Tharoeun Thap1, Kwon-Ha Yoon2, Jinseok Lee3. 1. Department of Biomedical Engineering, Wonkwang University School of Medicine, 460 Iksandeaero, Iksan, Jeonbuk 570-749, Korea. bami1314@wku.ac.kr. 2. Department of Radiology, Wonkwang University School of Medicine, 460 Iksandeaero, Iksan, Jeonbuk 570-749, Korea. khy1646@wku.ac.kr. 3. Department of Biomedical Engineering, Wonkwang University School of Medicine, 460 Iksandeaero, Iksan, Jeonbuk 570-749, Korea. gonasago@wku.ac.kr.
Abstract
We proposed new electrodes that are applicable for electrocardiogram (ECG) monitoring under freshwater- and saltwater-immersion conditions. Our proposed electrodes are made of graphite pencil lead (GPL), a general-purpose writing pencil. We have fabricated two types of electrode: a pencil lead solid type (PLS) electrode and a pencil lead powder type (PLP) electrode. In order to assess the qualities of the PLS and PLP electrodes, we compared their performance with that of a commercial Ag/AgCl electrode, under a total of seven different conditions: dry, freshwater immersion with/without movement, post-freshwater wet condition, saltwater immersion with/without movement, and post-saltwater wet condition. In both dry and post-freshwater wet conditions, all ECG-recorded PQRST waves were clearly discernible, with all types of electrodes, Ag/AgCl, PLS, and PLP. On the other hand, under the freshwater- and saltwater-immersion conditions with/without movement, as well as post-saltwater wet conditions, we found that the proposed PLS and PLP electrodes provided better ECG waveform quality, with significant statistical differences compared with the quality provided by Ag/AgCl electrodes.
We proposed new electrodes that are applicable for electrocardiogram (ECG) monitoring under freshwater- and saltwater-immersion conditions. Our proposed electrodes are made of graphite pencil lead (GPL), a general-purpose writing pencil. We have fabricated two types of electrode: a pencil lead solid type (PLS) electrode and a pencil lead powder type (PLP) electrode. In order to assess the qualities of the PLS and PLP electrodes, we compared their performance with that of a commercial Ag/AgCl electrode, under a total of seven different conditions: dry, freshwater immersion with/without movement, post-freshwater wet condition, saltwater immersion with/without movement, and post-saltwater wet condition. In both dry and post-freshwater wet conditions, all ECG-recorded PQRST waves were clearly discernible, with all types of electrodes, Ag/AgCl, PLS, and PLP. On the other hand, under the freshwater- and saltwater-immersion conditions with/without movement, as well as post-saltwater wet conditions, we found that the proposed PLS and PLP electrodes provided better ECG waveform quality, with significant statistical differences compared with the quality provided by Ag/AgCl electrodes.
One of the most important methods of monitoring vital signs that track an individual’s health is using an electrocardiogram (ECG or EKG). It provides meaningful information regarding person’s heart performance and functionality. ECG is a recording of the electrical activity of the myocardium during one cardiac cycle. It is characterized by a recurrent sequence of PQRST waves and a conditional U-wave, and it is recorded by placing electrodes on a body surface in a non-invasive manner [1]. Each different ECG wave signature correlates with a phase of the heart’s activity, and the observed characteristics are clinically important in diagnostic decision for various cardiac diseases [2]. The information extracted from the morphological changes in the ECG waveform can be used to diagnose cardiac ischemia, injury, and malignant arrhythmias [3,4]. The cardiovascular system, the heart and circulation, are mostly controlled by higher brain centers and cardiovascular control areas in the brain through the sympathetic and parasympathetic branches of the autonomic nervous system. Efferent sympathetic and vagal activities directed to the sinus node are characterized by discharges, largely synchronous with each cardiac cycle, which can be modulated by central and peripheral oscillators. These mechanisms generate rhythmic fluctuations in efferent neural discharge, which manifest short and long-term oscillations in the heart period. Heart rate variability (HRV) revealed by ECG analysis becomes a non-invasive technique for assessing autonomic influences on the heart [5,6,7,8].Based on various clinical needs, much effort has been spent lately to make ECG monitoring an easy and “anywhere-anytime” available procedure [9,10,11,12,13,14,15,16,17,18]. However, ECGs can be measured and analyzed in a limited variety of environments. Especially in the water-submerged condition, the signal quality is dramatically decreased; thus, the ECG signal analysis such as HRV becomes highly debatable. In underwater environments, pressure, temperatures, and depth have an impact on the human physiological control systems [19]. As the control systems break down due to prolonged hyperbaric exposures, humans are more susceptible to detrimental conditions such as hypothermia, hypoxia, and many neurological and cardiovascular problems including decompression sickness (DCS) [20,21,22,23]. It is reported that early detection of DCS