Clinical cone beam computed tomography compared to high-resolution peripheral computed tomography in the assessment of distal radius bone.

C de Charry1,2, S Boutroy3, R Ellouz3, F Duboeuf3, R Chapurlat1,3,4, H Follet3, J B Pialat5,6,7.   

Abstract

UNLABELLED: Clinical cone beam computed tomography (CBCT) was compared to high-resolution peripheral quantitative computed tomography (HR-pQCT) for the assessment of ex vivo radii. Strong correlations were found for geometry, volumetric density, and trabecular structure. Using CBCT, bone architecture assessment was feasible but compared to HR-pQCT, trabecular parameters were overestimated whereas cortical ones were underestimated.
INTRODUCTION: HR-pQCT is the most widely used technique to assess bone microarchitecture in vivo. Yet, this technology has been only applicable at peripheral sites, in only few research centers. Clinical CBCT is more widely available but quantitative assessment of the bone structure is usually not performed. We aimed to compare the assessment of bone structure with CBCT (NewTom 5G, QR, Verona, Italy) and HR-pQCT (XtremeCT, Scanco Medical AG, Brüttisellen, Switzerland).
METHODS: Twenty-four distal radius specimens were scanned with these two devices with a reconstructed voxel size of 75 μm for Newtom 5G and 82 μm for XtremeCT, respectively. A rescaling-registration scheme was used to define the common volume of interest. Cortical and trabecular compartments were separated using a semiautomated double contouring method. Density and microstructure were assessed with the HR-pQCT software on both modality images.
RESULTS: Strong correlations were found for geometry parameters (r = 0.98-0.99), volumetric density (r = 0.91-0.99), and trabecular structure (r = 0.94-0.99), all p < 0.001. Correlations were lower for cortical microstructure (r = 0.80-0.89), p < 0.001. However, absolute differences were observed between modalities for all parameters, with an overestimation of the trabecular structure (trabecular number, 1.62 ± 0.37 vs. 1.47 ± 0.36 mm(-1)) and an underestimation of the cortical microstructure (cortical porosity, 3.3 ± 1.3 vs. 4.4 ± 1.4 %) assessed on CBCT images compared to HR-pQCT images.
CONCLUSIONS: Clinical CBCT devices are able to analyze large portions of distal bones with good spatial resolution and limited irradiation. However, compared to dedicated HR-pQCT, the assessment of microarchitecture by NewTom 5G dental CBCT showed some discrepancies, for density measurements mainly. Further technical developments are required to reach optimal assessment of bone characteristics.

Keywords:  3D registration; Bone microarchitecture; CBCT; HR-pQCT

Mesh:

Year:  2016        PMID: 27121345     DOI: 10.1007/s00198-016-3609-4

Source DB:  PubMed          Journal:  Osteoporos Int        ISSN: 0937-941X            Impact factor:   4.507


  40 in total

1.  Visual grading of motion induced image degradation in high resolution peripheral computed tomography: impact of image quality on measures of bone density and micro-architecture.

Authors:  J B Pialat; A J Burghardt; M Sode; T M Link; S Majumdar
Journal:  Bone       Date:  2011-10-13       Impact factor: 4.398

2.  Dose reduction in maxillofacial imaging using low dose Cone Beam CT.

Authors:  Kostas Tsiklakis; Catherine Donta; Sophia Gavala; Kety Karayianni; Vasiliki Kamenopoulou; Costas J Hourdakis
Journal:  Eur J Radiol       Date:  2005-06-22       Impact factor: 3.528

3.  A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results.

Authors:  P Mozzo; C Procacci; A Tacconi; P T Martini; I A Andreis
Journal:  Eur Radiol       Date:  1998       Impact factor: 5.315

4.  Analysis of intensity variability in multislice and cone beam computed tomography.

Authors:  Olivia Nackaerts; Frederik Maes; Hua Yan; Paulo Couto Souza; Ruben Pauwels; Reinhilde Jacobs
Journal:  Clin Oral Implants Res       Date:  2011-01-18       Impact factor: 5.977

5.  Accuracy of trabecular structure by HR-pQCT compared to gold standard μCT in the radius and tibia of patients with osteoporosis and long-term bisphosphonate therapy.

