Hemolysis comparison between the OmniSonics OmniWave Endovascular System and the Possis AngioJet in a porcine model.

Elvira V Lang1, Anne M Kulis, Mario Villani, William Barnhart, Richard Balano, Robert Cohen.   

Abstract

PURPOSE: To compare the degree of hemolysis generated during use of the ultrasound-based OmniSonics OmniWave Endovascular System and the Possis AngioJet connected to an Xpeedior-6 catheter.
MATERIALS AND METHODS: Twelve 64-145-kg Yorkshire pigs were used as the model because their weight, blood volume, and vessel diameters are comparable to that of adult humans. Six pigs were treated with the OmniWave and six with the AngioJet; half of the pigs in each group were treated in the contralateral iliofemoral arteries and half in the ipsilateral iliofemoral veins. Devices were activated for 10 minutes in patent vessels to represent a worst-case scenario for hemolysis generation. Plasma-free hemoglobin (PfHgb) levels, red blood cell counts, white blood cell (WBC) counts, and platelet counts were measured before and after the procedure, and PfHgb levels were measured every 2 minutes during treatment.
RESULTS: The OmniWave generated significantly lower PfHgb levels than the AngioJet (average, 228 vs 1,367 mg/dL; P < .001). The location of activation (arterial vs venous) had no significant effect (P = .575). Compared with OmniWave use, AngioJet use was associated with significantly greater changes from pre- to posttreatment values in WBCs (P = .002), platelets (P = .004), and creatinine (P = .002). AngioJet use was associated with a significant postprocedure decrease in the WBC count (mean, -8,000/microl; 95% confidence interval [CI]: -3,830, -12,170; P = .004) and platelet count (mean, -72,000; 95% CI: -25,000, -120,000; P = .011). With the OmniWave device there were no significant differences between pre- and posttreatment WBC count (P = .127), platelet count (P = .493), and creatinine level (P = .317).
CONCLUSIONS: The OmniWave generated approximately one-sixth of the hemolysis the AngioJet created. In addition, animals treated with the AngioJet exhibited a decrease in WBC and platelet counts.

Entities:  

Mesh:

Year:  2008        PMID: 18656016     DOI: 10.1016/j.jvir.2008.04.007

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  5 in total

Review 1.  A systematic review of ultrasound-accelerated catheter-directed thrombolysis in the treatment of deep vein thrombosis.

Authors:  Yadong Shi; Wanyin Shi; Liang Chen; Jianping Gu
Journal:  J Thromb Thrombolysis       Date:  2018-04       Impact factor: 2.300

2.  Initial clinical use of a novel mechanical thrombectomy device, XCOILTM, in hemodialysis graft and fistula declot procedures.

Authors:  Wayne L Monsky; Richard E Latchaw
Journal:  Diagn Interv Radiol       Date:  2016 May-Jun       Impact factor: 2.630

3.  Hemodynamic and Hematologic Effects of Histotripsy of Free-Flowing Blood: Implications for Ultrasound-Mediated Thrombolysis.

Authors:  Rajiv Devanagondi; Xi Zhang; Zhen Xu; Kimberly Ives; Albert Levin; Hitinder Gurm; Gabe E Owens
Journal:  J Vasc Interv Radiol       Date:  2015-05-04       Impact factor: 3.464

4.  A novel mechanical thrombectomy device for retrieval of intravascular thrombus.

Authors:  Wayne L Monsky; Stephanos Finitsis; Dino De Cicco; John M Brock; John Kucharczyk; Richard E Latchaw
Journal:  Cardiovasc Intervent Radiol       Date:  2010-11-12       Impact factor: 2.740

5.  Modulation of soluble guanylate cyclase ameliorates pulmonary hypertension in a rat model of chronic thromboembolic pulmonary hypertension by stimulating angiogenesis.

Authors:  John Zagorski; Evandro Neto-Neves; Nathan J Alves; Amanda J Fisher; Jeffrey A Kline
Journal:  Physiol Rep       Date:  2022-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.