can be made from HRV analysis [23,24].Underwater ECG monitoring further can be extended to life vest equipment and aqua therapy with the main purpose to ensure safety of any water-related activity. A fully functional and reliable underwater ECG monitoring system, capable of recording ECG waveforms, does not currently exist to support either shallow or deep diving medical research. ECG electrodes that can operate in water-submerged environments have just been recently developed [14]. It is reported that the Ag/AgCl electrode was replaced by carbon black powder to obtain ECG signals underwater without the use of waterproof adhesive tape. However, there are no published studies yet that have analyzed the performance under saltwater immersion.Graphite is an allotrope of carbon and is a substance made of pure carbon. Other allotropes include diamonds, amorphous carbon, and charcoal. Graphite is mostly consumed for refractories, batteries, steelmaking, expanded graphite, brake linings, foundry facings, and lubricants. Graphite is also used as a pigment and a molding agent in glass manufacturing. Nuclear reactors also use graphite as an electron moderator. The detailed properties and characteristics of graphite have been specifically described elsewhere [25,26,27]. In this study, we evaluated graphite pencil lead (GPL) for ECG measurement under fresh- and saltwater. GPL is easily found and is widely used for writing and drawing. We have developed two types of dry electrodes from GPL and evaluated the usability of our electrodes in the seven different conditions: dry condition, freshwater immersion with/without movement condition, post-freshwater condition (i.e., freshwater wet condition), saltwater immersion with/without movement condition, and post-saltwater condition (i.e., saltwater wet condition). We have also described the fabrication process, the impedance characteristics, and the quality evaluation of the ECG morphology in each condition.
2. Materials and Methods
2.1. Fabrication of GPL Electrodes
In a pre-experiment, we simply cut and placed the GPL on the metal cap, and then held them altogether with duct tape as shown in Figure 1a. We connected the GPL with the PowerLab 8/35 and DualBioAmps FE135 (ADInstrument, Sydney, Australia), and contiguously placed for a standard 1-channel ECG monitor with sampling rate of 1 kHz and low pass filter cutoff frequency at 50 Hz. The resultant signal is shown in Figure 1b. With this evidence, we fabricated two types of GPL electrodes: pencil lead solid type (PLS) and pencil lead powder type (PLP).
Figure 1
Pre-experiment for testing the usability of a graphite pencil lead electrode. (a) GPL on the metal cap; (b) Resultant signal.
2.1.1. Pencil Lead Solid-Type (PLS) Electrode
PLS electrodes were fabricated as presented below:The 4B pencil leads were used as a material for making PLS electrodes. The pencil leads were cut into pieces as shown in Figure 2a.
Figure 2
(a) Pencil leads; (b) Flattened pencil leads; (c) Each lead and the metal cap attached using a bond and then supported with a piece of paper; and (d) Pencil lead solid-type electrode.
The cutting leads were flattened to form a rounded rectangle cross-sectional shape as shown in Figure 2b.The flattened pencil leads were arranged in a parallel manner with the metal cap placed on top and attached to each other by using an ethyl cyanoacrylate bond. The cap and lead surface are perfectly attached and conducted without bond infiltration as shown in Figure 2c. We applied the bond again around the metal cap and placed a piece of paper on it to support the pencil leads and cap.We circularly trimmed the edge. Figure 2d shows the face and back view of the PLS electrode.For the performance evaluation, we fabricated the PLS electrode with a diameter of 10 mm and a thickness of 1 mm, which is the same size as the commercial Ag/AgCl electrode (3M Health Care, 2223H, Seoul, Korea).
2.1.2. Pencil Lead Powder-Type (PLP) Electrode
PLP electrodes were fabricated as presented below:As same as PLS type, the 4B pencil leads were used as raw materials, and the pencil leads were peeled and grinded into powder as shown in Figure 3a.
Figure 3
(a) Pencil lead grinding powder; (b) small piece of acrylic tube within electrical wire being placed at the bottom; and (c) pencil lead powder-type electrode.
We mixed the pencil lead powder with chloroprene-rubber based bond and poured it into a piece of cylindrical acrylic tube within an electrical wire being placed at the bottom as shown in Figure 3b.We finally sanded the face of the electrode, made it smooth and flat, and subsequently applied the hot melted glue to support the wire on the backside as shown in Figure 3c.In order to evaluate its performance, we also fabricated the PLP electrode with a diameter of 10 mm and thickness of 1 mm in order to compare it with the PLS and the commercial available Ag/AgCl electrode. Figure 4 shows three types of electrodes that we evaluated and used to compare the signal quality: the commercial Ag/AgCl electrodes (3M Health Care), PLP electrodes, and PLS electrodes.
Figure 4
Three types of electrodes being used to perform the experiment: (a) commercial Ag/AgCl electrodes; (b) pencil lead powder-type electrodes; and (c) pencil lead solid-type electrodes.
