Ascorbic acid-A black hole of urine chemistry screening.

Adriana Unic1, Nora Nikolac Gabaj1, Marijana Miler1, Jelena Culej1, Adrijana Lisac2, Anita Horvat1, Nada Vrkic1.   

Abstract

BACKGROUND: Study was performed in order: (i) to assess the comparability of glucose, bilirubin, hemoglobin, leukocyte esterase, and protein; (ii) to assess accuracy of glucose, bilirubin, hemoglobin, leukocyte esterase, and protein; and (iii) to evaluate interference of ascorbic acid on the glucose, bilirubin, hemoglobin, and nitrite determination using 2 different dipsticks: iChem Velocity, Iris Diagnostics and Combur-10M, Roche Diagnostics.
METHODS: Random urine specimens were included in the study. Comparability, accuracy, and ascorbic acid interference testing were performed.
RESULTS: Obtained results have shown almost perfect agreement for all parameters between 2 dipsticks in samples with negative ascorbic acid. Agreement in samples with positive ascorbic acid was not acceptable for bilirubin, protein, nitrite, and hemoglobin. Accuracy was not acceptable for hemoglobin and leukocyte esterase on both dipsticks. Ascorbic acid interference examination has shown that intensity of interference differs between dipsticks. Ascorbic acid interferes with glucose, hemoglobin, nitrite, and bilirubin at different concentrations causing false-negative results.
CONCLUSION: Obtained results indicate that it is necessary to determine diagnostic accuracy of used dipstick in order to define purpose of urinalysis. It is very important to choose dipstick with ascorbic acid indicator and to examine ascorbic acid impact on dipstick analytes independently of manufacturer claims.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  ascorbic acid; diagnostic accuracy; interference; patient safety; urinalysis

Mesh:

Substances:

Year:  2018        PMID: 29479748      PMCID: PMC6816895          DOI: 10.1002/jcla.22390

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  19 in total

Review 1.  Introduction to urinalysis: historical perspectives and clinical application.

Authors:  Germán Echeverry; Glen L Hortin; Alex J Rai
Journal:  Methods Mol Biol       Date:  2010

2.  Influence of Vitamin C on Urine Dipstick Test Results.

Authors:  Dae-Hyun Ko; Tae-Dong Jeong; Sollip Kim; Hee-Jung Chung; Woochang Lee; Sail Chun; Won-Ki Min
Journal:  Ann Clin Lab Sci       Date:  2015       Impact factor: 1.256

3.  Accuracy of dipstick urinalysis as a screening method for detection of glucose, protein, nitrites and blood.

Authors:  B Zamanzad
Journal:  East Mediterr Health J       Date:  2009 Sep-Oct       Impact factor: 1.628

Review 4.  Clinical Chemistry Laboratory Automation in the 21st Century - Amat Victoria curam (Victory loves careful preparation).

Authors:  David A Armbruster; David R Overcash; Jaime Reyes
Journal:  Clin Biochem Rev       Date:  2014-08

5.  The influence of vitamin C on the urine dipstick tests in the clinical specimens: a multicenter study.

Authors:  Wonmok Lee; Yukyung Kim; Soonhee Chang; A-Jin Lee; Chang-Ho Jeon
Journal:  J Clin Lab Anal       Date:  2016-10-24       Impact factor: 2.352

6.  High incidence of significant urinary ascorbic acid concentrations in a west coast population--implications for routine urinalysis.

Authors:  M L Brigden; D Edgell; M McPherson; A Leadbeater; G Hoag
Journal:  Clin Chem       Date:  1992-03       Impact factor: 8.327

7.  Ascorbate interference in the estimation of urinary glucose by test strips.

Authors:  B Berg
Journal:  J Clin Chem Clin Biochem       Date:  1986-01

8.  Ascorbic acid interference in reagent-strip reactions for assay of urinary glucose and hemoglobin.

Authors:  M H Zweig; A Jackson
Journal:  Clin Chem       Date:  1986-04       Impact factor: 8.327

Review 9.  The urine dipstick test useful to rule out infections. A meta-analysis of the accuracy.

Authors:  Walter L J M Devillé; Joris C Yzermans; Nico P van Duijn; P Dick Bezemer; Daniëlle A W M van der Windt; Lex M Bouter
Journal:  BMC Urol       Date:  2004-06-02       Impact factor: 2.264

10.  Interrater reliability: the kappa statistic.

Authors:  Mary L McHugh
Journal:  Biochem Med (Zagreb)       Date:  2012       Impact factor: 2.313

View more
  3 in total

1.  Discrepancy in results between dipstick urinalysis and urine sediment microscopy.

Authors:  Anca Bacârea; Gyula László Fekete; Bianca Liana Grigorescu; Vladimir Constantin Bacârea
Journal:  Exp Ther Med       Date:  2021-03-23       Impact factor: 2.447

2.  Ascorbic acid-A black hole of urine chemistry screening.

Authors:  Adriana Unic; Nora Nikolac Gabaj; Marijana Miler; Jelena Culej; Adrijana Lisac; Anita Horvat; Nada Vrkic
Journal:  J Clin Lab Anal       Date:  2018-02-25       Impact factor: 2.352

3.  Direct microfluidic antibiotic resistance testing in urine with smartphone capture: significant variation in sample matrix interference between individual human urine samples.

Authors:  Sarah Helen Needs; Sultan İlayda Dönmez; Alexander Daniel Edwards
Journal:  RSC Adv       Date:  2021-11-29       Impact factor: 3.361

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.