Authors:  M Krause; O Museyko; S Breer; B Wulff; C Duckstein; E Vettorazzi; C Glueer; K Püschel; K Engelke; M Amling
Journal:  Osteoporos Int       Date:  2014-02-25       Impact factor: 4.507

6.  Effect of voxel size on the accuracy of 3D reconstructions with cone beam CT.

Authors:  D Maret; N Telmon; O A Peters; B Lepage; J Treil; J M Inglèse; A Peyre; J L Kahn; M Sixou
Journal:  Dentomaxillofac Radiol       Date:  2012-12       Impact factor: 2.419

7.  Risk factors for proximal humerus, forearm, and wrist fractures in elderly men and women: the Dubbo Osteoporosis Epidemiology Study.

Authors:  T V Nguyen; J R Center; P N Sambrook; J A Eisman
Journal:  Am J Epidemiol       Date:  2001-03-15       Impact factor: 4.897

8.  CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT).

Authors:  Seppo K Koskinen; Ville V Haapamäki; Jari Salo; Nina C Lindfors; Mika Kortesniemi; Lauri Seppälä; Kimmo T Mattila
Journal:  Skeletal Radiol       Date:  2012-09-19       Impact factor: 2.199

9.  In vivo high resolution 3D-QCT of the human forearm.

Authors:  A Laib; H J Häuselmann; P Rüegsegger
Journal:  Technol Health Care       Date:  1998-12       Impact factor: 1.285

10.  Cortical bone in the human femoral neck: three-dimensional appearance and porosity using synchrotron radiation.

Authors:  Valérie Bousson; Françoise Peyrin; Catherine Bergot; Marc Hausard; Alain Sautet; Jean-Denis Laredo
Journal:  J Bone Miner Res       Date:  2004-01-19       Impact factor: 6.741

View more
  8 in total

1.  High-Resolution Cone-Beam Computed Tomography is a Fast and Promising Technique to Quantify Bone Microstructure and Mechanics of the Distal Radius.

Authors:  Karen Mys; Peter Varga; Filip Stockmans; Boyko Gueorguiev; Verena Neumann; Olivier Vanovermeire; Caroline E Wyers; Joop P W van den Bergh; G Harry van Lenthe
Journal:  Calcif Tissue Int       Date:  2021-01-16       Impact factor: 4.333

Review 2.  Radiation dose in non-dental cone beam CT applications: a systematic review.

Authors:  Cosimo Nardi; Sergio Salerno; Roberto Molteni; Mariaelena Occhipinti; Giulia Grazzini; Niccolò Norberti; Cesare Cordopatri; Stefano Colagrande
Journal:  Radiol Med       Date:  2018-06-05       Impact factor: 3.469

3.  Guidelines for the assessment of bone density and microarchitecture in vivo using high-resolution peripheral quantitative computed tomography.

Authors:  D E Whittier; S K Boyd; A J Burghardt; J Paccou; A Ghasem-Zadeh; R Chapurlat; K Engelke; M L Bouxsein
Journal:  Osteoporos Int       Date:  2020-05-26       Impact factor: 4.507

Review 4.  The use of bone mineral density measured by dual energy X-ray absorptiometry (DXA) and peripheral quantitative computed microtomography in chronic kidney disease.

Authors:  Martin Jannot; Fabrice Mac-Way; Vanessa Lapierre; Marie-Helene Lafage-Proust
Journal:  J Nephrol       Date:  2017-09-12       Impact factor: 3.902

Review 5.  A Contemporary View of the Diagnosis of Osteoporosis in Patients With Axial Spondyloarthritis.

Authors:  Mie Jin Lim; Kwi Young Kang
Journal:  Front Med (Lausanne)       Date:  2020-12-11

6.  Cost-effectiveness of introducing cone-beam computed tomography (CBCT) in the management of complex phalangeal fractures: economic simulation.

Authors:  N Faccioli; E Santi; G Foti; G Mansueto; M Corain
Journal:  Musculoskelet Surg       Date:  2020-11-19

7.  Spine Bone Texture Assessed by Trabecular Bone Score in Active and Controlled Acromegaly: A Prospective Study.

Authors:  E Sala; E Malchiodi; G Carosi; E Verrua; E Cairoli; E Ferrante; M Filopanti; C Eller-Vainicher; F M Ulivieri; A Spada; M Arosio; I Chiodini; G Mantovani
Journal:  J Endocr Soc       Date:  2021-05-15

Review 8.  The clinical application of high-resolution peripheral computed tomography (HR-pQCT) in adults: state of the art and future directions.

Authors:  J P van den Bergh; P Szulc; A M Cheung; M Bouxsein; K Engelke; R Chapurlat
Journal:  Osteoporos Int       Date:  2021-05-22       Impact factor: 4.507

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.