Literature DB >> 27147372

Novel laser-processed CsI:Tl detector for SPECT.

H Sabet1, L Bläckberg2, D Uzun-Ozsahin1, G El-Fakhri1.   

Abstract

PURPOSE: The aim of this work is to demonstrate the feasibility of a novel technique for fabrication of high spatial resolution CsI:Tl scintillation detectors for single photon emission computed tomography systems.
METHODS: The scintillators are fabricated using laser-induced optical barriers technique to create optical microstructures (or optical barriers) inside the CsI:Tl crystal bulk. The laser-processed CsI:Tl crystals are 3, 5, and 10 mm in thickness. In this work, the authors focus on the simplest pattern of optical barriers in that the barriers are created in the crystal bulk to form pixel-like patterns resembling mechanically pixelated scintillators. The monolithic CsI:Tl scintillator samples are fabricated with optical barrier patterns with 1.0 × 1.0 mm(2) and 0.625 × 0.625 mm(2) pixels. Experiments were conducted to characterize the fabricated arrays in terms of pixel separation and energy resolution. A 4 × 4 array of multipixel photon counter was used to collect the scintillation light in all the experiments.
RESULTS: The process yield for fabricating the CsI:Tl arrays is 100% with processing time under 50 min. From the flood maps of the fabricated detectors exposed to 122 keV gammas, peak-to-valley (P/V) ratios of greater than 2.3 are calculated. The P/V values suggest that regardless of the crystal thickness, the pixels can be resolved.
CONCLUSIONS: The results suggest that optical barriers can be considered as a robust alternative to mechanically pixelated arrays and can provide high spatial resolution while maintaining the sensitivity in a high-throughput and cost-effective manner.

Mesh:

Year:  2016        PMID: 27147372      PMCID: PMC5148208          DOI: 10.1118/1.4947294

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  MRC-SPECT: A sub-500 μm resolution MR-compatible SPECT system for simultaneous dual-modality study of small animals.

Authors:  Liang Cai; Xiaochun Lai; Zengming Shen; Chin-Tu Chen; Ling-Jian Meng
Journal:  Nucl Instrum Methods Phys Res A       Date:  2014-01-11       Impact factor: 1.455

2.  A novel, SiPM-array-based, monolithic scintillator detector for PET.

Authors:  Dennis R Schaart; Herman T van Dam; Stefan Seifert; Ruud Vinke; Peter Dendooven; Herbert Löhner; Freek J Beekman
Journal:  Phys Med Biol       Date:  2009-05-15       Impact factor: 3.609

3.  Design of an efficient position sensitive gamma ray detector for nuclear medicine.

Authors:  J G Rogers; D P Saylor; R Harrop; X G Yao; C V Leitao; B D Pate
Journal:  Phys Med Biol       Date:  1986-10       Impact factor: 3.609

4.  Performance of a position-sensitive scintillation detector.

Authors:  J S Karp; G Muehllehner
Journal:  Phys Med Biol       Date:  1985-07       Impact factor: 3.609

5.  Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: first validation versus invasive coronary angiography.

Authors:  Michael Fiechter; Jelena R Ghadri; Silke M Kuest; Aju P Pazhenkottil; Mathias Wolfrum; Rene N Nkoulou; Robert Goetti; Oliver Gaemperli; Philipp A Kaufmann
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-07-15       Impact factor: 9.236

6.  Combining Surface Treatments with Shallow Slots to Improve the Spatial Resolution Performance of Continuous, Thick LYSO Detectors for PET.

Authors:  M Kaul; S Surti; J S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2013-02-01       Impact factor: 1.679

7.  Light-Sharing Interface for dMiCE Detectors Using Sub-Surface Laser Engraving.

Authors:  William C J Hunter; Robert S Miyaoka; Lawrence MacDonald; Wendy McDougald; Thomas K Lewellen
Journal:  IEEE Trans Nucl Sci       Date:  2015-02-06       Impact factor: 1.679

  7 in total
  3 in total

1.  Characterization or Monolithic Scintillation Detectors Etched with Laser Induced Optical Barriers.

Authors:  J V Panetta; S Surti; B Singh; J S Karp
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2018-10-10

2.  Simulation study of light transport in laser-processed LYSO:Ce detectors with single-side readout.

Authors:  L Bläckberg; G El Fakhri; H Sabet
Journal:  Phys Med Biol       Date:  2017-10-19       Impact factor: 3.609

3.  An edge-readout, multilayer detector for positron emission tomography.

Authors:  Xin Li; Maria Ruiz-Gonzalez; Lars R Furenlid
Journal:  Med Phys       Date:  2018-05-06       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.