2.2. Measurement Protocol and Signal Processing
In this study, twelve volunteer healthy subjects with ages ranging from 23 to 35 years (27.17 ± 3.33; mean ± STD), with average weight 74.75 ± 10.50 kg, and average height 173.33 ± 3.26 cm were included. The study protocol and data analysis were approved by the Institutional Review Board (IRB) at Wonkwang University of Hospital (WKUH201401-CTDV-001) and all volunteers consented to be subjects for the experiment.The PowerLab 8/35 and DualBioAmps FE135 (ADInstrument) were simultaneously used to acquire ECG signals from the subjects. The ECG signals were displayed and recorded by using LabChart software (Ver. 7.3.7 Professional). We selected a sampling rate of 1 kHz with low pass filter cutoff frequency at 50 Hz for the frequency content of a QRS complex, which extends up to 40 Hz [28,29,30,31]. A minimal sampling rate of 500 Hz with a resolution of 12 bits has been recommended for the adult ECG by the American Heart Association [32,33,34], but for pediatric ECGs, higher sampling rates have been suggested [35,36]. Practically, for portable devices (e.g., Holter systems), the tendency is to reduce the sampling rate in order to increase the speed of analysis and to limit the data storage [37,38]. Thus, the sampling rate can be set up in the range between 100 and 1000 Hz in considering speed of analysis, data storage and time-resolution requirement [38,39,40,41]. In this paper, we used a high sampling rate of 1 kHz for the sophisticated evaluation of our proposed electrodes, which provides high time-resolution of 1 ms in analysis of HRV or waveform pattern. The recorded ECG signals were analyzed for R-peak detection and heart rate variability (HRV) using Matlab (R2012b).ECG signals were simultaneously acquired and compared as follows: (Experiment-I) Ag/AgCl electrodes vs. PLS electrodes, and (Experiment-II) Ag/AgCl electrodes vs. PLP electrodes. As described in Table 1, each experiment was performed under two conditions, freshwater and saltwater. When all the experiments under freshwater were completed, the pool was washed and refilled with 400 liters of clean water mixed with 14 kg of coarse salt in order to make salt concentration similar to seawater (i.e., on average 35 g/kg) [42]. The electrodes were contiguously placed for a standard 1-channel ECG by two trained cardiologists in Wonkwang University Hospital. For the electrode attachment on the body, the attachment skin sites were cleaned with 83% ethanol. The PLS and PLP electrodes were simultaneously attached with a regular surgical tape (3M Micropore surgical tape 1530S-1). We add that the PLS and PLP electrodes were reused throughout the week when data were collected, while new Ag/AgCl electrodes were provided to every subject for each measurement.
Table 1
Experimental protocol.
Index
Experiment-I: Ag/AgCl vs. PLS
Experiment-II: Ag/AgCl vs. PLP
(i)
Dry Condition (5 min)
Dry Condition (5 min)
Case-I: Freshwater
Case-II: Saltwater
Case-I: Freshwater
Case-II: Saltwater
(ii)
Immersion without Movement (5 min)
Immersion without Movement (5 min)
(iii)
Immersion with Movement (3 min)
Immersion with Movement (3 min)
(iv)
Wet Condition (5 min)
Wet Condition (5 min)
Figure 5 summarizes all the steps, including Experiment I and II under freshwater and saltwater, respectively. Each experiment was performed for a total of 31 min. Each step was defined as follows:
Figure 5
Experimental protocol for electrocardiogram data acquisition: (a) dry condition; Immersed condition (b) without movement; and (c) Side-to-side movement; and (d) wet condition.
