are increasing. Out findings suggest that the use of Neoveil® as a buttressing material reinforces the staple line and increases burst pressure of the crotch when compared to a non-buttressed staple line. Neoveil® was found to perform comparably with clinically available buttress materials in this ex vivo model.

Ethical approval

None.

Funding

None.

Author contribution

Masahiro Kimura: study design, data collection, data analysis, writing. Yukio Terashita: data collection.

Conflicts of interest

None.

Guarantor

Masahiro Kimura.
  17 in total

Review 1.  The science of stapling and leaks.

Authors:  Randal S Baker; James Foote; Paul Kemmeter; Randall Brady; Todd Vroegop; Matt Serveld
Journal:  Obes Surg       Date:  2004 Nov-Dec       Impact factor: 4.129

2.  Experimental evaluation of the mechanical strength of stapling techniques.

Authors:  Kentaro Kawasaki; Yasuhiro Fujino; Kiyonori Kanemitsu; Tadahiro Goto; Takashi Kamigaki; Daisuke Kuroda; Yoshikazu Kuroda
Journal:  Surg Endosc       Date:  2007-03-13       Impact factor: 4.584

3.  Evaluation of the mechanical strength and patency of functional end-to-end anastomoses.

Authors:  T Goto; K Kawasaki; Y Fujino; K Kanemitsu; T Kamigaki; D Kuroda; Y Suzuki; Y Kuroda
Journal:  Surg Endosc       Date:  2007-02-07       Impact factor: 4.584

4.  Profiling surgical staplers: effect of staple height, buttress, and overlap on staple line failure.

Authors:  Carlos M Mery; Bilal M Shafi; Gary Binyamin; John M Morton; Michael Gertner
Journal:  Surg Obes Relat Dis       Date:  2008-01-28       Impact factor: 4.734

5.  The use of Seamguard to buttress the suture repair of a staple line leak following laparoscopic gastric bypass for obesity.

Authors:  N N Basu; D Leschinskey; D I Heath
Journal:  Obes Surg       Date:  2008-05-06       Impact factor: 4.129

6.  End-to-end cervical esophagogastric anastomoses are associated with a higher number of strictures compared with end-to-side anastomoses.

Authors:  Leonie Haverkamp; Pieter C van der Sluis; Roy J J Verhage; Peter D Siersema; Jelle P Ruurda; Richard van Hillegersberg
Journal:  J Gastrointest Surg       Date:  2013-02-12       Impact factor: 3.452

7.  The use of staplers in anatomical side-to-side and functional end-to-end enteroanastomoses.

Authors:  F M Steichen
Journal:  Surgery       Date:  1968-11       Impact factor: 3.982

8.  Comparison of laparoscopic linear staplers in clinical practice.

Authors:  Steven C Simper; Joanna M Erzinger; Sherman C Smith
Journal:  Surg Obes Relat Dis       Date:  2007-04-02       Impact factor: 4.734

9.  A new bioabsorbable sleeve for staple-line reinforcement: report of a clinical experience.

Authors:  Yukihito Saito; Hideyasu Omiya; Yuzo Shomura; Ken-ichiro Minami; Hiroji Imamura
Journal:  Surg Today       Date:  2002       Impact factor: 2.549

10.  Use of bioabsorbable staple reinforcement material in gastric bypass: a prospective randomized clinical trial.

Authors:  Karl A Miller; Antonia Pump
Journal:  Surg Obes Relat Dis       Date:  2007-06-12       Impact factor: 4.734

View more
  4 in total

1.  A new technology for reducing anastomotic fistula in the neck after esophageal cancer surgery.

Authors:  Ya-Nan Song; Yu Qi; Chun-Yang Zhang; Yin-Liang Sheng; Kai Wu; Sen-Lin Zhu; Lu Han; Ting-Ting Shan; Guan-Chao Ye; Qing-Yi Zhang; Yan-Li Chen; Jin-Wei Chen; Ya-Fei Liu; Lu-Bing Gao; Yang Yang; Zhan-Feng He; Deng-Yan Zhu; Dong-Lei Liu; Feng-Biao Wen; Tian-Liang Zheng; Ji-Lun Li; Song Zhao
Journal:  J Thorac Dis       Date:  2019-07       Impact factor: 2.895

2.  Sutureless functional end-to-end anastomosis using a linear stapler with polyglycolic acid felt for intestinal anastomoses.

Authors:  Masanori Naito; Hirohisa Miura; Takatoshi Nakamura; Takeo Sato; Takahiro Yamanashi; Atsuko Tsutsui; Masahiko Watanabe
Journal:  Ann Med Surg (Lond)       Date:  2017-03-31

3.  Prevention of anastomotic leakage using a polyglycolic acid sheet in double-stapling technique anastomosis for rectal surgery.

Authors:  Masatsune Shibutani; Hisashi Nagahara; Tatsunari Fukuoka; Yasuhito Iseki; Yuki Okazaki; Kosei Hirakawa; Masaichi Ohira
Journal:  Ann Med Surg (Lond)       Date:  2021-11-24

4.  Effect of Biologic Material Reinforcement on Surgical Anastomosis After Gastrectomy-A Pilot Study.

Authors:  Won Jun Kim; Chang Min Lee; Liang An; Jong-Han Kim; Sungsoo Park
Journal:  Front Oncol       Date:  2019-11-06       Impact factor: 6.244

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.