Wu Che1, Qian Chen1, Yi-Qun Ma1, Yun-Qi Jiang1, Wei Yuan1, Xiao-Gang Zhou1, Xi-Lei Li1, Jian Dong1.
Abstract
BACKGROUND: The purpose of this biomechanical in vitro study was to compare the kinematics and intradiscal pressure achieved with 2 methods: L4-L5 pedicle screw-rod fixation (PSRF) with an upper L3-L4 Coflex device and L4-L5 PSRF alone. The results were used to characterize the biomechanics of the topping-off operation with a Coflex device for the lumbar motion segment adjacent to single-level rigid fixation. MATERIAL/Entities:
Mesh:
Year: 2016 PMID: 27021044 PMCID: PMC4815995 DOI: 10.12659/msm.896706
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Figure 1X-ray examinations of the different groups under the biomechanical test. (A, B) Intact lumbar motion segment (Group A) under the biomechanical test in the X-ray examination; (C, D) Pedicle screw-rod fixation in the L4–L5 segment (group B) under the biomechanical test in the X-ray examination; (E, F) Pedicle screw-rod fixation (PSRF) in the L4–L5 segment with a Coflex device in the L3–L4 segment (group C) under the biomechanical test in the X-ray examination.
ROM of the L3–L4 segment under 8.0 Nm torque (mean ± standard deviation, n=6, unit: degree).
| Motion | Group A | Group B | Group C |
|---|---|---|---|
| Flexion | 2.93±0.26 | 3.47±0.42 | 3.31±0.26 |
| Extension | 2.41±0.23 | 2.75±0.30 | 2.43±0.19 |
| Lateral bending | 2.62±0.29 | 2.92±0.20 | 2.90±0.17 |
| Axial rotation | 2.15±0.24 | 2.34±0.14 | 2.23±0.14 |
ROM – range of motion;
Compared with Group A, P<0.05;
Compared with Group B, P<0.05;
Compared with Group A, P>0.05;
Compared with Group B, P>0.05.
ROM of the L5–S1 segment under 8.0 Nm torque (mean ± standard deviation, n=6, Unit: degree).
| Motion | Group A | Group B | Group C |
|---|---|---|---|
| Flexion | 3.34±0.40 | 3.72±0.30 | 3.60±0.30 |
| Extension | 2.62±0.35 | 2.84±0.25 | 2.96±0.27 |
| Lateral bending | 2.87±0.40 | 3.10±0.29 | 3.13±0.30 |
| Axial rotation | 2.32±0.23 | 2.46±0.20 | 2.47±0.15 |
ROM – range of motion;
Compared with Group A, P<0.05;
Compared with Group B, P<0.05;
Compared with Group A, P>0.05;
Compared with Group B, P>0.05.
IDP of the L3–L4 segment under 8.0 Nm torque (mean ± standard deviation, n=6, Unit: kPa).
| Motion | Group A | Group B | Group C |
|---|---|---|---|
| Flexion | 10.20±1.22 | 12.02±1.69 | 11.73±1.17 |
| Extension | 8.94±0.98 | 10.82±1.51 | 9.43±1.32 |
| Lateral bending | 9.42±0.69 | 11.26±1.13 | 11.13±1.00 |
| Axial rotation | 9.56±0.91 | 11.01±0.86 | 10.91±0.71 |
IDP – intradiscal pressure;
Compared with Group A, P<0.05;
Compared with Group B, P<0.05;
Compared with Group A, P>0.05;
Compared with Group B, P>0.05.
IDP of the L5–S1 segment under 8.0 Nm torque (mean ± standard deviation, n=6, Unit: kPa).
| Motion | Group A | Group B | Group C |
|---|---|---|---|
| Flexion | 10.51±1.27 | 11.72±0.94 | 11.33±0.96 |
| Extension | 9.23±0.85 | 10.17±0.66 | 10.07±0.17 |
| Lateral bending | 9.52±0.88 | 10.22±0.71 | 10.37±0.93 |
| Axial rotation | 9.72±0.91 | 10.53±0.79 | 10.71±0.88 |
IDP – intradiscal pressure;
Compared with Group A, P<0.05;
Compared with Group B, P<0.05;
Compared with Group A, P>0.05;
Compared with Group B, P>0.05.