Literature DB >> 27132802

Influence of exposure parameters on the detection of simulated root fractures in the presence of various intracanal materials.

M G O Pinto1, K A Rabelo1, S L Sousa Melo2, P S F Campos3, L S A F Oliveira4, P M Bento1, D P Melo1.   

Abstract

AIM: To investigate how the exposure parameters used when producing CBCT scans affect diagnostic accuracy when detecting simulated vertical root fractures and how various filling materials affect image quality.
METHODOLOGY: A total of 160 extracted single-rooted human teeth were divided into four groups based on the materials within the root canals: unrestored, gutta-percha, metallic post and fibreglass post. Half of the sample of each group was selected for the induction of vertical root fractures. Each tooth was placed in an empty socket of a dry skull which was scanned on a CBCT unit (9000 3D scanner; Kodak Dental Systems, Carestream Health, Rochester, NY, USA) using the following exposure parameters: 74 kV/12 mA; 74 kV/10 mA; 74 kV/8 mA; 74 kV/6.3 mA; 70 kV/12 mA; 70 kV/10 mA; 70 kV/8 mA; 70 kV/6.3 mA. Two observers assessed all images using a 5-point confidence scale for fracture detection and a 4-point score for the presence of artefacts. Sensitivity, specificity, accuracy and area under ROC curve were compared by two-way anova and Tukey's test. Artefact formation was evaluated by descriptive statistics.
RESULTS: There were no significant differences in the sensitivity (P = 0.370), specificity (P = 0.660), accuracy (P = 0.084) and area under the ROC curve (P = 0.674) values amongst the various exposure parameters within the same group. There were some significant differences when the groups were compared for each intracanal material. The sensitivity and accuracy of the group containing metallic posts were significantly lower than the unrestored and fibreglass post groups (P ≤ 0.017).
CONCLUSION: The variations in exposure parameters did not interfere with the diagnosis of vertical root fractures, independent of the root canal restorative status. Metallic posts were associated with greater artefact formation and compromised the diagnostic performance. It is possible to decrease the kVp/mA settings to reduce the probability of biological effects due to radiation, without losing diagnostic accuracy.
© 2016 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  artefacts; cone beam computed tomography; vertical root fracture

Mesh:

Substances:

Year:  2016        PMID: 27132802     DOI: 10.1111/iej.12655

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  12 in total

1.  Application of image processing techniques to aid in the detection of vertical root fractures in digital periapical radiography.

Authors:  Lucas Exposto Soares; Deborah Queiroz Freitas; Kaique Leite de Lima; Lorena Rosa Silva; Fernanda Paula Yamamoto-Silva; Marcelo Andrade da Costa Vieira
Journal:  Clin Oral Investig       Date:  2021-02-05       Impact factor: 3.573

2.  Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts.

Authors:  Deborah Queiroz Freitas; Rocharles Cavalcante Fontenele; Eduarda Helena Leandro Nascimento; Taruska Ventorini Vasconcelos; Marcel Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2018-06-27       Impact factor: 2.419

3.  Influence of sharpening filters on the detection of root fractures using low-dose cone-beam computed tomography.

Authors:  Alexandra Robles González; Guilherme Monteiro Tosoni; Deborah Queiroz Freitas; Matheus L Oliveira
Journal:  Clin Oral Investig       Date:  2022-03-15       Impact factor: 3.606

4.  The detection of vertical root fractures in post-core restored teeth with cone-beam CT: in vivo and ex vivo.

Authors:  Svetlana F Byakova; Nina E Novozhilova; Irina M Makeeva; Vasiliy I Grachev; Inna V Kasatkina
Journal:  Dentomaxillofac Radiol       Date:  2019-05-24       Impact factor: 2.419

5.  CBCT image artefacts generated by implants located inside the field of view or in the exomass.

Authors:  Husniye Demirturk Kocasarac; Lisa J Koenig; Gulbahar Ustaoglu; Matheus Lima Oliveira; Deborah Queiroz Freitas
Journal:  Dentomaxillofac Radiol       Date:  2021-07-29       Impact factor: 2.419

6.  Accuracy of three cone-beam CT devices and two software systems in the detection of vertical root fractures.

Authors:  Aline Pf Caetano; Thiago O Sousa; Mariana R Oliveira; Karine Evanglista; Juliano M Bueno; Maria Ag Silva
Journal:  Dentomaxillofac Radiol       Date:  2020-12-01       Impact factor: 2.419

7.  Does anthropomorphic model design in ex vivo studies affect diagnostic accuracy for dental root fracture using CBCT?

Authors:  Fedil Andraws Yalda; Rosalyn J Clarkson; Jonathan Davies; Peter G J Rout; Anita Sengupta; Keith Horner
Journal:  Dentomaxillofac Radiol       Date:  2020-06-09       Impact factor: 2.419

8.  Influence of size of field of view (FOV), position within the FOV, and scanning mode on the detection of root fracture and observer's perception of artifacts in CBCT images.

Authors:  Martina Gerlane de Oliveira Pinto; Saulo L Sousa Melo; Fernanda Clotilde Mariz Suassuna; Luiz Eduardo Marinho; José Bruno da Silva Leite; Andre Ulisses Dantas Batista; Patrícia Meira Bento; Daniela Pita Melo
Journal:  Dentomaxillofac Radiol       Date:  2021-02-10       Impact factor: 3.525

9.  Effects of various cone-beam computed tomography settings on the detection of recurrent caries under restorations in extracted primary teeth.

Authors:  Kıvanç Kamburoğlu; Gül Sönmez; Zeynep Serap Berktaş; Hakan Kurt; Doĝukan Özen
Journal:  Imaging Sci Dent       Date:  2017-06-22

10.  Quantitative assessment of image artifacts from root filling materials on CBCT scans made using several exposure parameters.

Authors:  Katharina Alves Rabelo; Yuri Wanderley Cavalcanti; Martina Gerlane de Oliveira Pinto; Saulo Leonardo Sousa Melo; Paulo Sérgio Flores Campos; Luciana Soares de Andrade Freitas Oliveira; Daniela Pita de Melo
Journal:  Imaging Sci Dent       Date:  2017-09-21
View more