Five minutes in a sitting position outside the pool (the dry condition).Five-minute relaxation (lying position) inside the water pool (the immersion condition without movement).Three-minute side-to-side moving condition inside the water pool (the immersion condition with movement).Five minutes in the standing position outside the water pool (the wet condition).The pool water temperature was 20.50 ± 1.30 °C; room temperature, 25 °C; and the outdoor temperature, 4.0 ± 1.0 °C. During the 5-min dry condition (Step 1), the subjects were asked to remain relaxed in the sitting position outside the pool; the electrodes at this point were completely dry. Step 1 is similar to a regular ECG measurement in a clinical environment. During the 5-min water-immersed condition (Step 2), the subjects were asked to immerse in the water pool, with water coming up to their neck so that all electrodes were fully immersed. Subsequently, the subjects were asked to perform side-to-side turning movements with the speed about 2 turns per second for 3 min (Step 3). Finally, the subjects were asked to exit the pool and remain relaxed in the standing position with the wet electrodes still applied to their skin for 5 min (Step 4).The acquired ECG data were automatically filtered in the LabChart software (Ver.7.3.7 Pro.) through lowpass filter with cutoff frequency at 50 Hz. The signals were saved as Matlab file for data analysis. R-wave peak detection was performed on each recorded 31 minute-ECG signal by incorporating a filter bank with variable cut-off frequencies, spectral estimates of the heart rate, rank-order nonlinear filters and direct analysis [28,43]. The resultant R-wave peaks were confirmed by two trained cardiologists. For the HRV analysis, both time domain and frequency domain analysis were performed. We used HRV analysis as a quantitative performance measure since most previous hyperbaric studies [21,23,24,44] were concerned with changes in the dynamics of the autonomic nervous system.HRV was measured in both time and frequency domains [45]. Time domain analysis included mean normal-to-normal (NN) interval (meanNN), standard deviation of all NN intervals (SDNN), the root mean square of the differences between adjacent NN intervals (RMSSD) and number of pairs of adjacent NN intervals differing by more than 50 ms in the entire recording (NN50). Frequency domain HRV analysis included low frequency (LF, 0.04–0.15 Hz), high frequency (HF, 0.15–0.4 Hz) bands, total power, and HF/LF ratio.In the frequency domain analysis, we used RRI series along the time axis and analyzed with the fast Fourier transform (FFT). The analysis was performed using RRI time series, then the FFT yields complex numbers spaced evenly along the frequency axis from zero to half the sampling rate. The squared magnitude of each of the complex values from the FFT was then computed (real part squared plus imaginary part squared), as this yields the power spectral density (PSD) function. For both time and frequency domain analysis, premature ventricular contractions (PVCs) and premature atrial contractions (PACs) have a particularly insidious effect on each result since the effect of a sudden, brief change in RR interval (or HR) adds substantial power density to all frequencies or increases the statistical values, often overwhelming the true variability that is assessed [46]. On the other hand, if the RR intervals corresponding to the ectopic beats are simply deleted, individual frequency components change because the deletion of time advances the phase. This has the effect of reducing the calculated power density at some frequency components and augmenting it at others. We replaced the RR intervals affected by an ectopic beat with the same number of RR intervals of the value equal to the mean of those replaced.For further performance analysis, we used cross-correlation coefficient measures. The ECG template obtained from Ag/AgCl electrode in the dry condition as a reference was used to quantify the signal quality obtained from PLS, PLP, and Ag/AgCl electrodes in the dry, fresh- and saltwater immersion and wet conditions by measuring cross-correlation coefficients r as defined by where represents the one cardiac cycle of ECG template obtained with the Ag/AgCl electrodes in the dry condition as a reference, which is clinically measured, and represents the one cardiac cycle of ECG samples with each different electrode in each different condition. We calculated the cross-correlation coefficient values in the entire measured ECG samples.Temporal and spectral measures of HRV and cross-correlation coefficients obtained from PLS and PLP electrodes were compared with those obtained from Ag/AgCl electrodes using the t-test. Statistical significance was accepted at p < 0.05.
2.3. Impedance Characterization
The electrode-to-skin contact impedances were analyzed for the three pairs of electrodes: PLS, PLP, and Ag/AgCl. The impedance was measured by using an impedance analyzer (Agilent 4294A precision impedance analyzer), which covers a broader test-frequency range from 40 Hz to 110 MHz, with a basic impedance accuracy of ±0.08%. The test signal level was 600 μA, and the frequency ranged from 40 Hz to 1 MHz. For each measurement, each pair of electrodes was placed on anterior forearm skin, 5 cm apart from each other. While ECG-power is mostly below 40 Hz; the electrodes for impedance measurements were chosen so that our results can be compared with the previous studies [11,47]. An elastic compression bandage was used to fix them to the skin. Prior to each measurement, the skin area was cleaned with a skin cleanser. For each of the subjects, all measurements were performed on the same day.
3. Results and Discussion
3.1. Dry and under Freshwater Condition (Case-I): PLS vs. Ag/AgCl and PLP vs. Ag/AgCl
Figure 6 shows approximately 5-s ECG data measurements using PLS and Ag/AgCl electrodes for each step: dry, freshwater immersion with/without movement, and freshwater wet condition. It is shown that the PLS electrodes provided all morphological components of ECG signals, in all conditions. Especially in freshwater immersion, with/without movement conditions, the ECG signal from PLS provided better quality than the one from Ag/AgCl, which is almost indiscernible.
Figure 6
(Experiment I—Pencil lead solid-type (PLS) electrode compared with Ag/AgCl electrode) Example of electrocardiogram (ECG) signals for each experimental condition: (a) dry condition; freshwater immersed condition (b) without movement; and (c) with movement; and (d) freshwater wet condition. ECG signals acquired with Ag/AgCl (blue line) and PLS electrodes (red line).
