Literature DB >> 27122925

Accordion Stent Deformation upon Retrieval of a Side-Branch Protective Guidewire.

Shih-Tsung Cheng1, Cheng-Wen Su2, Chiau-Suong Liau1, Yu-Lin Ko1.   

Abstract

UNLABELLED: In this case we herein report a dangerous complication from primary percutaneous coronary intervention, where an unnoticed loop of the guidewire was inadvertently made around the stent during provisional stenting. Since the guidewire and the stent were entangled, efforts to retrieve the guidewire only exacerbated the problem by compressing the stent like an accordion. We review those factors that may have influenced stent compression in our case, as well as possible ways to avoid it from occurring in the future. KEY WORDS: Catheterization; Coronary stenosis; Embolism; Myocardial infarction; Percutaneous coronary intervention; Stents.

Entities:  

Year:  2015        PMID: 27122925      PMCID: PMC4804985          DOI: 10.6515/acs20141201d

Source DB:  PubMed          Journal:  Acta Cardiol Sin        ISSN: 1011-6842            Impact factor:   2.672


  8 in total

1.  Stent longitudinal integrity bench insights into a clinical problem.

Authors:  John A Ormiston; Bruce Webber; Mark W I Webster
Journal:  JACC Cardiovasc Interv       Date:  2011-11-30       Impact factor: 11.195

2.  Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms.

Authors:  Paul D Williams; Mamas A Mamas; Kenneth P Morgan; Magdi El-Omar; Bernard Clarke; Anthony Bainbridge; Farzin Fath-Ordoubadi; Douglas G Fraser
Journal:  EuroIntervention       Date:  2012-06-20       Impact factor: 6.534

3.  Longitudinal stent deformation: insights on mechanisms, treatments and outcomes from the Food and Drug Administration Manufacturer and User Facility Device Experience database.

Authors:  Mamas A Mamas; Paul D Williams
Journal:  EuroIntervention       Date:  2012-06-20       Impact factor: 6.534

4.  Delayed surgical retrieval of retained guidewire following percutaneous coronary intervention.

Authors:  Amit Modi; Aleksejus Zorinas; Hunaid A Vohra; Markku Kaarne
Journal:  J Card Surg       Date:  2010-10-11       Impact factor: 1.620

5.  Unintended stent extraction from a coronary artery during bifurcation coronary angioplasty.

Authors:  Ivo Varvarovsky; Jan Matejka
Journal:  J Invasive Cardiol       Date:  2007-11       Impact factor: 2.022

6.  Entrapped devices after PCI.

Authors:  Ihab Alomari; Richard Snider; Sonia Ponce; Bina Ahmed
Journal:  Cardiovasc Revasc Med       Date:  2013-10-21

7.  Side-branch wire entrapment during bifurcation PCI: avoidance and management.

Authors:  Andrew T Burns; Jack Gutman; Rob Whitbourn
Journal:  Catheter Cardiovasc Interv       Date:  2010-02-15       Impact factor: 2.692

8.  Clinical and angiographic outcomes of patients undergoing entrapped guidewire retrieval in stent-jailed side branch using a balloon catheter.

Authors:  Shingo Sakamoto; Norimasa Taniguchi; Yukio Mizuguchi; Takeshi Yamada; Shunsuke Nakajima; Tetsuya Hata; Akihiko Takahashi
Journal:  Catheter Cardiovasc Interv       Date:  2014-01-29       Impact factor: 2.692

  8 in total
  1 in total

1.  Longitudinal Stent Deformation in Undeployed Stents: A Case Series.

Authors:  Parminder Singh Otaal; Dinakar Bootla
Journal:  J Tehran Heart Cent       Date:  2021-07
  1 in total