Literature DB >> 27015445

Initial clinical use of a novel mechanical thrombectomy device, XCOILTM, in hemodialysis graft and fistula declot procedures.

Wayne L Monsky1, Richard E Latchaw.   

Abstract

PURPOSE: We aimed to evaluate the safety and effectiveness of a novel catheter-based mechanical thrombectomy device, XCOILTM, as a first line therapy to restore patency of thrombosed dialysis grafts and fistulae.
METHODS: In 2010, 18 consecutive/sequential patients (11 male, 7 female; median age, 52 years; age range, 32-69 years) with occluded arteriovenous grafts (n=15) or fistulae (n=3) were treated with XCOILTM (NexGen Medical Systems Inc.) without adjunctive thrombolytic drugs. XCOILTM was advanced distal to the thrombus within the outflow vein as well as distal to the arterial inflow platelet thrombin plug, using a 4F angiographic catheter. The percentage of thrombus cleared, primary patency, procedure time, and XCOILTM performance were documented.
RESULTS: Thrombosis occurred 1-30 days prior to the procedure. Thrombosed segments of graft/fistula measured 10-50 cm. Pre- and postprocedure angiography demonstrated that in 15 of 18 cases (83%) XCOILTM removed 80%-100% of the venous outflow thrombus. In 11 of 14 cases (79%), the platelet thrombin plug was also removed. Thrombectomy procedure time averaged 8 min, with one to three passes with the XCOILTM required. No evidence of distal embolization or graft/vessel injury was found on angiography following clot removal. In four cases in whom patency was not restored with XCOILTM, subsequent use of other clot removal devices also failed to restore patency. In one case with severe venous stenosis, the device failed to deploy and the thrombus was not captured. No intraprocedural complications related to XCOILTM use occurred.
CONCLUSION: XCOILTM is an effective and safe first-line therapy option for the treatment of thrombosed hemodialysis grafts/fistulae. Rapid removal of intact thrombus and platelet thrombin plug can be achieved without adjunctive thrombolytics.

Entities:  

Mesh:

Year:  2016        PMID: 27015445      PMCID: PMC4859742          DOI: 10.5152/dir.2015.15158

Source DB:  PubMed          Journal:  Diagn Interv Radiol        ISSN: 1305-3825            Impact factor:   2.630


  19 in total

Review 1.  Reporting standards for percutaneous interventions in dialysis access.

Authors:  Richard J Gray; David Sacks; Louis G Martin; Scott O Trerotola
Journal:  J Vasc Interv Radiol       Date:  2003-09       Impact factor: 3.464

Review 2.  Mechanical thrombectomy as a first-line treatment for arterial occlusion.

Authors:  K Kasirajan; J M Marek; M Langsfeld
Journal:  Semin Vasc Surg       Date:  2001-06       Impact factor: 1.000

3.  Hydrodynamic thrombectomy system versus pulse-spray thrombolysis for thrombosed hemodialysis grafts: a multicenter prospective randomized comparison.

Authors:  K H Barth; M R Gosnell; A M Palestrant; L G Martin; J B Siegel; T A Matalon; S C Goodwin; P A Neese; T L Swan; R Uflacker
Journal:  Radiology       Date:  2000-12       Impact factor: 11.105

Review 4.  Nephrology, dialysis and transplantation in Turkey.

Authors:  Ekrem Erek; Gültekin Süleymanlar; Kamil Serdengeçti
Journal:  Nephrol Dial Transplant       Date:  2002-12       Impact factor: 5.992

Review 5.  The role of interventional radiology in management of patients with end-stage renal disease.

Authors:  M Surlan; P Popovic
Journal:  Eur J Radiol       Date:  2003-05       Impact factor: 3.528

Review 6.  Transcatheter treatment of thrombosed hemodialysis access grafts.

Authors:  K Valji
Journal:  AJR Am J Roentgenol       Date:  1995-04       Impact factor: 3.959

7.  Hemolysis comparison between the OmniSonics OmniWave Endovascular System and the Possis AngioJet in a porcine model.

Authors:  Elvira V Lang; Anne M Kulis; Mario Villani; William Barnhart; Richard Balano; Robert Cohen
Journal:  J Vasc Interv Radiol       Date:  2008-06-02       Impact factor: 3.464

8.  Comparative in vitro study of five mechanical embolectomy systems: effectiveness of clot removal and risk of distal embolization.

Authors:  Thomas Liebig; Jörg Reinartz; Ralf Hannes; Elina Miloslavski; Hans Henkes
Journal:  Neuroradiology       Date:  2007-10-12       Impact factor: 2.804

9.  Clinical and economic evaluation of the trellis thrombectomy device for arterial occlusions: preliminary analysis.

Authors:  Timur P Sarac; Daniel Hilleman; Frank R Arko; Christopher K Zarins; Kenneth Ouriel
Journal:  J Vasc Surg       Date:  2004-03       Impact factor: 4.268

10.  A novel mechanical thrombectomy device for retrieval of intravascular thrombus.

Authors:  Wayne L Monsky; Stephanos Finitsis; Dino De Cicco; John M Brock; John Kucharczyk; Richard E Latchaw
Journal:  Cardiovasc Intervent Radiol       Date:  2010-11-12       Impact factor: 2.740

View more