Figure 7 shows approximately 5-s ECG data measurements using PLP and Ag/AgCl electrodes for each step: dry, freshwater immersion with/without movement, and freshwater wet condition. It is also shown that the PLP electrodes provided all morphological components of ECG signals, in all conditions. Similar to PLS, in freshwater immersion with/without movement conditions, the ECG signal from PLP provided better quality than the one from Ag/AgCl, which is almost indiscernible.
Figure 7
(Experiment II—Pencil lead powder-type (PLP) electrode compared with Ag/AgCl electrode) Example of electrocardiogram (ECG) signals for each experimental condition: (a) dry condition; freshwater immersed condition (b) without movement; and (c) with movement; and (d) freshwater wet condition. ECG signals acquired with Ag/AgCl (blue line) and PLP electrodes (red line).
Figure 8 shows an example of Ag/AgCl- and PLS-based ECG template directly derived from Figure 6a,d. Similarly, Figure 9 shows an example of Ag/AgCl- and PLP-based ECG template directly derived from Figure 7a,d. Ag/AgCl electrodes were good for dry and wet conditions. However, for the freshwater immersion with/without movement condition, the signals were hardly recognized, and the R-wave was almost indiscernible.
Figure 8
Example of electrocardiogram templates for dry and freshwater wet condition: (a) template derived from Figure 6a; and (b) template derived from Figure 6d.
Figure 9
Example of electrocardiogram templates for dry and freshwater wet condition: (a) template derived from Figure 7a; and (b) template derived from Figure 7d.
Statistical results of HRV indices and cross-correlation coefficients obtained with Ag/AgCl, PLS, and PLP electrodes are summarized in Table 2, Table 3, Table 4 and Table 5, for the dry, freshwater immersion without/with movement, and freshwater wet conditions, respectively. Regarding the temporal and spectral measures of HRV, under the dry and wet conditions, no statistical differences (p > 0.05) were identified for any of the temporal or spectral measures of HRV obtained within each pair: Ag/AgCl vs. PLS and Ag/AgCl vs. PLP. However, in the freshwater immersion conditions, we found significant statistical differences. More specifically, without movement, the statistical differences were identified when the Ag/AgCl vs. PLS was with SDNN, RMSSD, LF, and HF, the Ag/AgCl vs. PLP was with SDNN, RMSSD, and LF/HF as summarized in Table 3. With movement, the statistical differences identified when the Ag/AgCl vs. PLS was with mean NN, SDNN, RMSSD, and NN50, and the Ag/AgCl vs. PLP was with mean NN, SDNN, RMSSD, and NN50, LF, HF and Total power, as summarized in Table 4. Regarding the cross-correlation coefficients, we found the strong correlation (>0.8) in the dry and wet conditions regardless of the electrode types. In addition, no statistical differences were identified. On the other hand, in the immersion without movement condition, the signal from Ag/AgCl electrode indicated weak correlation (<0.5) whereas the signal from PLS and PLP showed the strong correlation higher than 0.9. Definitely, the correlation coefficients in the immersion with movement condition were significantly reduced. However, as can be shown in Table 4, we found PLS and PLP showed higher correlation of 0.51 and 0.45, respectively, than that of Ag/AgCl with 0.19, which was reduced by approximately 0.3. Moreover, the statistical differences were identified within each pair: Ag/AgCl vs. PLS and Ag/AgCl vs. PLP in the freshwater immersion with/without movement conditions. Thus, the signals obtained from PLS and PLP electrodes showed stronger correlation than the ones from Ag/AgCl electrode.
Table 2
Electrocardiogram signal quality—dry condition.
Parameter
Ag/AgCl
PLS
p-Value
Ag/AgCl
PLP
p-Value
Temporal measures of HRV
meanNN (ms)
700.5 ± 89.6
700.1 ± 91.1
0.9816
723.4 ± 86.2
723.0 ± 83.8
0.9853
SDNN (ms)
59.2 ± 22.5
55.8 ± 27.3
0.7819
54.2 ± 16.0
48.8 ± 16.3
0.4171
RMSSD (ms)
58.7 ± 39.8
51.3 ± 42.4
0.7075
50.4 ± 27.9
38.3 ± 27.8
0.2986
NN50
23.22 ± 19.44
21.77 ± 19.33
0.8764
28.83 ± 21.88
27.75 ± 23.31
0.9076
Spectral measures of HRV
LF (ms2)
1258 ± 783
1066 ± 757
0.6042
992 ± 760
895 ± 632
0.7371
HF (ms2)
1391 ± 2003
701 ± 1287
0.3976
757 ± 812
498 ± 623
0.3914
Total (ms2)
431562 ± 81172
429086 ± 79566
0.9487
450714 ± 101211
449358 ± 100911
0.9740
LF/HF
3.66 ± 2.69
4.29 ± 2.54
0.6182
2.25 ± 1.71
3.61 ± 2.84
0.1685
Corr-Coef
r (unitless)
-
0.99 ± 0.005
-
-
0.99 ± 0.001
-
Table 3
Electrocardiogram signal quality—freshwater immersion without movement condition.
