Urinalysis: The Automated Versus Manual Techniques; Is It Time To Change?.

Asmaa Ismail Ahmed, Heba Baz, Sarah Lotfy.   

Abstract

BACKGROUND: Urinalysis is the third major test in clinical laboratory. Manual technique imprecision urges the need for a rapid reliable automated test. We evaluated the H800-FUSIOO automatic urine sediment analyzer and compared it to the manual urinalysis technique to determine if it may be a competitive substitute in laboratories of central hospitals.
METHODS: 1000 urine samples were examined by the two methods in parallel. Agreement, precision, carryover, drift, sensitivity, specificity, and practicability criteria were tested.
RESULTS: Agreement ranged from excellent to good for all urine semi-quantitative components (K > 0.4, p = 0.000), except for granular casts (K = 0.317, p = 0.000). Specific gravity results correlated well between the two methods (r = 0.884, p = 0.000). RBCS and WBCs showed moderate correlation (r = 0.42, p = 0.000) and (r = 0.44, p = 0.000), respectively. The auto-analyzer's within-run precision was > 75% for all semi-quantitative components except for proteins (50% precision). This finding in addition to the granular casts poor agreement indicate the necessity of operator interference at the critical cutoff values. As regards quantitative contents, RBCs showed a mean of 69.8 +/- 3.95, C.V. = 5.7, WBCs showed a mean of 38.9 +/- 1.9, C.V. = 4.9). Specific gravity, pH, microalbumin, and creatinine also showed good precision results with C.Vs of 0.000, 2.6, 9.1, and 0.00 respectively. In the between run precision, positive control showed good precision (C.V. = 2.9), while negative control's C.V. was strikingly high (C.V. = 127). Carryover and drift studies were satisfactory. Manual examination of inter-observer results showed major discrepancies (< 60% similar readings), while intra-observer's results correlated well with each other (r = 0.99, p = 0.000).
CONCLUSIONS: Automation of urinalysis decreases observer-associated variation and offers prompt competitive results when standardized for screening away from the borderline cutoffs.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27012033     DOI: 10.7754/clin.lab.2015.150520

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  3 in total

1.  Comments on evaluation of an oral sodium bicarbonate protocol for high-dose methotrexate urine alkalinization.

Authors:  Joanna Chyu; Zain Ali; Mohamad M Al-Rahawan
Journal:  Support Care Cancer       Date:  2022-10-19       Impact factor: 3.359

2.  Ascorbic acid-A black hole of urine chemistry screening.

Authors:  Adriana Unic; Nora Nikolac Gabaj; Marijana Miler; Jelena Culej; Adrijana Lisac; Anita Horvat; Nada Vrkic
Journal:  J Clin Lab Anal       Date:  2018-02-25       Impact factor: 2.352

3.  Automated urinalysis combining physicochemical analysis, on-board centrifugation, and digital imaging in one system: A multicenter performance evaluation of the cobas 6500 urine work area.

Authors:  Christa M Cobbaert; Figen Arslan; Imma Caballé Martín; Antoni Alsius Serra; Ester Picó-Plana; Víctor Sánchez-Margalet; Antonio Carmona-Fernández; John Burden; André Ziegler; Walter Bechel
Journal:  Pract Lab Med       Date:  2019-09-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.