| Literature DB >> 27074450 |
Jie Yang1, Grace Tang, Pengpeng Zhang, Margie Hunt, Seng B Lim, Thomas LoSasso, Gig Mageras.
Abstract
Hypofractionated treatments generally increase the complexity of a treatment plan due to the more stringent constraints of normal tissues and target coverage. As a result, treatment plans contain more modulated MLC motions that may require extra efforts for accurate dose calculation. This study explores methods to minimize the differences between in-house dose calculation and actual delivery of hypofractionated volumetric-modulated arc therapy (VMAT), by focusing on arc approximation and tongue-and-groove (TG) modeling. For dose calculation, the continuous delivery arc is typically approximated by a series of static beams with an angular spacing of 2°. This causes significant error when there is large MLC movement from one beam to the next. While increasing the number of beams will minimize the dose error, calculation time will increase significantly. We propose a solution by inserting two additional apertures at each of the beam angle for dose calculation. These additional apertures were interpolated at two-thirds' degree before and after each beam. Effectively, there were a total of three MLC apertures at each beam angle, and the weighted average fluence from the three apertures was used for calculation. Because the number of beams was kept the same, calculation time was only increased by about 6%-8%. For a lung plan, areas of high local dose differences (> 4%) between film measurement and calculation with one aperture were significantly reduced in calculation with three apertures. Ion chamber measurement also showed similar results, where improvements were seen with calculations using additional apertures. Dose calculation accuracy was further improved for TG modeling by developing a sampling method for beam fluence matrix. Single element point sampling for fluence transmitted through MLC was used for our fluence matrix with 1 mm resolution. For Varian HDMLC, grid alignment can cause fluence sampling error. To correct this, transmission volume averaging was applied. For three paraspinal HDMLC cases, the average dose difference was greatly reduced in film and calculation comparisons with our new approach. The gamma (3%, 3 mm) pass rates have improved significantly from 74.1%, 90.0%, and 90.4% to 99.2%, 97.9%, and 97.3% for three cases, for calculation without volume averaging and calculation with volume averaging, respectively. Our results indicate that more accurate MLC leaf position and transmission sampling can improve accuracy and agreement between calculation and measurement, and are particularly important for hypofractionated VMAT that consists of large MLC movement.Entities:
Mesh:
Year: 2016 PMID: 27074450 PMCID: PMC4831077 DOI: 10.1120/jacmp.v17i2.4989
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Illustration of volume averaging effects in sampling fluence through TG of HDMLC 2.5 mm wide leaves in 1 mm grid. The cross sections of three MLC leaves at the isocenter plane are shown. A section through 1 mm matrix is also shown. For fluence sampling purpose, parallel beams are illustrated in dotted lines. Volume averaging is required for the matrix elements filled with diagonal line pattern. Those elements are divided into two equal subelements by the dotted lines inside them for volume averaging.
Figure 2MLC apertures for the middle three control points of the simple arc are shown in white lines in (a), (b), and (c), respectively. Jaws are in yellow lines. White cross is the isocenter. Yellow circle is a point at 1 cm to the left of the isocenter. White dots are grid points on grid. The collimator angle of the arc is 90°.
Chamber measurement compared to four different sets of calculations for a simple 5‐CP arc
|
|
| |||
|---|---|---|---|---|
|
|
|
|
| |
| Measurement | 49.9 | ‐ | 86.0 | ‐ |
| Calc 1 | 50.4 | 0.990 | 85.6 | 1.004 |
| Calc 2 | 42.3 | 1.180 | 85.4 | 1.006 |
| Calc 3 | 51.3 | 0.971 | 85.8 | 1.001 |
| Calc 4 | 43.5 | 1.148 | 85.6 | 1.003 |
a Calculation 1 is a calculation with both corrections of averaging fluence from three apertures and volume averaging of fluence over finer fluence resolution.
b Calculation 2 is a calculation with only volume averaging correction.
c Calculation 3 is a calculation with only three aperture average fluence correction.
d Calculation 4 is with no correction.
Figure 3Dosimetric effects of one vs. three apertures at each beam. Panels (a) and (b) show the planar dose distributions in cGy from single aperture and three apertures overlay with film, respectively. The film measurement is in dotted lines and calculation is represented by solid lines. Panels (c) and (d) show the corresponding dose difference (film ‐ calculation) for (a) and (b), respectively. The dotted display grid size is .
Chamber measurement compared to three different sets of calculations for 6 VMAT arcs
|
|
| |||||
|---|---|---|---|---|---|---|
| Dose (cGy) | Arc 1 | Arc 2 | Arc 3 | Arc 4 | Arc 1 | Arc 2 |
| Measurement | 106.8 | 120.6 | 66.2 | 69.8 | 125.2 | 134.7 |
| Calc (1aper)a | 102.9 | 117.1 | 68.5 | 68.6 | 125.1 | 139.3 |
| Calc(3apers)b | 107.3 | 119.0 | 67.3 | 68.9 | 125.6 | 138.4 |
|
| 107.2 | 118.9 | 67.4 | 68.9 | 125.6 | 138.2 |
|
| ||||||
| Calc (1aper)a | 1.039 | 1.030 | 0.966 | 1.018 | 1.001 | 0.967 |
| Calc(3apers)b | 0.995 | 1.014 | 0.983 | 1.013 | 0.996 | 0.973 |
|
| 0.997 | 1.014 | 0.982 | 1.013 | 0.996 | 0.975 |
a Calc(1aper) is calculation with original aperture at each beam.
b Calc(3apers) is calculation with averaging fluences from the original aperture and two inserted interpolated apertures at each beam.
c is calculation with tripled number of beams. Two additional beams are inserted linearly between two original beams.
MLC leaf movement statistics of the VMAT arcs and calculation time
|
|
| |||||
|---|---|---|---|---|---|---|
| Leaf movement of leaf pair 30 and 31 | Arc 1 | Arc 2 | Arc 3 | Arc 4 | Arc 1 | Arc 2 |
| Average (cm) | 0.8 | 0.5 | 0.5 | 0.6 | 0.2 | 0.7 |
| SD (cm) | 1.2 | 1.1 | 0.7 | 0.7 | 0.5 | 0.8 |
|
| 17 | 12 | 0 | 1 | 0 | 2 |
|
| ||||||
| Calc(3apers) | 1.06 | 1.08 | 1.08 | 1.05 | 1.08 | 1.08 |
|
| 2.9 | 3.0 | 3.1 | 3.0 | 3.0 | 2.8 |
a Calc(1aper) is calculation with original aperture at each CP.
b Calc(3apers) is calculation with averaging fluences from the original aperture and two inserted interpolated apertures at each CP.
c is calculation with tripled number of beams. Two additional beams are inserted linearly between two original beams. Percent of movement greater than 2 cm is calculated by the number of leaf movements that are greater than 2 cm divided by total number of leaf movements.
Figure 4Comparison of dose calculations with film measurement for paraspinal case 1 with HDMLC. Panels (a) and (b) show the planar dose distribution in cGy calculated without volume‐averaging correction and with volume‐averaging correction, respectively. The film measurement is in dotted lines and calculation is represented by solid lines. Panels (c) and (d) show the corresponding dose difference as a percentage of the average dose (film ‐ calculation) for (a) and (b), respectively. The thick black line defines the ROI for average dose and gamma pass rate. The dotted display grid size is .