Parameter
Ag/AgCl
PLS
p-Value
Ag/AgCl
PLP
p-Value
Temporal measures of HRV
meanNN (ms)
840.0 ± 155.6
828.0 ± 97.8
0.8358
778.0 ± 275.1
746.9 ± 119.1
0.4171
SDNN (ms)
112.9 ± 64.0
53.5 ± 13.8
0.0223
225.4 ± 135.2
72.8 ± 27.6
0.0044
RMSSD (ms)
144.9 ± 101.7
54.5 ± 20.0
0.0271
304.5 ± 189.0
75.9 ± 41.8
0.0027
NN50
61.37 ± 39.28
43.0 ± 33.6
0.3318
162.33 ± 108.57
102.88 ± 99.19
0.2429
Spectral measures of HRV
LF (ms2)
3809 ± 3317
1105 ± 686
0.0404
11390 ± 15613
1514 ± 1238
0.0767
HF (ms2)
7168 ± 7290
886 ± 600
0.0292
24366 ± 38574
1678 ± 2052
0.0971
Total (ms2)
667061 ± 15484
638996 ± 15272
0.7205
584884 ± 181314
492856 ± 147666
0.2549
LF/HF
1.03 ± 0.74
1.42 ± 0.66
0.2799
0.62 ± 0.29
1.47 ± 0.89
0.0154
Corr-Coef
r (unitless)
0.41 ± 0.23
0.96 ± 0.02
0.0000
0.41 ± 0.23
0.93 ± 0.04
0.0000
Table 4
Electrocardiogram signal quality—freshwater immersion with movement condition.
Parameter
Ag/AgCl
PLS
p-Value
Ag/AgCl
PLP
p-Value
Temporal measures of HRV
meanNN(ms)
1130.0 ± 799.9
753.1 ± 171.5
0.0205
1146.3 ± 1055.1
728.0 ± 139.0
0.0145
SDNN (ms)
754.4 ± 383.7
152.6 ± 70.1
0.0087
834.0 ± 620.1
122.2 ± 53.5
0.0060
RMSSD (ms)
1010.2 ± 579.7
199.0 ± 103.1
0.0151
1109.0 ± 832.3
156.3 ± 72.8
0.0061
NN50
117.4 ± 17.03
48.6 ± 32.02
0.0028
83.12 ± 51.48
28.62 ± 18.2
0.0136
Spectral measures of HRV
LF (ms2)
223079 ± 24173
4007 ± 3893
0.0773
168449 ± 189442
3319 ± 2135
0.0272
HF (ms2)
354946 ± 44101
8967 ± 12222
0.1176
348641 ± 392893
6166 ± 4855
0.0272
Total (ms2)
2336098 ± 1864
535985 ± 10881
0.0633
2268620 ± 2114
475750 ± 69503
0.0310
LF/HF
0.72 ± 0.72
0.74 ± 0.61
0.9582
0.61 ± 0.22
0.80 ± 0.61
0.4278
Corr-Coef
r (unitless)
0.19 ± 0.11
0.51 ± 0.16
0.0000
0.19 ± 0.11
0.45 ± 0.21
0.0000
Table 5
Electrocardiogram signal quality—freshwater wet condition.
Parameter
Ag/AgCl
PLS
p-Value
Ag/AgCl
PLP
p-Value
Temporal measures of HRV
meanNN(ms)
811.6 ± 100.6
810.8 ± 97.2
0.9670
794.6 ± 130.2
792.5 ± 116.9
0.9609
SDNN (ms)
73.7 ± 26.6
69.4 ± 24.7
0.7767
84.2 ± 41.8
70.7 ± 24.9
0.4465
RMSSD (ms)
70.4 ± 38.0
61.2 ± 21.6
0.6163
86.8 ± 62.5
63.7 ± 28.2
0.3563
NN50
56.83 ± 46.69
55.16 ± 46.08
0.9516
62.12 ± 50.25
58.87 ± 50.14
0.8988
Spectral measures of HRV
LF (ms2)
2293 ± 2142
2182 ± 2098
0.9295
2344 ± 2406
2156 ± 1962
0.8660
HF (ms2)
1722 ± 2032
819 ± 837
0.3377
1408 ± 1778
1306 ± 942
0.8888
Total (ms2)
626275 ± 13981
621357 ± 13557
0.9518
585003 ± 147931
584516 ± 145797
0.9947
LF/HF
2.44 ± 2.10
3.29 ± 2.32
0.5219
3.27 ± 3.57
3.01 ± 3.54
0.8840
Corr-Coef
r (unitless)
0.89 ± 0.01
0.88 ± 0.01
0.2551
0.89 ± 0.01
0.90 ± 0.01
0.2565
3.2. Under Saltwater Condition (Case-II): PLS vs. Ag/AgCl and PLP vs. Ag/AgCl
Figure 10a–c shows approximately 5-s ECG data measurements using PLS and Ag/AgCl electrodes under the saltwater immersion condition, followed by the movement condition and the wet condition, respectively. It is shown that the PLS electrodes provided better quality of ECG signals with clear defined R-waves in all three conditions. On the other hand, the ECG signals from Ag/AgCl were almost indiscernible. Similarly, Figure 10d through f show approximately 5-s ECG data measurements measured by PLP and Ag/AgCl electrodes under the same conditions. It is also shown that the PLP electrodes provided better quality of ECG signals in all three conditions.
