Literature DB >> 27114250

Assessment of the Reliability of the Sysmex XE-5000 Analyzer to Detect Platelet Clumps.

Jennifer Hawkins1, Gene Gulati2, Guldeep Uppal3, Jerald Gong3.   

Abstract

OBJECTIVE: Platelet counts generated by automated analyzers on blood specimens that contain platelet clumps are often inaccurate and require verification by blood-smear review. In this study, we assessed the reliability of the Sysmex XE-5000 instrument to detect platelet clumps.
METHOD: We reviewed automated complete blood count (CBC) results and the findings of the microscopic review of corresponding blood smears of 600 blood specimens specifically selected from the routine laboratory workload. The sensitivity, specificity, efficiency, positive predictive value (PPV), and negative predictive value (NPV) of its 2 platelet-associated flags (abnormal platelet-size distribution [PAD] flag and platelet-clumps [CLP] flag) were determined.
RESULTS: The respective values for the sensitivity, specificity, efficiency, and PPV were 42%, 83%, 63%, and 1% for the PAD flag and 57%, 99%, 78%, and 37% for the CLP flag. The NPV was 100%.
CONCLUSION: The overall reliability of the CLP flag is superior than that of the PAD flag but there is room for further improvement. © American Society for Clinical Pathology, 2016. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.

Entities:  

Keywords:  CBC; Sysmex XE-5000; assessment; automated platelet counts; platelet clumps; reliability

Mesh:

Year:  2016        PMID: 27114250      PMCID: PMC4985765          DOI: 10.1093/labmed/lmw016

Source DB:  PubMed          Journal:  Lab Med        ISSN: 0007-5027


  3 in total

1.  Platelet counting by the coulter LH 750, sysmex XE 2100, and advia 120: a comparative analysis using the RBC/platelet ratio reference method.

Authors:  Linda M Sandhaus; Ebenezer S Osei; Neeta N Agrawal; Chrisine A Dillman; Howard J Meyerson
Journal:  Am J Clin Pathol       Date:  2002-08       Impact factor: 2.493

2.  Detection of Platelet Clumps on Peripheral Blood Smears by CellaVision DM96 System and Microscopic Review.

Authors:  Gene Gulati; Guldeep Uppal; Alina Dulau Florea; Jerald Gong
Journal:  Lab Med       Date:  2014

3.  Technology and new fluorescence flow cytometry parameters in hematological analyzers.

Authors:  Borros M Arneth; Mario Menschikowki
Journal:  J Clin Lab Anal       Date:  2014-05-05       Impact factor: 2.352

  3 in total
  2 in total

1.  The diagnostic accuracy of Sysmex XN for identification of pseudothrombocytopenia using various thresholds for definition of platelet aggregation.

Authors:  Hanne Elisabeth Lunde; Agnete Nyborg Hjelmtvedt; Erik Koldberg Amundsen
Journal:  Int J Lab Hematol       Date:  2022-06-24       Impact factor: 3.450

2.  Peculiarities of Platelet Metabolism in Patients with Acute Coronary Syndrome with Anxiety-Depressive Disorders and Informativity of Enzymes in the Forecast of Development of Cardiovascular Complications.

Authors:  Natalya Yu Shimokhina; Andrey A Savchenko; Marina M Petrova
Journal:  Pharmaceuticals (Basel)       Date:  2020-07-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.