Figure 10
Example of electrocardiogram signals for each experimental condition. (a–c) Pencil lead solid-type electrode compared with Ag/AgCl electrode; (d–f) Pencil lead powder-type electrode compared with Ag/AgCl electrode; Saltwater immersed condition: (a,d) without movement; and (b,e) with movement; (c,f) Saltwater wet condition.
Statistical results of temporal and spectral HRV indices obtained with Ag/AgCl, PLS, and PLP electrodes are summarized in Table 6, Table 7 and Table 8 for the saltwater immersion without movement, followed by immersion with movement, and the wet conditions, respectively. We identified statistical differences in the all conditions. More specifically, without movement, the statistical differences were identified when the Ag/AgCl vs. PLS was with RMSSD, LF and HF, and the Ag/AgCl vs. PLP was with SDNN, RMSSD, LF and HF as summarized in Table 6. With movement, the statistical differences were identified when the Ag/AgCl vs. PLS was with mean NN, SDNN, RMSSD, and total power, and the Ag/AgCl vs. PLP was with mean NN, SDNN, RMSSD, LF, HF, and total power as summarized in Table 7. In the saltwater wet condition, the significant statistical differences were identified when the Ag/AgCl vs. PLS was with SDNN, RMSSD, LF, HF, and LF/HF, and the Ag/AgCl vs. PLP is with SDNN, RMSSD, and HF as summarized in Table 8. Furthermore, we found the strong correlation (>0.8) under the wet condition regardless of the electrode types as shown in Table 8. On the other hand, in the immersion with/without movement conditions, the correlation coefficients from PLS and PLP showed higher values than that of Ag/AgCl, as shown in Table 6 and Table 7. The statistical differences were identified in the saltwater immersion with/without movement conditions.
Table 6
Electrocardiogram signal quality—saltwater immersion without movement condition.
Parameter
Ag/AgCl
PLS
p-Value
Ag/AgCl
PLP
p-Value
Temporal measures of HRV
meanNN (ms)
882 ± 162
823 ± 86
0.3275
841 ± 103
813 ± 110
0.4972
SDNN (ms)
359 ± 383
130 ± 43
0.0776
313 ± 154
124 ± 82
0.0004
RMSSD (ms)
428 ± 396
175 ± 60
0.0410
421 ± 210
158 ± 117
0.0004
NN50
92.4 ± 73.5
62.5 ± 38.3
0.2903
246.2 ± 184.5
159.6 ± 157.5
0.1998
Spectral measures of HRV
LF (ms2)
21595 ± 31984
2845 ± 1473
0.0405
25621 ± 24694
5994 ± 7756
0.0087
HF (ms2)
40166 ± 60509
6642 ± 5420
0.0480
43893 ± 39516
10705 ± 14515
0.0066
Total (ms2)
753959 ± 34646
606370 ± 14328
0.2291
652235 ± 238080
557039 ± 206664
0.2688
LF/HF
0.5562 ± 0.1238
0.5097 ± 0.1648
0.4850
0.5673 ± 0.1620
1.2010 ± 1.8080
0.2029
Corr-Coef
r (unitless)
0.57 ± 0.19
0.73 ± 0.10
0.0000
0.57 ± 0.19
0.69 ± 0.06
0.0000
Table 7
Electrocardiogram signal quality—saltwater immersion with movement condition.
Parameter
Ag/AgCl
PLS
p-Value
Ag/AgCl
PLP
p-Value
Temporal measures of HRV
meanNN(ms)
1454 ± 223
938 ± 124
0.0020
1277 ± 355
851 ± 120
7.11e-04
SDNN (ms)
1102 ± 315
497 ± 190
0.0054
872 ± 412
335 ± 243
7.92e-04
RMSSD (ms)
1481 ± 309
677 ± 265
0.0023
1197 ± 558
448 ± 293
4.53e-04
NN50
132.2 ± 41.2
117.6 ± 22.9
0.5072
171.8 ± 116.5
102.1 ± 58.5
0.0785
Spectral measures of HRV
LF (ms2)
296431 ± 35372
45736 ± 29809
0.1529
174843 ± 241251
26346 ± 30832
0.0459
HF (ms2)
531535 ± 37302
146587 ± 10063
0.0564
439754 ± 499125
57801 ± 58326
0.0151
Total (ms2)
2984351 ± 1396
1088667 ± 4357
0.0199
2459872 ± 20369
748602 ± 261051
0.0085
LF/HF
0.4661 ± 0.2187
0.3257 ± 0.0298
0.1928
0.4170 ± 0.1328
0.4440 ± 0.1370
0.6284
Corr-Coef
r (unitless)
0.21 ± 0.14
0.51 ± 0.17
0.0000
0.21 ± 0.14
0.63 ± 0.13
0.0000
Table 8
Electrocardiogram signal quality—saltwater wet condition.
Parameter
Ag/AgCl
PLS
p-Value
Ag/AgCl
PLP
p-Value
Temporal measures of HRV
meanNN (ms)
789 ± 60
756 ± 67
0.4320
848 ± 121
823 ± 114
0.5957
SDNN (ms)
228 ± 89
60 ± 17.8
0.0032
173 ± 107
77.2 ± 37.4
0.0055
RMSSD (ms)
297 ± 123
55.8 ± 28.5
0.0027
219 ± 142
86.2 ± 69.7
0.0059
NN50
70.8 ± 44.7
44.4 ± 47.7
0.4028
82.5 ± 49.6
68.1 ± 57.1
0.5011
Spectral measures of HRV
LF (ms2)
12189 ± 8218
926 ± 568
0.0156
12170 ± 19375
1476 ± 1108
0.0584
HF (ms2)
19837 ± 16445
633 ± 458
0.0311
16844 ± 25268
2123 ± 3183
0.0479
Total (ms2)
589723 ± 85193
497650 ± 95412
0.1461
686362 ± 196078
620400 ± 186430
0.3881
LF/HF
0.7470 ± 0.2407
1.8791 ± 0.9075
0.0272
1.0658 ± 0.7633
1.5443 ± 1.0340
0.1920
Corr-Coef
r (unitless)
0.97 ± 0.01
0.92 ± 0.008
0.2039
0.97 ± 0.01
0.98 ± 0.003
0.2951
3.3. Impedance Characterization
The electrode-to-skin contact impedances for PLP, PLS, and Ag/AgCl were measured and compared. The impedance magnitude |Z| values were plotted versus frequency in Figure 11, where the x-axis is a log scale. The PLP electrodes showed lower electrode-to-skin impedance magnitude followed by the PLS and Ag/AgCl. At 40 Hz, the electrode-skin impedance was 49.47 kΩ, 51.37 kΩ, and 86.12 kΩ for PLP, PLS, and Ag/AgCl, respectively, which shows that the PLP provides 1.03 and 1.74 times lower impedance than PLS and Ag/AgCl do.
Figure 11
The impedance performance of Ag/AgCl, pencil lead powder-type, and pencil lead solid-type electrodes.
4. Conclusions and Discussion
We have developed two types of dry electrodes from graphite material using pencil lead. Type-1 is the PLS electrode and Type-2, the PLP electrode. The performances of PLS and PLP electrodes have been compared with the commercial Ag/AgCl electrode, and the electrode-to-skin contact impedance have also been analyzed. The impedances of PLS and PLP are almost identical, and the values are smaller than those of the Ag/AgCl electrode. The comparison was analyzed for various conditions: dry surface, freshwater immersion with/without movement, post-freshwater (wet), saltwater immersion with/without movement, and post-saltwater (wet) conditions. We were able to capture all morphological ECG waveforms with our dry electrodes in all conditions.When PLS and PLP were compared to Ag/AgCl, we identified significant statistical differences under all conditions, except for dry and freshwater wet conditions. Especially, under the saltwater wet condition after the immersion, R-peak from Ag/AgCl was observed, but the overall morphology was more unstable compared with that from PLS and PLP. Under the freshwater- and saltwater-immersion without movement conditions, we observed an R-wave amplitude reduction with PLS and PLP, but the signal had a good quality and the reduction was not severe enough to obscure the readability of the recordings. We believe that the graphite-based electrodes can be used for the new research area concerning underwater vital sign monitoring (including sea water).
Authors: Paul Kligfield; Leonard S Gettes; James J Bailey; Rory Childers; Barbara J Deal; E William Hancock; Gerard van Herpen; Jan A Kors; Peter Macfarlane; David M Mirvis; Olle Pahlm; Pentti Rautaharju; Galen S Wagner; Mark Josephson; Jay W Mason; Peter Okin; Borys Surawicz; Hein Wellens Journal: J Am Coll Cardiol Date: 2007-03-13 Impact factor: 24.094